+8618117273997weixin
InggrisInggris
中文简体 中文简体 en English ru Русский es Español pt Português tr Türkçe ar العربية de Deutsch pl Polski it Italiano fr Français ko 한국어 th ไทย vi Tiếng Việt ja 日本語

Pendeteksi Bergerak Goniophotometer

Pengguna manual

berlaku LISUN Model: LSG-6000,LSG-6000B.LSG-6000BCCD,LSG-6000CCD,LSG-6000L、LSG-6000LCCD、LSG-6000S、LSG-6000SCCD

1. Membuka Kemasan dan Mengenal Peralatan Secara Dasar

1.1 Membuka Kemasan dan Pemeriksaan

When unpacking the device, handle it gently to avoid scratching the casing with sharp tools. After unpacking, first inspect the exterior of the device to ensure there are no visible deformations or dents, and that the buttons are not damaged or loose. If you notice any abnormalities with the device or its accessories, do not turn it on. Please contact us immediately.

1.2 System Configuration and Overview

1.2.1 Model Classification

Pilih Model Kategori Ukuran Differentiation by Core Features
LSG-6000S Kecil Includes near-field photometric probes and far-field photometric probes; no Spectroradiometer
LSG-6000SCCD Includes near-field photometric probes and far-field photometric probes; features a Spectroradiometer
LSG-6000 Model Standar Includes near-field and far-field photometric probes; no Spectroradiometer
LSG-6000CCD Includes near-field photometric probes and far-field photometric probes; features a Spectroradiometer
LSG-6000B Skala besar Includes near-field photometric probes and far-field photometric probes; no Spectroradiometer
LSG-6000BCCD Includes near-field photometric probes and far-field photometric probes; features a Spectroradiometer
LSG-6000L Sangat besar Includes near-field photometric probes and far-field photometric probes; no Spectroradiometer
LSG-6000LCCD Includes near-field photometric probes and far-field photometric probes; features a Spectroradiometer

1.2.2 Komponen Perangkat Keras Inti

Note: The table below lists the core hardware components of the entire system. The equipment you purchased may include only some of these components; please keep this in mind when following the instructions in this manual.

Nama komponen Pilih Model Fitur Inti
Goniophotomoeter Main Unit LSG-6000 Seri Used in conjunction with photometric probes and Spectroradiometers to measure photometric parameters and their spatial distribution, as well as chromaticity parameters and their spatial distribution.
Spectroradiometer LMS-9000CG Pre-installed in the LSG-6000 main unit. Used to test colorimetric parameters and their spatial distribution, such as color temperature and spatial color distribution.
Near-field photometric probe PM400C (N) Suitable for testing near-field photometric parameters and the spatial distribution of photometric parameters
Far-field photometric probe PM400C (F) Suitable for testing far-field photometric parameters and the spatial distribution of photometric parameters
Lampu standar SLS-150W Used for calibrating darkrooms; includes a photometric probe and a Spectroradiometer
E27 Fixture / For C-Gamma testing; suitable for incandescent bulbs
Penjepit Silang For C-Gamma testing; suitable for other indoor and outdoor lighting fixtures
Penjepit Persegi For B-Beta testing; suitable for floodlights
Cross-Line Laser Used during installation and commissioning; used during calibration and testing to determine the position and orientation of the standard lamp and the sample luminaire
Kabinet CASE-19 Used to house angle controllers, power meters, and DC and AC power supplies. Includes a power strip, an RS-485 communication module, and other components.
Pengontrol Sudut RT-300C Sudut di mana Goniophotometer diposisikan untuk tampilan dan kontrol. Umumnya, tidak diperlukan penyesuaian manual; kontrol ditangani oleh perangkat lunak.
Pengukur daya Seri LS Measure electrical parameters such as voltage, current, power, and power factor of the sample
DC Power Supply Seri DC Provides a stable DC power supply for DC samples, standard lamps, and auxiliary lamps
AC Power LSP Series Provides a stable, pure sine-wave AC power supply for AC samples

1.2.3 Software, manuals, certificates, etc.

At the time of shipment, a download link containing all the documents was sent via email. Please download them promptly. If you missed the email, you can also contact us directly to obtain the download link again.

Includes software, user manuals, warranty cards, standard lamp calibration certificates, and other important documents, as well as other relevant documents.

1.3 Parameter Pengukuran

LSG-6000 Series: Illuminance data, luminous intensity distribution, area luminous flux, luminaire efficiency, luminance distribution (optional), utilization factor, luminance limit curve, glare rating, isocandela curves, maximum allowable distance-to-height ratio, luminaire curve versus illumination area, isointensity curves, effective beam angle, EEI, UGR, etc.

In addition to testing the parameters listed above, the LSG-6000*CCD series of built-in spectroradiometers can also measure: CCT, CRI, TM-30-24, Spectrum, (x, y)/(u, v), PAR, PPF, and PPFD spatial distribution, among others.

2. Catatan Penting

2.1 Keamanan Listrik

The system must be connected to a stable AC power source of 220 V ±5%, with a frequency of 50 Hz or 60 Hz. Ensure proper grounding, with a grounding resistance of ≤4 Ω. Do not use the equipment if the power connections are loose or the cables are damaged.

2.2 Keselamatan Lingkungan

Avoid using the device in environments where humidity exceeds 70% (no condensation) or where the temperature is outside the range of 0–35°C. The optimal operating temperature is 25±2°C.

Baik ruang kontrol maupun ruang gelap harus dijaga agar tetap kering, berventilasi baik, bebas debu, dan bebas dari gas korosif, dengan suhu lingkungan 15–35°C dan kelembapan relatif ≤85%;

Keep the equipment away from strong magnetic fields, strong electric fields, or corrosive gases;

2.3 Prosedur Operasi Ruang Gelap

2.3.1 Deskripsi Ruang Gelap

The interior of the darkroom must be completely painted matte black—that is, a non-reflective, pure black—including all walls, ceilings, floors, partitions, doors, and so on. A small number of components, such as lighting fixtures, air conditioning units, and power switches, may not be painted black, provided they do not appear in the optical path between the Goniophotomoeter’s main unit and the photometric probe;

Once the equipment has been installed and commissioned, do not move any hardware within the darkroom, including the main unit base, the photometer probe base, the orientation of the photometer probe, or the orientation of the Optical Fiber; doing so may result in inaccurate test results;

When putting the darkroom into use, please keep it as dust-free as possible;

Ensure that the darkroom door is closed and that no external light enters the darkroom when the interior lighting is turned off.

2.3.2 Selama Penggunaan

Remove your shoes or put on shoe covers when entering the darkroom to prevent dust from being brought inside;

Set the air conditioner in the darkroom to 25°C and switch it to a gentle breeze setting to prevent strong air currents from affecting the test results;

The sample luminaire must be securely mounted to prevent it from falling and damaging the Goniophotomoeter during testing;

Jangan meletakkan benda apa pun di atas unit Goniophotometer atau di dalam jangkauan putarannya untuk menghindari kerusakan pada Goniophotometer;

During testing, the darkroom must remain sealed to prevent external light from interfering with the test results.

2.3.3 When Not in Use

Keep the darkroom closed when not in use; unauthorized persons are not allowed to enter, to prevent dust from getting into the darkroom.

2.4 Prosedur Pengoperasian Lampu Standar

The standard lamp has a lifespan of one year and is used only during installation and commissioning to calibrate photometric probes and Spectroradiometers. If the conditions of your darkroom remain unchanged (such as the color of the interior coating, internal dimensions, and the location and orientation of hardware), recalibration is not necessary;

If your laboratory requires annual calibration of the darkroom, you may purchase a new standard lamp;

The standard lamp is hot immediately after use; do not touch it. Allow it to cool for 10 minutes before returning it to the lamp case;

Avoid getting stains on the surface of standard light bulbs; if stains are present, they can no longer be used as standard lights;

The standard lamp used in the darkroom is a directional standard lamp, meaning that its color and luminous intensity data are specific to a particular direction. Do not remove the standard lamp bulb from its holder; doing so will render the standard lamp invalid.

2.5 Optical Fiber Operating Procedures

Optical Fiber must not be bent excessively (minimum bend radius ≥ 10 cm). Do not step on or squeeze the Optical Fiber, as this may damage it and cause abnormal test results.

2.6 Lainnya

Untuk hal-hal yang tidak tercakup dalam panduan ini, harap berhati-hati atau hubungi kami.

3. Instalasi Perangkat Lunak

3.1 Lingkungan Eksekusi Perangkat Lunak

Persyaratan Sistem: Windows 7/8/10/11 (32-bit/64-bit); komputer harus memiliki setidaknya satu port USB.

Instalasi 3.2

Jumlah Note
LSG Series Goniophotomoeter Software—please double-click to install
This APK file can be copied to an Android phone for installation. The app is intended solely for rotating the Goniophotomoeter and cannot be used for testing. Since the software includes all the app’s features, this manual uses the software as an example for operation.
RS485 Communication Module Driver—Double-click to install

Moving Detector Goniophotometer-Figure1

Gambar 0

4.1.3 Penanganan Pengecualian

Jika perangkat lunak antivirus atau Windows Firewall Anda menandai sebuah file sebagai mencurigakan selama proses instalasi perangkat lunak, harap nonaktifkan sementara perangkat lunak antivirus atau Windows Firewall tersebut, lalu coba instal perangkat lunak itu lagi.

4. Perakitan Perangkat Keras

4.1 Persiapan

4.1.1 Preparing Tools

Nama Note
Pita pengukur At least 10 meters in total length
Bor Palu Drill bit sizes: φ 8, φ 10, and φ 12
Obeng One set each of Phillips screwdrivers, flathead screwdrivers, and Allen wrenches
Double-ended wrench Satu set
12-inch Adjustable Wrench Bukaan maksimum: 36 mm
Electric or Manual Forklifts Load capacity: 2 metric tons; lifting height: 1.6 meters. Front fork length: 1.2 meters; width: 0.6 meters
Other Common Tools Gloves, hammers, pliers, electrical tape, paper, pens, etc.

Forklift Listrik:

Moving Detector Goniophotometer-Figure2

Gambar 1

Truk Palet Manual:

Moving Detector Goniophotometer-Figure3

Gambar 2

4.1.2 Darkroom Preparation

Please confirm that the darkroom has been constructed exactly according to the darkroom drawings we provided. If there are any discrepancies in the actual dimensions, please let us know in advance so we can adjust the installation plan to ensure that installation and commissioning proceed smoothly and that test results are accurate.

According to the darkroom drawings, there is a circular black cover in the exact center of the light-passing opening in the partition near the LSG-6000 Goniophotomoeter main unit. However, please do not install this cover yet; wait until all installation and commissioning steps are complete before installing it.

Moving Detector Goniophotometer-Figure4

Gambar 2.1

4.2 Cabinet Equipment Assembly

4.2.1 Introduction to the Cabinet

Place the cabinet on a level, sturdy workbench, leaving sufficient space for operation.

Jumlah Note
Wiring Diagram: Please assemble the equipment according to the wiring diagram.
Power distribution panel, used to connect the power cords of all devices inside the cabinet
Cooling fan power cord: Please connect it to the power strip inside the cabinet.
RS-485 Communication Box, used to connect all devices via an RS-485 communication cable
Please connect the power adapter for the RS485 communication box to the power strip inside the cabinet.
The USB cable for the RS485 communication box needs to be connected to the computer.
RT-300C Angle Controller Wiring
LS Series Power Meter Wiring
Koneksi Daya DC
AC Power Wiring
The main power cable for the cabinet, which supplies power to all equipment inside the cabinet, should be routed through cable opening ② and connected to the power source only after all equipment has been fully assembled.
1 Cable opening—the main power cord can be routed through here. Note: The main power cord plug is a Chinese standard plug; you may need to provide an adapter.

Moving Detector Goniophotometer-Figure5

Gambar 3

4.2.2 RT-300C Installation

For now, all you need to do is plug it in.

Moving Detector Goniophotometer-Figure6

Gambar 4

4.2.3 Installation of the LS Power Meter

Jumlah Note
Connected via an RS-485 communication cable; the other end is already connected to the RS-485 communication box inside the cabinet.
The power cord is connected; the other end is already connected to the power distribution panel inside the cabinet.
V1 test lead, connected to the red terminal of the power meter’s voltage V
V2 test lead, connected to the black terminal of the power meter’s voltage V
A1 test lead, connected to the red terminal of the power meter’s current A
A2 test lead, connected to the black terminal of the power meter’s current A

Moving Detector Goniophotometer-Figure7

Gambar 5

4.2.4 DC Power Supply Installation

Jumlah Note
Connected via an RS-485 communication cable; the other end is already connected to the RS-485 communication box inside the cabinet.
Connect the DC power output wires, making sure to match the positive and negative terminals.
Kabel daya sudah terpasang; ujung lainnya sudah terhubung ke stopkontak di dalam kabinet.

Moving Detector Goniophotometer-Figure8

Gambar 6

4.2.5 Instalasi Daya AC

Please locate the following AC power cord among the accessories:

Moving Detector Goniophotometer-Figure9

Gambar 7

Jumlah Note
Connected via an RS-485 communication cable; the other end is already connected to the RS-485 communication box inside the cabinet.
The power cord is connected; the other end is already connected to the power strip inside the cabinet.
Run the AC power output cord through port ④ to connect it to position ⑤.
Lubang kawat
Run the AC power output cable through cable opening ④ and connect it to position ③.

Moving Detector Goniophotometer-Figure10

Gambar 8

Moving Detector Goniophotometer-Figure11

Gambar 9

4.3 Assembly of the LSG-6000 Goniophotomoeter

4.3.1 Assembly of the Main Unit, Light Shaft, and Base

4.3.1.1 Main Unit and Base

Position the main unit and the base in the appropriate locations according to the darkroom drawings. Use the drawings below as an example.

Jumlah Note
LSG-6000 location—this is where you can place the main unit and the base
Position of the far-field photometric probe; assembly will be done later

Moving Detector Goniophotometer-Figure12

Gambar 9.1

To facilitate shipping, the LSG-6000 and higher-model units are disassembled into two parts for transport. Place the base and the main unit as shown in the illustration below. The front and rear cover plates on the base and the main unit can be removed directly to facilitate assembly.

Jumlah Note
Host Base
The side with the “FRONT” sticker should face the far end of the darkroom—that is, in the direction of the far-field probe.
Load-bearing beams, one on each side
Main unit. Note that the side with the rotating arm faces the far end of the darkroom—that is, the direction of the far-field probe.
Some of the wires left exposed here are high-voltage wires that have been wrapped in electrical tape. Please do not remove the electrical tape for the time being to prevent electric shock.

Moving Detector Goniophotometer-Figure13

Gambar 10

Note: The base and main unit of the LSG-6000S compact device are assembled before shipment.

Install and secure the two load-bearing crossbeams to the base. The long end of the crossbeams—indicated by the arrows in the figure below—is the far end of the anechoic chamber, which is the direction of the far-field probe.

Moving Detector Goniophotometer-Figure14

Gambar 11

Moving Detector Goniophotometer-Figure15

Gambar 12

4.3.1.2 Lamp Shaft Installation

Then locate the light shaft as shown in the figure below.

Moving Detector Goniophotometer-Figure16

Gambar 13

Remove the screws from the cover here, and pull the cover down.

Moving Detector Goniophotometer-Figure17

Gambar 14

Remove the six Allen screws from the main unit here, then pull out the terminal block and all the wires from inside the center hole.

Moving Detector Goniophotometer-Figure18

Gambar 15

Jumlah Note
The signal line is on the left.
The high-voltage wires are on the right; they must be kept separate from the signal wires to prevent interference.

Moving Detector Goniophotometer-Figure19

Gambar 16

With the lamp shaft facing to the right, lift it into place on the main unit and secure it firmly using the six Allen screws you removed earlier.

At least two people are needed to support the lamp shaft. Slightly loosen the two large Allen screws on the right side of the main unit (as shown in the figure below) by turning them counterclockwise. Then, manually lift the lamp shaft counterclockwise until it is in a vertical position. Finally, tighten the two screws on the right side of the main unit securely to lock the lamp shaft in the vertical position.

Moving Detector Goniophotometer-Figure20

Gambar 17

Moving Detector Goniophotometer-Figure21

Gambar 18

Note: These two screws are used to secure the lamp shaft; it takes considerable force to loosen them, so be sure to apply sufficient force when tightening them.

Then connect the cables on the host lamp shaft according to the labels.

Jumlah Note
The signal line is on the left.
The high-voltage wires are on the right; they must be kept separate from the signal wires and routed through two different openings in the light fixture to prevent interference.
Temperature and humidity probe—insert it into the dedicated port on the left side of the lamp shaft

Moving Detector Goniophotometer-Figure22

Gambar 19

4.3.1.3 Assembling the Main Unit and Base

The unit is secured to the bottom of the wooden crate with screws; please remove the mounting screws.

Moving Detector Goniophotometer-Figure23

Gambar 20

Next, use a forklift to lift the main unit and place it onto the base. Note: If the main unit is equipped with a light shaft at the front, its center of gravity will be shifted forward; take special care to prevent it from tipping over while lifting. Then, securely fasten the main unit to the base at both the front and rear using Allen screws.

Moving Detector Goniophotometer-Figure24

Gambar 21

4.3.1.4 Connecting the Host to the Cabinet
Jumlah Note
Connect the sample power cables and sampling cables to the cabinet, matching them one-to-one with their respective serial numbers.
Kabel Listrik
Please connect a separate ground wire to the main unit of the Goniophotomoeter whenever possible.
RS-485 communication cable, with the other end connected to the RS-485 communication box inside the cabinet
If you have a CCD-series model that includes a Spectroradiometer, this package includes an additional RS-485 communication cable, the other end of which connects to the RS-485 communication box inside the cabinet.
This is the RS-485 communication cable for the near-field photometer probe; the other end connects to the RS-485 communication box inside the cabinet.
Once everything is connected, turn on the host computer.

Moving Detector Goniophotometer-Figure25

Gambar 22

Next, turn on the main power switch for the cabinet, and then turn on the power to each piece of equipment inside the cabinet one by one.

4.3.2 Assembly of Spindle Components

4.3.2.1 Installing the Rearview Mirror Bracket

The arrow points to the main axis of the goniophotomoeter (G/Gamma axis).② The end with the wire is used to connect to the near-field probe. The end ①without the wire is used to connect to the reflector.

Jumlah Note
The end without the pre-installed cable is used to mount the mirror bracket and the mirror.
Use this end of the reserved cable to install the near-field mount, the near-field photometer, and the Spectroradiometer.

Moving Detector Goniophotometer-Figure26

Gambar 23

Double-click to open the software; the first time you open it, it will automatically open the system configuration screen.

Jumlah Note
For the model of the goniophotomoeter, click the drop-down list on the right to select one. The model you purchased is the LSG-6000.
Near-field photometer probe; select photometer probe model PM400C
Far-field photometer probe; select photometer probe model PM400C
Near-field spectroradiometer; select spectroradiometer model LMS-9000CG (if available)
AC power supply models are typically from the LSP Series
DC power supply models are generally designated as the DC Series
Power meter model: The model number is printed on the front panel of the power meter. Please check it and select the appropriate model.
Once you have selected all the correct models, click “Auto-Detect Port.”
Click “OK” to save and exit.

Moving Detector Goniophotometer-Figure27

Gambar 24

Note: If the near-field photometer, far-field photometer, and Spectroradiometer are not currently connected, a “Communication Failed” message will appear during the automatic port search. Simply click “OK.” Once all devices are connected, you can attempt communication again.

The software automatically opens the screen for saving the database file. After that, all test reports will be automatically saved in this database file.

Jumlah Note
Please select the path where you want to save the database file. Please do not save it to the desktop or inside the desktop folder.
Masukkan nama file
Just click “Save”

Moving Detector Goniophotometer-Figure28

Gambar 25

The software will then automatically open to its main interface.

Jumlah Note
Click to switch angle modes
Click ① to display the C-Gamma angle here instead of the B-Beta angle.
Then click to open the angle control interface

Moving Detector Goniophotometer-Figure29

Gambar 26

Jumlah Note
Select a far-field photometric probe from the Gamma angle drop-down list
You can enter 0
Then click in any direction to rotate the mirror bracket’s mounting end so that it faces downward for easier installation.

Moving Detector Goniophotometer-Figure30

Gambar 27

Find the mirror bracket.

Moving Detector Goniophotometer-Figure31

Gambar 28

When installing, secure all screw holes firmly with hex screws.

Moving Detector Goniophotometer-Figure32

Gambar 29

Moving Detector Goniophotometer-Figure33

Gambar 30

4.3.2.2 Installing the Near-Field Bracket

However, the mounting end of the near-field bracket can be rotated to a horizontal position to facilitate installation.

Moving Detector Goniophotometer-Figure34

Gambar 31

Note: Before all components are installed on the spindle, the weight distribution at both ends is uneven; therefore, the spindle may not rotate when controlled by the software. This manifests as the software freezing with no response and the spindle failing to rotate. If this occurs, turn off the power to the LSG-6000 Goniophotomoeter main unit, wait one minute, and then turn it back on. Apply a constant force to push or pull the spindle, then simultaneously use the software to control its rotation. Ensure that the external force you apply is in the same direction as the spindle’s rotation. Throughout the process of installing the spindle accessories, you may need to repeat this procedure several times to complete the installation.

Be sure to thread the reserved cable through the holes in the near-field bracket. Then, secure all screw holes firmly using Allen screws.

Moving Detector Goniophotometer-Figure35

Gambar 32

4.3.2.3 Connecting the Near-Field Probe and The Spectroradiometer

Be sure to turn off the power to the LSG-6000 Goniophotomoeter main unit first.

Jumlah Note
The red and black wires are the power wires; connect them according to the labels.
The RS485 communication cable is used to connect the LMS-9000CG Spectroradiometer

Moving Detector Goniophotometer-Figure36

Gambar 34

Moving Detector Goniophotometer-Figure37

Gambar 35

Moving Detector Goniophotometer-Figure38

Gambar 36

I found the following two photometric probes.

Moving Detector Goniophotometer-Figure39

Gambar 37

F stands for far field, and N stands for near field.

Moving Detector Goniophotometer-Figure40

Gambar 38

Install the near-field photometer and the Optical Fibre of the Spectroradiometer as shown in the figure below.

Moving Detector Goniophotometer-Figure41

Gambar 39

Moving Detector Goniophotometer-Figure42

Gambar 40

4.3.2.4 Installing Reflectors

Silakan mulai ulang LSG-6000 goniophotomoeter. The software will control the mirror mount to rotate it to the appropriate lower position.

Moving Detector Goniophotometer-Figure43

Gambar 41

Locate the reflector. When handling the reflector, be sure to grip the metal bracket on the back of the reflector rather than the reflector itself to avoid damaging it.

Moving Detector Goniophotometer-Figure44

Gambar 42

Align the three screw posts on the reflector with the three holes on the bracket, insert them, and secure each screw post with an inner and an outer screw, as shown in the figure below.

Moving Detector Goniophotometer-Figure45

Gambar 43

Once all components on the spindle have been fully installed and the counterweights at both ends are balanced, the spindle will rotate smoothly when the rotation is restarted via software, without the need for any external force.

4.3.3 Installing the Far-Field Photometric Probe

Matikan LSG-6000 Goniophotomoeter main unit to facilitate wiring. The image below shows the internal cables of the main unit; we secure them with cable ties before shipment. Please remove the cable ties first to facilitate wiring.

Moving Detector Goniophotometer-Figure46

Gambar 44

After removing the cable ties, unscrew the mounting screws on the junction box, and you can pull the junction box out. Then remove the cover to make it easier to connect the wires.

Moving Detector Goniophotometer-Figure47

Gambar 45

Moving Detector Goniophotometer-Figure48

Gambar 46

Locate the far-field photometric probe mount and position it roughly where indicated on the darkroom drawings.

Moving Detector Goniophotometer-Figure49

Gambar 47

Cables required to connect the far-field photometric probe.

Jumlah Note
Power Cable for Far-Field Photometric Probe
Trigger wire: According to the cabinet wiring diagram, this must be connected to the host terminal block.
The RS-485 communication cable must be connected to the RS-485 communication box inside the cabinet, as indicated in the cabinet wiring diagram.

Moving Detector Goniophotometer-Figure50

Gambar 48

The remaining three wires need to be connected to the junction box inside the main unit.

Moving Detector Goniophotometer-Figure51

Gambar 49

Starting from the bottom of the console:

Moving Detector Goniophotometer-Figure52

Gambar 50

and feed them through the reserved wiring openings:

Moving Detector Goniophotometer-Figure53

Gambar 51

Then connect these three wires correctly to the three locations shown in the figure below.

Moving Detector Goniophotometer-Figure54

Gambar 52

Note: When connecting the red, yellow, white, and black wires to the terminal block, a wiring tool is provided at that terminal.

Remove the outer sleeve from the far-field probe.

Moving Detector Goniophotometer-Figure55

Gambar 53

If the angle is incorrect and it is difficult to disassemble, please turn on the power to all devices.

Jumlah Note
The synchronization axis is the angle of the far-field photometric probe
You can enter 90
Click in any direction to rotate it 90 degrees, making it easier to remove the screws.

Moving Detector Goniophotometer-Figure56

Gambar 54

Install the far-field photometric probe and connect all the wires on the back of the probe.

Moving Detector Goniophotometer-Figure57

Gambar 55

All hardware connections are complete. All devices are plugged in, and make sure all devices are turned on.

Jumlah Note
Click the red logo in the upper-left corner of the software
Click to return to the system configuration screen.

Moving Detector Goniophotometer-Figure58

Gambar 56

As described in Section 4.3.2.1, please ensure that all devices are successfully communicating.

5. Calibration of darkroom equipment (Each step must be performed carefully and thoroughly; otherwise, it may result in incorrect calibration data and abnormal test results.)

5.1 Single-Line Laser Calibration

Turn on the single-button laser power switch.

Moving Detector Goniophotometer-Figure59

Gambar 57

The single-character laser was turned on.

Moving Detector Goniophotometer-Figure60

Gambar 58

Rotate the main shaft of the Goniophotomoeter to the position shown in the figure below to facilitate the following steps.

Moving Detector Goniophotometer-Figure61

Gambar 59

Please hold the lamp shaft steady and loosen these two screws again.

Moving Detector Goniophotometer-Figure62

Gambar 60

Hold the lamp shaft steady and slowly rotate it 180 degrees clockwise in the direction shown in the figure below.

Moving Detector Goniophotometer-Figure63

Gambar 61

Note: A counterclockwise stop has been installed at this point on the lamp shaft, so it can only be rotated clockwise.

The straight laser beam from the main unit will then trace a semicircular path on the wall opposite the darkroom. You can attach a sheet of white paper to the wall and trace the laser path with a water-based marker. If the radius of the semicircle is very small (for example, 0.5 centimeters) or even close to a single point, no adjustment is necessary. If the radius of the semicircle is larger, adjust the four small Allen screws around the straight-line laser to position the laser spot at the center of the semicircle.

Moving Detector Goniophotometer-Figure64

Gambar 62

Once adjusted, hold the lamp shaft steady and rotate it 180 degrees counterclockwise to return it to its original position, while rechecking that the straight laser beam’s path meets the requirements. After debugging, ensure that all four small Allen screws on the straight laser are fully tightened. Rotate the lamp shaft back to its initial position and hold it steady, then fully tighten the two large screws on the side of the main unit to secure the lamp shaft.

5.2 Adjusting the Host’s Position

First, move the main unit to its approximate location based on the darkroom blueprints. Then, fine-tune the position of the main unit and adjust the height of all the large screws on the support beam to adjust the main unit’s tilt angle.

Moving Detector Goniophotometer-Figure65

Gambar 63

Ensure that the host’s final position meets the following two conditions.

① The single-line laser passes through the exact center of the light-transmitting holes in all the partitions.

Moving Detector Goniophotometer-Figure66

Gambar 64

② The light shaft must be vertical (with no tilt in any direction); you can use a cross-line laser to verify this.

Find the cross-hair laser and the tripod.

Moving Detector Goniophotometer-Figure67

Gambar 65

Remove the cross-line laser and the batteries. Insert the batteries into the cross-line laser. The power switch is located on the side of the device. Turn the switch to the “unlock” position to turn on the device.

Jumlah Note
Power switch: Move it to the “Unlock” position to turn on the power.
VH refers to vertical and horizontal laser switches, respectively

Moving Detector Goniophotometer-Figure68

Gambar 66

Note: The cross-line laser is self-leveling. If it is not self-leveled, an audible alarm will sound. This may be due to an uneven surface or an improperly set up tripod. Please ensure that the cross-line laser remains self-leveled at all times during use.

5.3 Adjust and Secure the Far-Field Luminance Probe Mount

There is a small mirror inside the standard light box; remove it.

Moving Detector Goniophotometer-Figure69

Gambar 67

Moving Detector Goniophotometer-Figure70

Gambar 68

Temporarily remove the far-field photometer probe, then install the small mirror in the position shown in the figure below.

Moving Detector Goniophotometer-Figure71

Gambar 69

Please adjust the position of the far-field probe bracket. Then remove the screws shown in the figure below.

Moving Detector Goniophotometer-Figure72

Gambar 70

You can adjust the height of the probe bracket by turning the two pairs of large screws here.

Moving Detector Goniophotometer-Figure73

Gambar 71

The goal is to have the straight laser beam from the host unit strike the exact center of the small mirror, so that the reflected straight laser beam travels back along the same path.

Moving Detector Goniophotometer-Figure74

Gambar 72

Once the location has been determined, use expansion screws to secure the probe bracket base to the floor.

Moving Detector Goniophotometer-Figure75

Gambar 73

Once it is securely fastened, double-check that the straight laser beam coming from the main unit hits the small mirror and that the reflected beam follows the same path back. After adjusting and securing it, you can remove the small mirror.

Reinstall the far-field photometric probe. Then attach the small mirror to the probe, as shown in the figure below.

Moving Detector Goniophotometer-Figure76

Gambar 74

5.4 Setting the Spindle’s 0-Degree Position

Jumlah Note
Still opting for a far-field photometric probe
Masukkan 0
Check the “Synchronize Axis” box
Click any button to simultaneously rotate the main shaft and the synchronous shaft (i.e., the far-field photometric probe) to the 0-degree position; at this point, the reflector should be directly below the main shaft.

Moving Detector Goniophotometer-Figure77

Gambar 75

Use the cross-hair laser to verify that the main axis is perpendicular to the lamp axis. If not, use the following method to fine-tune the main axis until it is perpendicular, then set it to 0 degrees.

Jumlah Note
Still opting for a far-field photometer
Please always check the “Sync Axis” box.
Based on the actual situation, enter a small angle, such as 1, 0.5, 0.1, 0.05, etc.
Based on the actual situation, choose a direction to rotate. Repeat steps ③ and ④ until the spindle is rotated to a vertical position.
Click to set the current angle to 0 degrees. Password: LISUNGROUP

Moving Detector Goniophotometer-Figure78

Gambar 76

5.5 Adjusting the Direction of the Reflector and the Far-Field Photometric Probe

Jumlah Note
There is a switch at the top of the lamp post; turn the switch on.
A straight laser beam will be projected downward from the center of the lamp shaft.

Moving Detector Goniophotometer-Figure79

Gambar 77

You need to adjust the three sets of screws on the fixed rearview mirror to adjust the mirror’s reflection angle.

Moving Detector Goniophotometer-Figure80

Gambar 78

At the same time, you’ll need to adjust the screws on the far-field photometer probe bracket to change the height and orientation of the far-field photometer probe.

Moving Detector Goniophotometer-Figure81

Gambar 79

This causes the straight laser beam emitted downward from the lamp shaft to strike the reflector; after being reflected by the reflector, it strikes the exact center of the small mirror on the far-field photometer probe, and after being reflected once more, it returns along the same path.

Moving Detector Goniophotometer-Figure82

Gambar 80

Moving Detector Goniophotometer-Figure83

Gambar 81

Once calibration is complete, all screws should be fully tightened. Remove the small mirror attached to the far-field photometric probe, and reinstall the far-field probe sleeve and housing.

5.6 Adjusting the Near-Field Photometric Probe

Jumlah Note
Selecting a Near-Field Photometric Probe
Masukkan 0
Choose one of the following: Rotate the near-field photometer probe to the 0-degree position.

Moving Detector Goniophotometer-Figure84

Gambar 82

Attach the small mirror to the near-field photometer. Adjust the position and orientation of the near-field photometer by turning the screws on its mounting bracket so that the laser beam from the lamp axis strikes the exact center of the small mirror and the reflected light returns along the same path. Once adjusted, all screws should be fully tightened.

Moving Detector Goniophotometer-Figure85

Gambar 83

If the angle is too far off and cannot be adjusted to the correct position, you may need to do the following.

Jumlah Note
Still opting for a near-field photometer
Based on the actual situation, enter a smaller angle, such as 1, 0.5, 0.1, etc.
Based on the actual conditions, rotate the near-field photometer in one direction by the angle entered above. Then readjust the position and orientation of the near-field photometer until the laser beam from the lamp axis hits the exact center of the small reflector and the reflected light returns along the same path. Once adjusted, all screws should be fully tightened.
Then click to reset the 0-degree position for the near-field photometric probe. Password: LISUNGROUP

Moving Detector Goniophotometer-Figure86

Gambar 84

5.7 Adjusting the Direction of The Optical Fiber Cable

Select a spectroradiometer.

Moving Detector Goniophotometer-Figure87

Gambar 85

For the remaining steps, please refer to Section 5.6. Once the adjustment is complete, you can turn off the flat-head laser switch on the lamp shaft.

5.8 Installing the Baffle and Cross-Line Laser

The dimensions marked in the figure below are the dimensions of the baffle.

Moving Detector Goniophotometer-Figure88

Gambar 86

Jumlah Note
Install the baffle so that it is centered over the light hole in the first partition.
The straight laser beam from the host center should also be aimed at the exact center of the baffle; please mark it.

Moving Detector Goniophotometer-Figure89

Gambar 87

Find the small cross-shaped laser shown below.

Moving Detector Goniophotometer-Figure90

Gambar 88

Jumlah Note
Remove the straight-line laser from the center of the treadmill and replace it with a small cross-shaped laser.
After connecting the power cord, turn on the laser switch. Adjust the direction of the crosshair laser so that its center aligns with the mark you just made and the crosshair is perfectly level and plumb, then secure the small crosshair laser in place.

Moving Detector Goniophotometer-Figure91

Gambar 89

5.9 Equipment and Darkroom Cleaning

Once all installation and commissioning steps are complete, shut down all equipment. Note: The shutdown sequence is as follows: first, turn off the power to each piece of equipment one by one, and then turn off the main power switch for the cabinet.

After cleaning the equipment with a dry, soft cloth, reinstall all covers, screws, and other parts.

Then, cover all equipment with dust-proof cloths and clean the darkroom and control room. Once everything has been cleaned, turn on all the equipment.

6. Kalibrasi

6.1 Installation and Commissioning of Standard Lamps

6.1.1 Installation of Standard Lamps

Find the standard lamp and its extension pole.

Moving Detector Goniophotometer-Figure92

Gambar 90

Assemble the standard lamp as shown in the figure below, and install the small mirror as well.

Moving Detector Goniophotometer-Figure93

Gambar 91

Go to the software angle control interface and rotate the C-axis to the 0-degree position.

Moving Detector Goniophotometer-Figure94

Gambar 92

Attach the standard lamp to the extension rod, and then mount the extension rod to the lamp shaft.

Jumlah Note
There is a slot on the extension rod. After removing the bolt from this slot, insert the extension rod all the way to the bottom and then loosen the bolt; this will lock it in place and prevent it from falling out.
Turn the screw by hand; turning it counterclockwise loosens it, and turning it clockwise tightens it.

Moving Detector Goniophotometer-Figure95

Gambar 93

Secure the standard lamp in the orientation shown in the figure below.

Moving Detector Goniophotometer-Figure96

Gambar 94

There are four terminals on the light fixture. Please follow the diagram on the light fixture: short-circuit the red and yellow terminals together, and short-circuit the blue and black terminals together, then connect the red and blue wires to the standard light fixture power cord.

Moving Detector Goniophotometer-Figure97

Gambar 95

Note: The connection method for the power supply of the standard lamp and the sample lamp is the same.

6.1.2 Adjusting the Orientation of the Standard Lamp

Jumlah Note
Turn on the switch here
Use the up/down switch to adjust the lamp shaft—that is, the height of the standard lamp.

Moving Detector Goniophotometer-Figure98

Gambar 96

The three screws on the standard lamp bracket allow you to adjust the pitch angle of the standard lamp.

Moving Detector Goniophotometer-Figure99

Gambar 97

The standard lamp should be adjusted so that the cross-shaped laser beam is directed at the center of the small mirror, and the horizontal and vertical laser beams reflected from the mirror coincide exactly with the original cross-shaped laser beam.

Moving Detector Goniophotometer-Figure100

Gambar 98

Based on our experience, follow these steps for calibration: First, adjust the standard lamp to the appropriate height. The vertical reflected laser can be calibrated by rotating the extension rod, while the horizontal reflected laser can be calibrated by adjusting the standard lamp’s pitch. After calibration, the center height of the standard lamp may change, requiring you to readjust its height. Repeat this process until the crosshair laser is projected onto the center of the small mirror, and the reflected horizontal and vertical laser beams perfectly align with the original crosshair laser.

Once the standard lamp is properly aligned, carefully remove the small reflector, taking care not to alter the alignment of the standard lamp during removal. Then turn off the cross-hair laser.

6.2 Calibration of the Far-Field Photometric Probe

All standard lamps are constant-current DC light sources. Set the AC/DC switch on the 19-inch Standard Instrument Cabinet to DC.

Moving Detector Goniophotometer-Figure101

Gambar 99

Open the software’s angle control interface.

Jumlah Note
Rotate the C-axis to the 90-degree position
Be sure to check the “Synchronize Axes” box.
Then select the far-field photometer probe and rotate it to the 90-degree position.
The synchronizing shaft rotates in sync with the rearview mirror to the 90-degree position.

Moving Detector Goniophotometer-Figure102

Gambar 100

Once you have moved to the target position, the round hole in the standard light bracket should be directly aligned with the rearview mirror.

Moving Detector Goniophotometer-Figure103

Gambar 101

Turn on the calibration lamp and enter the calibration screen.

Jumlah Note
Select a DC Power Supply
The standard lamp is a constant-current light source. Select CC (constant-current mode) and enter the reference voltage and rated current as specified in the standard lamp certificate; these are typically 30 V and 6.1 A.
Click to turn on the standard lamp and turn off all other lights in the darkroom. The standard lamp takes 15 minutes to reach a steady state.
Once the standard lamp has stabilized, click to enter the photometer calibration interface.

Moving Detector Goniophotometer-Figure104

Gambar 102

Close the darkroom door and make sure that all light sources in the darkroom are turned off, except for the standard lamp.

Jumlah Note
Selecting a Far-Field Photometric Probe
Luminous flux calibration—our software includes this option, but current standard practices require the use of luminous intensity calibration, and the standard lamps included in our standard configuration are also luminous intensity standard lamps. Therefore, there is no need to use luminous flux calibration at this time.
Light Intensity Calibration Method
Refer to the standard lamp calibration certificate and enter its calibrated luminous intensity value.
Klik untuk Memulai
The software will automatically calculate the test distance.
To verify the results, click “Check,” and the software will display the current measured light intensity value. The measured light intensity value should match the calibrated light intensity value.
Click “OK” to save and exit.

Moving Detector Goniophotometer-Figure105

Gambar 103

Note: The test distance for the far-field photometric probe is the distance from the center of the Goniophotomoeter to the reflector, plus the distance from the reflector to the surface of the far-field photometric probe. The distance calculated by the software should be close to the actual distance; an error of 1% or less is considered normal. If the error is too large, please refer to Section 2.3.1 to verify that the darkroom is functioning properly, and refer to Chapter 5 to review the calibration procedure.

6.3 Calibration of Near-Field Photometric Probes

The reference lamp remains lit and stable. Enter the angle control interface.

Jumlah Note
The C-axis is still at the 90-degree position
Select the near-field photometer probe and rotate it to the 90-degree position
. Note: When calibrating the near-field, you do not need to worry about the position of the synchronization axis.

Moving Detector Goniophotometer-Figure106

Gambar 104

Once the lamp has been rotated to the target position, the round hole in the standard lamp bracket is directly aligned with the near-field photometer probe.

Moving Detector Goniophotometer-Figure107

Gambar 105

Go to the photometric probe calibration interface and select the near-field photometric probe. The remaining steps are exactly the same as those for calibrating the far-field photometric probe; please refer to Section 6.2.

Moving Detector Goniophotometer-Figure108

Gambar 106

Note: The standard distance for the LSG-6000S near-field photometric probe is 1.25 m, and the standard distance for the LSG-6000 near-field photometric probe is 1.5 m.while the standard measurement distance for the LSG-6000Pita LSG-6000L near-field photometric probes is 1.75 m.

6.4 Calibration of The Spectroradiometer (if applicable)

The reference lamp remains lit and stable. Enter the angle control interface.

Jumlah Note
The C-axis is still at the 90-degree position
Select the spectroradiometer and rotate it to the 90-degree position.

Moving Detector Goniophotometer-Figure109

Gambar 107

Once the device has been moved to the target position, the circular opening in the standard lamp mount should be aligned directly with the Optical Fiber path of the Spectroradiometer.

Click to open the Spectroradiometer calibration interface.

Moving Detector Goniophotometer-Figure110

Gambar 108

Jumlah Note
For the points accumulation time, be sure to always select “Automatic.”
First, click “Sampling.” Once sampling is complete, the corresponding curve will appear on the screen. After sampling is complete, the “Calibrate” button will become clickable.
Refer to the standard lamp calibration certificate and enter its calibrated color temperature value.
Click “Calibrate.” Once calibration is complete, the corresponding curve will appear on the screen.
Click OK to save and exit.

Moving Detector Goniophotometer-Figure111

Gambar 109

After calibration is complete, you can turn off the standard lamp using the software. Since the standard lamp becomes quite hot during use, wait for it to cool down before placing it back in its packaging and storing it safely along with the extension rod.

6.5 Kalibrasi dan Pemeliharaan

No changes in the darkroom If there have been no changes to the computer, recalibration is not necessary.
No changes were made to the computer; simply unplug and replug the communication cable or reinstall the Goniophotomoeter software—no need to recalibrate.
If you replace your computer or reinstall the operating system, the photometric probe does not need to be recalibrated; simply enter the previous measurement distance and save it. However, the Spectroradiometer does need to be recalibrated (if you do not have the means to recalibrate it, please contact us to restore the calibration data).
Changes in the Darkroom The structure of the darkroom remains unchanged; however, after prolonged use, the interior paint may fade or peel. Depending on how well you maintain the darkroom and your requirements for the accuracy of test results, you may repaint the darkroom periodically. After repainting, the darkroom must be recalibrated.
There has been a change in the darkroom configuration, so recalibration is required.

7. Persiapan Sebelum Pengujian

7.1 Lamp Selection and Aging

Select representative luminaires for testing that comply with national or corporate standards; new luminaires must undergo aging treatment (for LED luminaires, a 24-hour aging period is recommended);

Check that the light fixture has no visible damage, that the internal wiring is properly connected, and clean any stains from the surface of the fixture.

7.2 Introduction to Test Modes

C-Gamma Mode: Suitable for most lighting fixtures, such as streetlights, downlights, panel lights, and plant grow lights.

Moving Detector Goniophotometer-Figure112

Gambar 109.1

Beta Mode: Suitable for spotlights, floodlights, and similar fixtures; requires installation of an auxiliary arm.

Moving Detector Goniophotometer-Figure113

Gambar 109.2

Alpha Mode: Suitable for automotive lights and signal lights.

Moving Detector Goniophotometer-Figure114

Gambar 109.3

Catatan: LSG-6000 series can perform C-Gamma and B-Beta tests. For A-Alpha testing, a Type A Goniophotomoeter is required; please contact us if needed.

8. C-Gamma Test

8.1 Luminous Intensity Distribution Test

8.1.1 Installing and Turning On the Light Fixture

8.1.1.1 Pemasangan Perlengkapan Lampu

We have two C-Gamma test fixtures; this procedure uses the cross-shaped fixture as an example.

Jumlah Note
Large cross-shaped clamps are used for various types of lighting fixtures
E27 Fixture for Testing Globe Bulbs

Moving Detector Goniophotometer-Figure115

Gambar 110

Moving Detector Goniophotometer-Figure116

Gambar 111

Attach the fixture to the Goniophotomoeter using the same method as for installing the standard lamp extension rod.

Moving Detector Goniophotometer-Figure117

Gambar 112

Adjust the height of the light axis and the cross-shaped clamp so that the light-emitting surface of the sample is level with the cross-shaped laser.

Moving Detector Goniophotometer-Figure118

Gambar 113

Caution is required when installing directional luminaires. When the angle of the Goniophotomoeter’s C-plane is 0 degrees, according to streetlight installation standards, the light pole should face the Goniophotomoeter’s main unit; when the angle of the Goniophotomoeter’s C-plane is 0 degrees, according to industry standards for linear luminaires, the light pole should be perpendicular to the Goniophotomoeter’s main unit.

Once the light fixture is installed, you can turn off the crosshair laser.

8.1.1.2 Lighting the Lamps

Connect the power cord to the luminaire under test. The connection method for the luminaire under test is the same as that for the standard luminaire; refer to Section 6.1.1.

Adjust the cabinet switch according to the parameters of the luminaire being tested.

Jumlah Note
If the sample luminaire is powered by AC, switch to the AC position; if it is powered by DC, switch to the DC position.
Internal/External Sampling Switch for Electrical Parameters. External sampling uses a four-wire configuration, which provides more accurate voltage readings. Unless there are specific requirements, please set this switch to the external sampling position.

Moving Detector Goniophotometer-Figure119

Gambar 114

If the sample luminaire is AC-powered.

Jumlah Note
Select AC Power
Based on the sample parameters, enter the correct voltage and frequency.

Moving Detector Goniophotometer-Figure120

Gambar 115

If the sample luminaire is DC-powered.

Jumlah Note
Select a DC Power Supply
Based on the sample parameters, enter the correct voltage and current.
Select the output mode: CC stands for constant current, and CV stands for constant voltage. If you select constant current, you need to enter the sample’s rated current value; for voltage, simply enter the maximum output voltage of this DC power supply. If you select constant voltage, enter the sample’s rated voltage; for the current value, simply enter the DC power supply’s maximum output current.
Note: If your DC power supply model is DC3010Tegangan keluaran maksimumnya adalah 30 V dan arus keluaran maksimumnya adalah 10 A.

Moving Detector Goniophotometer-Figure121

Gambar 116

The software offers two methods for turning the sample lamp on and off, and users can choose either one.

Jumlah Note
If you check the “Auto-On” and “Auto-Off” options, the software will automatically power on the sample according to the selected power source and the entered electrical parameters when the test begins, and will automatically cut off the sample’s power when the test ends.
If the “Auto On” and “Auto Off” options are not checked, you can still click this button; the software will power on the sample based on the selected power source and the entered electrical parameters. Clicking this button again will cut off the power to the sample.

Moving Detector Goniophotometer-Figure122

Gambar 117

8.1.2 Pengaturan Parameter Pemanasan Awal

Jumlah Note
If “Preheat” is checked, the software will preheat the luminaire according to the configured preheat parameters when the test begins, and the test will start automatically once preheating is complete.
If “Preheat” is not checked, the software will begin the test immediately.
Preheating Parameter Settings

Moving Detector Goniophotometer-Figure123

Gambar 118

Click “Preheat Parameter Settings.”

Jumlah Note
Warm-up Duration Setting: The purpose of the warm-up is to allow the light fixture to remain on for a period of time until its luminous and chromatic parameters stabilize, after which testing begins automatically.
Sampling interval: The software will automatically take samples at the set interval throughout the entire warm-up period to obtain real-time data. We recommend setting the interval to 5 seconds or longer.
If checked, the test will automatically skip the remaining warm-up time and begin once the light fixture has stabilized; if unchecked, the warm-up will complete according to the set duration before the test begins.
Definition of a luminaire’s steady state. If the values are set to 20 and 0.5%, the luminaire is considered to be in a steady state if the variation in the parameter does not exceed 0.5% over the past 20 minutes.
Anda dapat memilih parameter mana yang harus digunakan perangkat lunak untuk menentukan apakah perlengkapan lampu stabil: parameter optik, parameter listrik.
The software will save the change curves for the selected parameters during the warm-up process in the test report.
Setelah Anda selesai mengkonfigurasi pengaturan, klik OK untuk menyimpan dan keluar.

Moving Detector Goniophotometer-Figure124

Gambar 119

8.1.3 Configure Test Parameters and Start the Test

Click “Start Test.”

Moving Detector Goniophotometer-Figure125

Gambar 120

Buka layar pengaturan parameter pengujian.

Jumlah Note
Lighting fixture types: indoor lights, outdoor lights, or plant grow lights—these three types of lighting fixtures are subject to C-Gamma testing.
Select whether the luminaire is symmetrical. Unless there are specific requirements, select the first option, “asymmetrical,” regardless of whether the luminaire being tested is symmetrical or not (to obtain the luminaire’s actual data).
C-plane angle range: Select 0 to 180 degrees; no modification is needed. For data ranging from 180 to 360 degrees, the software will automatically convert the
C-plane angle intervals. For symmetrical luminaires, such as indoor lights, you can select a larger interval, such as 30 degrees; for asymmetrical luminaires, such as streetlights, you can select a smaller interval, such as 10 degrees. The smaller the angle increment, the higher the testing accuracy, but the longer the testing time. If time permits, we recommend selecting a smaller angle increment.
Gamma plane angle range: If the sample’s light irradiation angle exceeds 180 degrees (e.g., incandescent bulbs), select -180 degrees to 180 degrees; if the sample’s light irradiation angle is less than 180 degrees (e.g., panel lights), you may select a
Gamma plane angle increment of -90 degrees to 90 degrees. Unless there are special requirements, a 1-degree increment is sufficient. Different angle increments do not affect the test speed.
In the vast majority of cases, you should choose to use a far-field photometer. Near-field photometers are intended only for light sources with very low luminous intensity, such as a single LED.
For routine testing of standard lighting fixtures, select “Normal Speed”; if the lighting fixture being tested has an extremely narrow beam angle, select “Slow Speed” to obtain more accurate test results; for expedited testing, select “Fast Speed.”
Fixed Range: If you are consistently measuring the same type of light and the luminous intensity data does not vary significantly, you can check this box to save time. If you have ample time or frequently test lamps with varying luminous intensities, it is recommended that you leave this box unchecked. In this case, the photometric probe will automatically select the appropriate range for each lamp.
Continuous Scan Mode is suitable when the angular interval on the Gamma plane is between -180 degrees and 180 degrees. During the test, when the Gamma angle approaches ±180 degrees, while simultaneously rotating the C-axis to the next scanning position—that is, the Gamma axis rotates continuously without stopping to save time. If you have ample time, we recommend leaving this option unchecked.
To set the monitoring angle during the lamp’s warm-up phase, simply set it to (0, 0).
Normalization refers to the decision of whether to apply a uniform value to the first row of data in the luminous intensity table of a test report; certain standards, such as the U. S. streetlight standard, may require you to perform a similar adjustment. There are three options: no normalization, normalization to the maximum value (which will slightly increase the luminaire’s total luminous flux), and normalization to the average value (which leaves the luminous flux essentially unchanged).
If you need to remove all stray light outside a specific angle, check this option and enter the desired angle
. For example: If you check this option and enter 60, data outside ±60 degrees of the luminaire’s Gamma plane will automatically be set to 0.
Enter the number of light sources in the luminaire and the rated luminous flux of each light source. If you do not know this information, you may leave it blank; it will not affect the test results.
1 Enter the dimensions of the sample’s light-emitting surface. For indoor lighting fixtures, the software requires this data to calculate the luminance limit curve, UGR, and other parameters.
Enter information about the sample luminaire
The software automatically reads the temperature and humidity data from the host’s thermometer and hygrometer.
3 Click OK to go to the warm-up screen, or go to the test screen.

Moving Detector Goniophotometer-Figure126

Gambar 121

Then wait for the software test to finish. Once the test is complete, refer to Section 8.2 to perform the chromaticity distribution test.

8.2 Chromaticity Distribution Test

After completing the C-Gamma luminous intensity distribution test, a chromaticity distribution test can be conducted based on the test report.

Note: The software must automatically calculate the spatial angle range required for the chromaticity distribution test based on the photometric distribution test report. Therefore, if a photometric distribution test is not performed, a chromaticity distribution test cannot be conducted directly.

Select the photometric distribution test report for this luminaire, then click to open the chromaticity distribution test parameter settings interface.

Moving Detector Goniophotometer-Figure127

Gambar 122

Jumlah Note
If checked, only the chromaticity characteristics test will be performed; the chromaticity distribution test
will not be performed. If unchecked, both the chromaticity characteristics test and the chromaticity distribution test will be performed.
Selection of test positions for colorimetric characteristics. The “optical axis” refers to the position directly in front of the light-emitting surface of the sample luminaire; the “point of maximum luminous intensity” refers to the position where the luminous intensity of the sample luminaire is greatest.
The angle range for the C-plane is 0 to 360 degrees. There is no need to adjust
the angle increment for the C-plane. Symmetrical fixtures, such as indoor lights, can use a larger increment—for example, 30 degrees—while asymmetrical fixtures, such as streetlights, can use a smaller increment—for example, 10 degrees. The smaller the increment, the higher the test accuracy, but the longer the test takes.
The gamma plane angle range is automatically calculated by the software based on the results of the photometric distribution test. There is no need to adjust the angle intervals of the
gamma plane. You may also select a larger angle interval; symmetrical luminaires, such as indoor lights, can use a larger interval—for example, 10 degrees—while asymmetrical luminaires, such as streetlights, can use a smaller interval—for example, 5 degrees. The smaller the angle interval, the higher the test accuracy; however, the test will take longer.
Once you’ve finished setting it up, click “Start” to begin the test, and wait for the software to automatically complete the self-test.

Moving Detector Goniophotometer-Figure128

Gambar 123

9. B-Beta Testing

9.1 Luminous Intensity Distribution Test

9.1.1 Installing and Turning On the Lights

9.1.1.1 Pemasangan Perlengkapan Lampu

Find the Square B-Beta Test Fixture.

Moving Detector Goniophotometer-Figure129

Gambar 124

Jumlah Note
Click to switch the viewing angle
Click ① to display the B-Beta angle here instead of the C-Gamma angle.
Click to open the angle control interface

Moving Detector Goniophotometer-Figure130

Gambar 125

Similarly, in the vast majority of cases, you should choose to use a far-field photometer. Select the far-field photometer and set both the B and Beta angles to 0 degrees.

Moving Detector Goniophotometer-Figure131

Gambar 126

Install and secure the B-Beta fixture to the lamp shaft as shown in the figure below, and secure the sample to the fixture. When using a far-field photometric probe, install the lamp so that its light-emitting surface faces the reflector (Note: When using a near-field photometric probe, install the lamp so that its light-emitting surface faces the near-field photometric probe). To ensure the lamp’s light-emitting surface is centered, set the B angle to 90 degrees in the software and align the lamp’s light-emitting surface with the central crosshair laser on the main unit to verify proper positioning.

Moving Detector Goniophotometer-Figure132

Gambar 127

Once the light fixture is installed, you can turn off the cross-line laser.

9.1.1.2 Turning on the Lights

Please refer to 8.1.1.2.

9.1.2 Pengaturan Parameter Pemanasan Awal

Please refer to 8.1.2.

9.1.3 Configure Test Parameters and Start the Test

Click “Start Test” to go to the test parameter settings screen.

Jumlah Note
Please select “Floodlight,” and the test mode will automatically switch to B-Beta.
The angle range for Plane B is -90 degrees to 90 degrees. There is no need to adjust
the angle increments for Plane B; you can select smaller increments, such as 5 degrees or 10 degrees. The smaller the increment, the higher the test accuracy, but the longer it takes.
The Beta plane angle range is -90 degrees to 90 degrees. There is no need to adjust the
Beta plane angle increment; unless there are special requirements, select 1 degree. Different angle increments here have no effect on test speed.

Moving Detector Goniophotometer-Figure133

Gambar 128

For other parameter settings, refer to Section 8.1.3. Once the settings are configured, click “Start” and wait for the software to complete the test. After the test is complete, refer to Section 9.2 to perform a chromaticity distribution test.

9.2 Chromaticity Distribution Test

After completing the B-Beta luminous intensity distribution test, a chromaticity distribution test can be conducted based on the test report.

Select the photometric distribution test report for this luminaire, then click to open the chromaticity distribution test parameter settings screen.

Jumlah Note
The angle range for Plane B is automatically calculated by the software based on the results of the photometric distribution test; there is no need to adjust the angle intervals for Plane
B. Options include 5 degrees and 10 degrees, for example. The smaller the interval, the higher the test accuracy, but the longer it takes.
The beta-plane angle range is automatically calculated by the software based on the results of the photometric distribution test; there is no need to adjust the
beta-plane angle intervals, which can be set to values such as 5 degrees or 10 degrees. The smaller the interval, the higher the test accuracy, but the longer the test takes.

Moving Detector Goniophotometer-Figure134

Gambar 129

For other parameter settings, please refer to Section 8.2. Once you have configured the settings, click “Start” and wait for the software to automatically complete the test.

10. Processing of Test Reports

10.1 Organizing Database Files

Jumlah Note
By default, all test reports are automatically stored in the currently open database file in the order in which the tests were run.
To delete a report, select it and then click “Delete.”
To make it easier to find test reports, click the blank button shown in the image below to create a new database file, and store reports for different lighting fixtures in separate database files. After creating a new database file, you can proceed directly with testing without needing to recalibrate or reconfigure the system.
Click “Open” to view test reports from other database files.

Moving Detector Goniophotometer-Figure135

Gambar 130

Note 1: The next time the software is opened, it will automatically load the database file used last time. If the path to that database file has changed or its name has been modified, the software will prompt you to create a new database file.

Note 2: Avoid placing database files on your computer’s desktop whenever possible.

10.2 Test Report Display Settings

Jumlah Note
You can double-click to upload a picture of the light fixture.
You can double-click the content displayed in this area to edit it.

Moving Detector Goniophotometer-Figure136

Gambar 131

Jumlah Note
Klik "Pengaturan"
For unit selection, “cd” is generally used; however, some older standards—such as those regarding energy-saving light bulbs—may require the use of “cd/klm.”
Streetlight C0 angle setting. The CIE/GB and IESNA standards have inconsistent requirements; please select the standard you are following.
Customize the divergence angle. The software automatically displays the half-peak angle (commonly referred to as the beam angle, calculated as 50% of the maximum luminous intensity) and the effective beam angle (calculated as 10% of the maximum luminous intensity). If you wish to customize the divergence angle, enter the percentage of the maximum light intensity and check the box (e.g., 0.25 for 25%). In this case, the test report will display not only the half-peak angle and effective beam angle but also your custom divergence angle.
Check this box to display the maximum luminous intensity cone angle in the report.

Moving Detector Goniophotometer-Figure137

Gambar 132

Jumlah Note
Switch to the “Energy Efficiency Rating” settings screen, then select the applicable standard and luminaire type.
Click to apply to the current test report
Click OK to save and exit

Moving Detector Goniophotometer-Figure138

Gambar 133

Each display window has several settings that allow you to customize the display. Users can experiment with adjusting these settings to achieve the desired display effects.

Moving Detector Goniophotometer-Figure139

Gambar 134

10.3 Processing Test Data

10.3.1 Double-Sided Light Fixtures

For lamps that emit light from both sides, please use the indoor lighting test mode to test each side individually—that is, set the gamma plane angle range to -90 degrees to 90 degrees during testing—and then generate two test reports.

Jumlah Note
Hold down the “Ctrl” key on your computer keyboard, then click to select the two test reports you just received.
Click “Data Synthesis” in the software navigation bar
In the pop-up window, you can click to toggle between the upward and downward light beams.
Select “Composite”
Click OK to save and exit

Moving Detector Goniophotometer-Figure140

Gambar 135

This will generate a new synthetic test report, as shown in the figure below.

Moving Detector Goniophotometer-Figure141

Gambar 136

10.3.2 Conversion of Test Report Types

Jumlah Note
Click “Type Conversion”
Test reports can be converted among four types
When converting to a C-Gamma test report, you can use the “termination angle” of the C-plane to generate a symmetric test report. If the termination angle is set to 360 degrees, there will be no symmetry; if the termination angle is set to 180 degrees, the report can be converted into an axisymmetric test report; if the termination angle is set to 90 degrees, the report can be converted into a quadrant-symmetric test report; if the termination angle is set to 0 degrees, the report can be converted into a fully symmetric test report. This feature is primarily intended for symmetric indoor lighting fixtures; certain lighting design software programs, such as DIALUX, may require you to use it when importing fully symmetric IES files.

Moving Detector Goniophotometer-Figure142

Gambar 137

10.4 Exporting Test Data and Printing Test Reports

10.4.1 Exporting Test Data

Jumlah Note
Click the red logo in the upper-left corner of the software
Click “Export Data” in the drop-down menu

Moving Detector Goniophotometer-Figure143

Gambar 138

Files can be exported in a variety of standard formats.

Moving Detector Goniophotometer-Figure144

Gambar 139

10.4.2 Printing Test Reports

Print settings.

Jumlah Note
Click to open the “Print Settings” screen
The report displays language options; you can choose between Chinese and English.
Format Nomor Halaman
Report Header Information
Report Titles in Chinese and English
General Report Content, Applicable to Reports for All Types of Lighting Fixtures
Since indoor lights, streetlights, floodlights, and plant grow lights have different key parameters, the report content will vary accordingly. You can select the specific items you want to include in the report.
TM-30 Options
CIE color space diagram or SDCM color tolerance diagram—only one of the two may be selected

Moving Detector Goniophotometer-Figure145

Gambar 140

Jumlah Note
Click to print the currently selected test report
Click to print all test reports in the current database file.

Moving Detector Goniophotometer-Figure146

Gambar 141

11. Pemeliharaan dan Perawatan Peralatan Harian

11.1 Perawatan Rutin

Start-up procedure: First, turn on the main power switch for the cabinet. Next, turn on the power switches for the equipment inside the cabinet and in the darkroom one by one. Finally, launch the software. Shutdown procedure: First, close the software. Next, turn off the power switches for the equipment inside the cabinet and in the darkroom one by one. Finally, turn off the main power switch for the cabinet.

Keep the darkroom control room clean and dust-free to prevent dust from entering the equipment. If there is dust on the surface of the equipment, wipe it off with a dry, soft cloth.

Avoid exposing the darkroom to high humidity for extended periods; keep a dehumidifier on hand if necessary.

11.2 Pemeliharaan Reguler

11.2.1 Standard Lamp

The standard lamp is valid for one year. If you still need to calibrate the darkroom after one year, we recommend purchasing a new standard lamp.

Perangkat Lunak 11.2.2

We update the software from time to time to add features and fix bugs. If you encounter any issues while using the software, please feel free to contact us at any time to obtain the latest version.

11.3 Penonaktifan dan Pemeliharaan Jangka Panjang

11.3.1 Equipment Cleaning and Protection

Thoroughly clean all equipment surfaces, and cover the main unit, photometric probe, Optical Fibre, and cabinet with dust-proof cloths; disconnect all power sources and unplug power cords and communication cables.

11.3.2 Perlindungan Lingkungan

Ensure that the darkroom is dry and well-ventilated to prevent equipment from becoming damp. If the ambient humidity is high, keep a dehumidifier on hand.

11.3.3 Monthly Power-On

Lakukan uji penyalaan daya sebulan sekali untuk memastikan semua perangkat keras sistem berfungsi dengan baik.

12. Pemecahan Masalah dan Penyelesaian Masalah Umum

12.1 Kesalahan Komunikasi

Cause of the problem: Check that the communication cable is securely connected and that it is plugged into the correct port; verify that the drivers are installed correctly; try a different USB port or communication cable to rule out hardware issues.

Solution: Reinstall the driver, connect the communication cable according to the device identifier, restart the software, and try again.

12.2 Kesalahan Pengujian

12.2.1 Luminous Flux Anomalies

Cause of the malfunction: Check whether the light source is stable and whether the warm-up time is sufficient; verify that the equipment calibration is valid, that the test distance is correct, and that the standard light is functioning properly; check whether there is any external light interference in the darkroom and whether the light path is obstructed.

Solution: Extend the warm-up time, recalibrate the instrument, turn off other light sources in the darkroom, and remove any obstructions from the light path.

12.2.2 UGR Anomalies

Cause of error: Dimensions of the luminaire’s light-emitting surface were not entered; the luminaire’s beam angle is <45°; the calculation standard was selected incorrectly.

Solution: Enter the dimensions of the luminaire’s light-emitting surface (length/width/height) correctly; verify that the luminaire is an indoor luminaire (the beam angle must be sufficient); select the CIE 190:2010 standard and recalculate.

12.2.3 Abnormal Chromaticity Distribution Test Results

Cause of the malfunction: The test was not conducted based on the photometric distribution test report for the current luminaire; the light intensity was too low, preventing the Spectroradiometer from accurately measuring the colorimetric parameters.

Procedure: Please first conduct a luminous intensity distribution test, and then perform a color distribution test based on that report; if the luminaire’s luminous intensity is too low, please contact us for the best solution.

12.3 The Goniophotomoeter Does Not Rotate

Cause of the malfunction: Check whether the host power is on and whether the wiring is correct; check whether the angle controller inside the cabinet is functioning properly; confirm whether any foreign objects are jammed in the rotating shaft.

Solution: Check the power supply and wiring, and remove any foreign objects; restart the device and software, then try again.

12.4 Software Failures

12.4.1 Kerusakan Perangkat Lunak

Cause of the problem: Check whether you are logged in to the computer’s operating system with an administrator account; verify that the software is fully installed; check whether the database files are corrupted.

Solution: Log in to the computer using an administrator account; reinstall the software and drivers; create a new database file.

12.4.2 The software displays the message “File is not a database”

Cause of the problem: The database file was deleted or moved; the default program used to open LGF files was changed.

Solution: Create a new database file and save it; restore the default program for opening LGF files.

12.5 Kegagalan Daya

12.5.1 DC Power Output Abnormalities

Causes of the malfunction: The AC/DC switch on the cabinet was not set to DC; the sample voltage or current exceeded the DC power supply limits; a short circuit in the sample;

Troubleshooting: Please verify that the sample’s voltage and current are within the DC power supply’s output range. For example, the DC3005 Model ini memiliki tegangan keluaran maksimum 30 V dan arus keluaran maksimum 5 A. Periksa apakah sampel mengalami korsleting; jika ya, perbaiki sampel sebelum melakukan pengujian ulang.

12.5.2 No AC Output

Cause of the malfunction: The AC/DC switch in the cabinet was not set to the AC side; the overload protection was triggered;

Troubleshooting Steps: Check the AC power output wiring and sample connections to ensure there are no loose connections or short circuits; if the overload protection is triggered, turn off the power, let the unit cool for 5 minutes, and then restart it. Ensure that the sample power remains within the AC power output range; for example, the LSP-500VARC has a maximum output power of 500W.

Lampiran & Unduhan

Video

LSG-6000 LM-79-19 Detektor Bergerak Goniophotometer - Video Operasi

LSG-5000 LM-79-09 Detektor Bergerak Goniophotometer - Video Operasi

LSG-3000 LM-79 Detektor Bergerak Goniophotometer - Video Operasi