+8618117273997weixin
InggrisInggris
中文简体 中文简体 en English ru Русский es Español pt Português tr Türkçe ar العربية de Deutsch pl Polski it Italiano fr Français ko 한국어 th ไทย vi Tiếng Việt ja 日本語

Rotasi Presisi Tinggi Luminer Goniophotometer

Pengguna manual

berlaku LISUN Model: LSG-1890B

1. Membuka Kemasan dan Mengenal Peralatan Secara Dasar

1.1 Membuka Kemasan dan Pemeriksaan

When unpacking the device, handle it gently to avoid scratching the casing with sharp objects. After unboxing, first inspect the exterior of the device to ensure there is no visible deformation or dents, and that the buttons are not damaged or loose. If you notice any abnormalities with the device or its accessories, do not turn it on. Please contact us immediately.

1.2 Konfigurasi dan Gambaran Umum Sistem

1.2.1 Komponen Perangkat Keras Inti

Catatan: Tabel di bawah ini mencantumkan komponen perangkat keras inti dari keseluruhan sistem. Peralatan yang Anda beli mungkin hanya mencakup beberapa komponen ini; harap perhatikan hal ini saat mengikuti petunjuk dalam manual ini.

Nama komponen Pilih Model Fitur Inti
Unit Utama Goniophotometer LSG-1890B Used in conjunction with a photometric probe to measure photometric parameters and their spatial distribution, such as luminous flux and light distribution curves.
Probe fotometri medan dekat PM400C (Utara) Suitable for standard darkrooms, such as those around 10 meters in size
Far-Field Photometric Probe PM400C (F) Optional. Suitable for extra-long darkrooms, such as those longer than 20 meters.
Lampu Standar SLS-150W Used to calibrate the darkroom (luminance sensor)
E27 Fixture / Untuk pengujian C-Gamma; cocok untuk lampu pijar.
Penjepit Silang Untuk pengujian C-Gamma; cocok untuk perlengkapan pencahayaan dalam dan luar ruangan lainnya.
Penjepit Persegi Untuk pengujian B-Beta; cocok untuk lampu sorot
Laser Garis Silang Digunakan selama instalasi dan pengoperasian awal; digunakan selama kalibrasi dan pengujian untuk menentukan posisi dan orientasi lampu standar dan luminer sampel.
Kabinet CASE-19 Used to house angle controllers, power meters, and DC and AC power supplies. Includes a power strip, an RS485 communication module, and more.
Pengontrol Sudut RT-200C Sudut di mana Goniophotometer diposisikan untuk tampilan dan kontrol. Umumnya, tidak diperlukan penyesuaian manual; kontrol ditangani oleh perangkat lunak.
Pengukur daya Seri LS Measure electrical parameters such as voltage, current, power, and power factor of the sample
DC Power Supply Seri DC Menyediakan catu daya DC yang stabil untuk sampel DC, lampu standar, dan lampu bantu.
AC Power LSP Series Provides a stable, pure sine-wave AC power supply for AC samples

1.2.2 Perangkat lunak, manual, sertifikat, dll.

At the time of shipment, a download link containing all the documents was sent via email. Please download them promptly. If you missed the email, you can contact us to request a new download link.

Includes software, user manuals, warranty cards, standard lamp calibration certificates, and other important documents, as well as other relevant documents.

1.4 Parameter Pengukuran

Illuminance data, illuminance distribution, area luminous flux, luminaire efficiency, luminance distribution (optional), utilization factor, luminance limit curve, glare rating, isophote curves, maximum allowable height-to-distance ratio, luminaire curve vs. illumination area, isointensity curves, effective beam angle, EEI, UGR, PAR, PPF, and PPFD spatial distribution tests, etc.

2. Catatan Penting

2.1 Keamanan Listrik

The system must be connected to a stable AC power source rated at 220 V ±5% with a frequency of 50 Hz or 60 Hz. Ensure proper grounding with a grounding resistance of ≤4 Ω. Do not use the equipment if the power connections are loose or the cables are damaged.

2.2 Keselamatan Lingkungan

When using the optical darkroom, the ambient temperature must be maintained at 25±1°C, and the relative humidity must be maintained between 10% and 65% RH;

Both the control room and the darkroom must be kept dry, well-ventilated, dust-free, and free of corrosive gases;

Avoid placing the device near strong magnetic or electric fields; it must be placed on a level, sturdy surface.

2.3 Prosedur Operasi Ruang Gelap

2.3.1 Deskripsi Ruang Gelap

Please confirm that the darkroom has been constructed exactly according to the darkroom drawings we provided. If there are any discrepancies in the actual dimensions, please inform us before installing and commissioning the equipment so that we can adjust the plan, ensuring that the installation and commissioning proceed smoothly and that the test results are accurate;

The interior of the darkroom must be completely painted matte black—that is, a non-reflective pure black—including all walls, ceilings, floors, partitions, doors, and so on. A small number of components, such as lighting fixtures, air conditioning units, and power switches, may not be painted black, provided they do not appear in the optical path between the Goniophotomoeter’s main unit and the photometric probe;

Once the equipment has been installed and commissioned, do not move any of the hardware components inside the darkroom, including the main unit base, the photometer probe base, the photometer probe orientation, and the cross-laser base. Doing so may result in inaccurate test results.

2.3.2 Precautions for Using the Darkroom

Lepaskan sepatu Anda atau kenakan penutup sepatu saat memasuki ruang gelap untuk mencegah debu masuk;

Luminer sampel harus dipasang dengan aman untuk mencegahnya jatuh dan merusak Goniophotometer selama pengujian;

Do not place any objects on the main unit of the Goniophotomoeter, and do not place any objects within the Goniophotomoeter’s rotation range;

During testing, the darkroom must remain sealed to prevent external light from interfering with the test results.

2.3.3 Precautions When the Darkroom Is Not in Use

Keep the darkroom closed; unauthorized persons are not allowed to enter; prevent dust from entering the darkroom.

2.4 Prosedur Pengoperasian Lampu Standar

The standard lamp has a lifespan of one year and is used for calibrating the photometer probe in a darkroom during installation and commissioning;

Jika laboratorium Anda diharuskan untuk mengkalibrasi ruang gelapnya setiap tahun, Anda dapat membeli lampu standar baru;

Lampu standar akan terasa panas segera setelah digunakan; jangan disentuh. Biarkan dingin selama 10 menit sebelum mengembalikannya ke wadah lampu;

The standard lamp used in the darkroom is a directional standard lamp, and its direction is controlled by its bracket. Do not remove the bulb from its bracket; doing so will render the standard lamp inoperable.

2.5 Lainnya

Untuk hal-hal yang tidak tercakup dalam panduan ini, harap berhati-hati atau hubungi kami.

3. Perakitan Perangkat Keras

3.1 Cabinet Equipment Assembly

3.1.1 Introduction to the Cabinet

Place the cabinet on a level, sturdy workbench, leaving sufficient space for operation.

Jumlah Note
Wiring Diagram: Please assemble the equipment according to the wiring diagram.
Panel distribusi daya, digunakan untuk menghubungkan kabel daya semua perangkat di dalam kabinet.
Cooling fan power cord: Please connect it to the power strip inside the cabinet.
Kotak Komunikasi RS-485, digunakan untuk menghubungkan semua perangkat melalui kabel komunikasi RS-485.
Harap sambungkan adaptor daya untuk kotak komunikasi RS485 ke stopkontak di dalam kabinet.
The USB cable for the RS485 communication box must be connected to the computer.
Pengkabelan Pengontrol Sudut RT-200C
LS Series Power Meter Wiring
Koneksi Daya DC
Pengkabelan Daya AC
The main power cable for the cabinet, which supplies power to all equipment inside the cabinet, should be routed out through cable opening ② and connected to the power source only after all equipment has been fully assembled.
1 Cable opening—the main power cord can be routed through here. Note: The main power cord plug is a Chinese standard plug; you may need to provide your own power adapter.

High Precision Rotation Luminaire Goniophotometer-Figure1

Gambar 1

3.1.2 Instalasi RT-200C

Untuk saat ini, yang perlu Anda lakukan hanyalah mencolokkannya.

High Precision Rotation Luminaire Goniophotometer-Figure2

Gambar 2

3.1.3 Installation of the LS Power Meter

Jumlah Note
Connected via an RS-485 communication cable; the other end is connected to the RS-485 communication box inside the cabinet.
Kabel daya sudah terhubung; ujung lainnya sudah terhubung ke panel distribusi daya di dalam kabinet.
V1 test lead, connected to the red terminal of the power meter’s voltage V
Kabel uji V2, dihubungkan ke terminal hitam tegangan V pada meteran daya.
A1 test lead, connected to the red terminal of the power meter’s current A
A2 test lead, connected to the black terminal for current A on the power meter

High Precision Rotation Luminaire Goniophotometer-Figure3

Gambar 3

3.1.4 Instalasi Catu Daya DC

Jumlah Note
Terhubung melalui kabel komunikasi RS-485; ujung lainnya sudah terhubung ke kotak komunikasi RS-485 di dalam kabinet.
Hubungkan kabel keluaran daya DC, pastikan untuk mencocokkan terminal positif dan negatif.
The power cord is connected; the other end is already connected to the power strip inside the cabinet.

High Precision Rotation Luminaire Goniophotometer-Figure4

Gambar 4

3.1.5 Instalasi Daya AC

Please locate the following AC power output cable in the accessories:

High Precision Rotation Luminaire Goniophotometer-Figure5

Gambar 5

Jumlah Note
Connected via an RS485 communication cable; the other end is already connected to the RS485 communication box inside the cabinet.
The power cord is connected; the other end is already connected to the power strip inside the cabinet.
Lewatkan kabel keluaran daya AC melalui lubang kabel ④ dan sambungkan ke posisi ⑤.
Lubang kawat
AC power output cable: Pass it through cable hole ④ and connect it to position ③.

High Precision Rotation Luminaire Goniophotometer-Figure6

Gambar 6

High Precision Rotation Luminaire Goniophotometer-Figure7

Gambar 7

3.2 Pemasangan Unit Utama Goniophotometer dan Probe Goniophotometer

3.2.1 Host Assembly

The same number is engraved on the main unit’s base tapered sleeve ① and the corresponding position on the main unit ②. Rotate the tapered sleeve until the two numbers are aligned, then secure the tapered sleeve using the hex socket screws provided as standard.

High Precision Rotation Luminaire Goniophotometer-Figure8

Gambar 8

High Precision Rotation Luminaire Goniophotometer-Figure9

Gambar 9

Remove the screws securing the dust cover on the turntable, and pull the dust cover off.

High Precision Rotation Luminaire Goniophotometer-Figure10

Gambar 10

Use a forklift to lift the turntable; since the turntable is quite heavy, three to four people can also carefully lift it. Align the dust cover on the turntable with the terminal block on the base, then slowly lower the turntable onto the base. Hold the terminal block in place with your hand so that it slides into the turntable along with the tapered sleeve, preventing the terminal block from being crushed by the turntable.

High Precision Rotation Luminaire Goniophotometer-Figure11

Gambar 11

To fine-tune the turntable’s direction, follow these steps in the order listed below.

Jumlah Note
Align the six screw holes on the turntable with the six screw holes on the base, and use the standard hex screws and washers provided to secure the turntable to the base
. Note: Be sure to use all the standard washers; otherwise, the screws may not fit properly, which could damage the equipment during turntable rotation.
Plug the base terminal block into the corresponding terminal block socket on the turntable.
Then put the dust cover back in place

High Precision Rotation Luminaire Goniophotometer-Figure12

Gambar 12

High Precision Rotation Luminaire Goniophotometer-Figure13

Gambar 13

The four feet on the bottom of the unit are height-adjustable.

Jumlah Note
Turn the screw at the bottom to raise the leg until it is off the ground.
At the same time, make sure the top screw isn’t pressing against anything; you can unscrew it slightly.

High Precision Rotation Luminaire Goniophotometer-Figure14

Gambar 14

Position the goniophotomoeter in the correct location according to the darkroom blueprints. Align the four sides of the base as closely as possible with the darkroom walls. Then remove the auxiliary arm.

High Precision Rotation Luminaire Goniophotometer-Figure15

Gambar 15

Next, please turn all four legs of the Goniophotomoeter until they are in contact with the floor.

3.2.2 Assembly of the Photometric Probe and Probe Holder

Take, for example, a darkroom with only one photometer.

Rakitlah probe fotometri dan dudukan probe, lalu posisikan keduanya di lokasi yang sesuai menurut gambar ruang gelap; untuk saat ini tidak perlu mengencangkannya dengan sekrup.

High Precision Rotation Luminaire Goniophotometer-Figure16

Gambar 16

High Precision Rotation Luminaire Goniophotometer-Figure17

Gambar 17

3.2.3 Pengkabelan Sistem

Jumlah Note
The four terminal blocks ①on the cabinet are connected to the Goniophotomoeter⑩; simply match them one-to-one based on the numbers.
Kedua terminal tersebut terhubung ke blok terminal di dalam kabinet dan digunakan untuk memasok daya ke semua peralatan di ruang gelap, termasuk Goniophotometer dan probe fotometrik. Hal ini memungkinkan kontrol yang mudah terhadap semua peralatan ruang gelap melalui sakelar daya kabinet.
Catatan: Peralatan kamar gelap juga dapat dioperasikan langsung dari listrik utama.
Cabinet grounding terminal: Please connect it to ground using a wire.
Kabel daya kabinet, terhubung ke sumber daya 220V, 50/60 Hz.
Power Cord for the Main Unit of the Goniophotomoeter
Hubungkan terminal ground Goniophotometer ke ground menggunakan kawat.
Ini adalah kabel komunikasi satu-ke-dua; satu ujung kabel RS-485 terhubung ke kotak komunikasi RS-485 di dalam kabinet, dan ujung lainnya terhubung ke posisi ⑭ di bagian belakang RT-200C.
Photometer probe trigger wire; connect to probe ⑬. When only one probe is installed, the default setting is near-field (N).
If you have only one photometer in your darkroom, there is no need to connect the far-field (F) sensor.
Keempat blok terminal ⑩ pada Goniophotometer terhubung ke kabinet ①; cukup pasangkan satu per satu sesuai dengan nomornya.
Adaptor Daya untuk Probe Fotometri
1 Kabel komunikasi RS-485 untuk probe fotometri; ujung lainnya terhubung ke kotak komunikasi RS-485 di dalam kabinet.
Kabel pemicu probe fotometri, terhubung ke unit utama ⑧ dari Goniophotometer
Kabel komunikasi RT-200C, dengan ujung lainnya terhubung ke unit utama ⑦ dari Goniophotometer

High Precision Rotation Luminaire Goniophotometer-Figure18

Gambar 18

High Precision Rotation Luminaire Goniophotometer-Figure19

Gambar 19

High Precision Rotation Luminaire Goniophotometer-Figure20

Gambar 20

High Precision Rotation Luminaire Goniophotometer-Figure21

Gambar 21

4. Instalasi dan Konfigurasi Perangkat Lunak

4.1 Instalasi Perangkat Lunak dan Driver

4.1.1 Lingkungan Eksekusi Perangkat Lunak

Persyaratan Sistem: Windows 7/8/10/11 (32-bit/64-bit); komputer harus memiliki setidaknya satu port USB.

Instalasi 4.1.2

Jumlah Note
LSG Series Goniophotomoeter Software—Double-click to install
File APK ini dapat disalin ke ponsel Android untuk diinstal. Aplikasi ini hanya ditujukan untuk memutar Goniophotomoeter dan tidak dapat digunakan untuk pengujian. Karena perangkat lunak ini mencakup semua fitur aplikasi, manual ini menggunakan perangkat lunak sebagai contoh pengoperasian.
RS485 Communication Box Driver—Please double-click to install

High Precision Rotation Luminaire Goniophotometer-Figure22

Gambar 27

4.1.3 Penanganan Pengecualian

If your antivirus software or the Windows Firewall flags a file as suspicious during the software installation process, please temporarily disable the antivirus software or the Windows Firewall and try installing the software again.

If the software fails to open when you double-click its icon after a successful installation, please check whether you are logged in to your computer with an administrator account. If the software still does not open after logging in with an administrator account, right-click the software icon to open its Properties window. On the Compatibility tab, check the box next to “Run this program as an administrator,” click “Apply,” and then double-click the icon again to open the software.

High Precision Rotation Luminaire Goniophotometer-Figure23

Gambar 28

4.2 Pengaturan Perangkat Lunak

First, turn on the cabinet’s power switch, then power on all system devices, including all equipment inside the cabinet, the Goniophotomoeter main unit, and the photometric probes. Double-click to launch the software. The first time you run the software, it will automatically open the system configuration interface. Select the correct model for each device you have purchased.

Jumlah Note
To select a goniophotomoeter model, click the drop-down list on the right. The model you purchased is the LSG-1890B.
Probe fotometer medan dekat; pilih model probe fotometer PM400C
Far-field photometric probe—if you have not purchased one, please select “Not Installed.”
The LSG-1890B does not include a near-field Spectroradiometer; please select “Not Installed.”
AC power supply models are generally from the LSP Series
Model catu daya DC umumnya disebut sebagai Seri DC.
Model meteran daya: Nomor model tercetak di panel depan meteran daya. Silakan periksa dan pilih model yang sesuai.
Once you have selected all the correct models, click “Auto-detect Port.”
Setelah pencarian semua port selesai, klik “OK” untuk menyimpan dan keluar.

High Precision Rotation Luminaire Goniophotometer-Figure24

Gambar 29

Note: If a communication error occurs during the automatic port search, please read the error message carefully, check whether the corresponding device is powered on, and verify that the RS485 communication cable is properly connected. Once you have confirmed these points, click “Automatic Port Search” again until communication is established with all devices.

After clicking OK, the software automatically opens the database file save dialog. From then on, all test reports will be automatically saved in this database file.

Jumlah Note
Please select the path where you want to save the database file. Please do not save it to the desktop or inside the Desktop folder.
Masukkan nama file
Just click “Save”

High Precision Rotation Luminaire Goniophotometer-Figure25

Gambar 30

5. Calibration of darkroom equipment (Each step must be performed carefully and thoroughly; otherwise, calibration data may be inaccurate and test results may be abnormal.)

5.1 Horizontal Commissioning of the Main Unit

5.1.1 Laser Ready

There is a laser switch on the back of the goniophotomoeter’s main arm. When you turn on the line laser switch, you will see a red line laser beam emitted directly in front of the main arm.

High Precision Rotation Luminaire Goniophotometer-Figure26

Gambar 31

Find the crosshair laser device and tripod.

High Precision Rotation Luminaire Goniophotometer-Figure27

Gambar 32

Remove the cross-line laser and the batteries. Insert the batteries into the cross-line laser. The power switch is located on the side of the device. The device turns on when the switch is set to the “unlocked” position.

Jumlah Note
Power switch: Turn it to the “Unlock” position to turn on the power.
VH refers to vertical and horizontal laser switches, respectively.

High Precision Rotation Luminaire Goniophotometer-Figure28

Gambar 33

Mount the crosshair laser onto the tripod.

Note: The cross-line laser is self-leveling. If the cross-line laser is not self-leveled, an audible alarm will sound. This may be due to an uneven surface or a tripod that is not properly set up. Please ensure that the cross-line laser remains self-leveled at all times during use.

5.1.2 Steps for Horizontal Commissioning of the Host Machine

According to the darkroom drawings, the center height of the LSG-1890B goniophotomoeter is 1510 mm. Using a tripod, adjust the height of the horizontal laser beam from the crosshair laser to 1510 mm. At this height, the horizontal laser beam should pass through the center of all light-transmitting holes in the darkroom partition.

High Precision Rotation Luminaire Goniophotometer-Figure29

Gambar 34

Please note: Do not place any objects on the Goniophotomoeter or within its rotation range to prevent damage to the equipment during rotation.

Buka perangkat lunak.

Jumlah Note
Click to switch angle modes
Click ① to display the C-Gamma angle here instead of the B-Beta angle.
Kemudian klik untuk membuka antarmuka kontrol sudut.

High Precision Rotation Luminaire Goniophotometer-Figure30

Gambar 35

Click the continuous rotation button for any direction. The Gamma axis of the Goniophotomoeter will then begin to rotate continuously.

High Precision Rotation Luminaire Goniophotometer-Figure31

Gambar 36

At this point, observe the horizontal laser beams emitted by the single-line and cross-line lasers from the main axis of the Goniophotomoeter.

High Precision Rotation Luminaire Goniophotometer-Figure32

Gambar 37

Then, using an appropriate wrench, raise the goniophotomoeter and level it by adjusting the height of its four legs. The final result of the calibration: Regardless of the Gamma angle of the goniophotomoeter, its straight laser beam perfectly aligns with the horizontal laser beam, as shown in the figure below.

High Precision Rotation Luminaire Goniophotometer-Figure33

Gambar 38

You’ll need plenty of patience to complete this step. Once adjusted, make sure all four legs of the main unit are firmly on the ground. Then click the “Stop” button on the software’s angle control interface to stop the Goniophotomoeter from rotating. Next, lock the four legs of the Goniophotomoeter in place using the screws at the top. Once the position and level of the Goniophotomoeter main unit have been properly adjusted, do not move it again, as this will cause deviations in the test results.

5.2 Kalibrasi Probe Fotometri

5.2.1 Calibration of the Photometric Probe at the 0-Degree Position

Jumlah Note
Masukkan 0
Click any rotation direction to rotate the Gamma axis back to the 0-degree position.

High Precision Rotation Luminaire Goniophotometer-Figure34

Gambar 39

Verify that the single-line laser beam is passing through the exact horizontal center of the last partition at this point.

High Precision Rotation Luminaire Goniophotometer-Figure35

Gambar 40

If not, follow the steps below.

Jumlah Note
Please enter a small angle value here based on the actual situation, such as 1, 0.1, 0.05, etc.
Select a rotation direction and rotate the Gamma axis until the single-line laser passes through the exact center of the last partition horizontally.
Click “Set to 0 Degrees” to set the current position of the Gamma axis on the near-field luminance probe to 0 degrees. The password is LISUNGROUP.

High Precision Rotation Luminaire Goniophotometer-Figure36

Gambar 40.1

Note 1: After setting the Gamma axis of the near-field photometer to 0 degrees, do not arbitrarily set it to 0 degrees again, as this will cause deviations in the test results.

Note 2: Definition of the Gamma angle for a goniophotomoeter. The direction directly facing the photometer probe is defined as 0 degrees/360 degrees. When rotating clockwise, the angle value increases from 0 degrees to 360 degrees; when rotating counterclockwise, the angle value decreases from 360 degrees to 0 degrees. Since angle notation may vary, -90 degrees and 270 degrees correspond to the same position.

5.2.2 Mounting the Photometric Probe

Rotate the light sensor’s lens cover counterclockwise and remove it.

High Precision Rotation Luminaire Goniophotometer-Figure37

Gambar 41

Kotak kemasan lampu standar berisi cermin kecil.

High Precision Rotation Luminaire Goniophotometer-Figure38

Gambar 42

High Precision Rotation Luminaire Goniophotometer-Figure39

Gambar 43

Remove the small reflector and attach it to the probe.

High Precision Rotation Luminaire Goniophotometer-Figure40

Gambar 44

At this point, the Gamma axis of the near-field photometer is still at the 0-degree position. Slightly move the probe base and adjust the screws at points ① and ② to adjust the probe’s height and orientation. Position the probe so that the straight laser beam coming from the main axis hits the exact center of the mirror, and ensure that the reflected beam travels back along the same path.

High Precision Rotation Luminaire Goniophotometer-Figure41

Gambar 45

Kemudian, kencangkan alas braket probe ke lantai dengan sekrup.

High Precision Rotation Luminaire Goniophotometer-Figure42

Gambar 46

Once secured, you may need to make slight adjustments to the probe’s height and orientation so that the straight laser beam continues to hit the exact center of the reflector and the reflected beam returns along the same path. After calibration is complete, ensure that all screws are tightened. Then remove the reflector and reinstall the light probe’s light shield.

Note: Once the probe’s position and orientation have been set, do not adjust them, as doing so may cause the test data to be inaccurate.

You can secure the photometer probe’s power adapter to the probe mount as shown in the figure below to protect the probe’s power jack.

High Precision Rotation Luminaire Goniophotometer-Figure43

Gambar 47

5.3 Cross-Line Laser Installation

Finally, the crosshair laser must be mounted on the side wall of the darkroom. Depending on the specific layout of each darkroom, the crosshair laser can be mounted on either the left or right wall. Take the darkroom layout shown in the figure below as an example.

High Precision Rotation Luminaire Goniophotometer-Figure44

Gambar 52

Go to the angle control screen.

Jumlah Note
Enter 90 (if the installation position of the crosshair laser in your darkroom blueprint is different, enter -90)
Click to rotate the gamma axis to a 90-degree position, so that the main-axis laser is directly aligned with the position where the crosshair laser is installed.

High Precision Rotation Luminaire Goniophotometer-Figure45

Gambar 53

Secure the crosshair laser to the triangular mount we provide.

High Precision Rotation Luminaire Goniophotometer-Figure46

Gambar 54

Turn on the crosshair laser and use the direction of the single-line laser to roughly determine the position of the crosshair laser.

High Precision Rotation Luminaire Goniophotometer-Figure47

Gambar 55

Fine-Tune the Position of the Crosshair Laser: The centerline of the straight laser should be perfectly aligned with the centerline of the crosshair laser. You can verify and adjust the position using the following method. Place a sheet of white paper at the exit of the cross-shaped laser and at the exit of the single-line laser. If the centerline of the single-line laser aligns with the centerline of the cross-shaped laser at both locations, you can confirm that the position and orientation of the cross-shaped laser are correct.

High Precision Rotation Luminaire Goniophotometer-Figure48

Gambar 56

High Precision Rotation Luminaire Goniophotometer-Figure49

Gambar 57

Next, secure the cross-laser tripod to the wall using screws. Once secured, you can still loosen the screws if necessary to make slight adjustments to the cross-laser in both the horizontal and vertical directions, ensuring it remains properly aligned. After calibration is complete, all screws should be tightened securely.

High Precision Rotation Luminaire Goniophotometer-Figure50

Gambar 58

Then rotate the Gamma axis back to the 0-degree position. The cross-shaped laser beam is projected onto the opposite wall. You can use this to create a permanent mark. That is, when the cross-shaped laser beam is projected in this direction, it indicates the horizontal and vertical center positions of the Goniophotomoeter. This mark is needed for both calibration and testing.

High Precision Rotation Luminaire Goniophotometer-Figure51

Gambar 59

Installation and commissioning are complete. Turn off the line laser on the main shaft of the Goniophotomoeter.

5.4 Membersihkan

You may shut down all equipment. When doing so, first turn off the power switch on each individual device, and then turn off the main power switch for the cabinet. Next, thoroughly clean the darkroom and control room. To prevent dust from entering the equipment, cover the devices with dust-proof cloths before cleaning. After cleaning, turn all equipment back on and continue with the calibration and testing procedures.

Catatan: Jaga agar ruang gelap sebisa mungkin bebas debu selama penggunaan.

6. Kalibrasi

6.1 Pemasangan dan Pengoperasian Lampu Standar

6.1.1 Pemasangan Lampu Standar

First, turn on the crosshair laser on the wall and rotate it to ensure that it is aimed at the mark on the opposite wall.

Open the Standard Lamp box.

High Precision Rotation Luminaire Goniophotometer-Figure52

Gambar 60

Then assemble the standard lamp bracket as shown in the figure below.

High Precision Rotation Luminaire Goniophotometer-Figure53

Gambar 61

The main shaft of the Goniophotomoeter should still be aligned with the photometric probe, i.e., the probe’s 0-degree position. Locate the calibration rod and install it at the center of the Goniophotomoeter. Then remove the top section.

High Precision Rotation Luminaire Goniophotometer-Figure54

Gambar 62

Please follow the instructions below to connect the wires.

Jumlah Note
The main shaft of the goniophotomoeter has four terminals for connecting the power cables and sampling cables of the standard lamp and sample luminaire. Use short wires to bridge the red and yellow terminals together, and the black and blue terminals together.
Connect the power cord of the standard lamp/sample fixture to the red and blue terminals on the main shaft.

High Precision Rotation Luminaire Goniophotometer-Figure55

Gambar 63

6.1.2 Calibration of the Standard Light Direction

Jumlah Note
Loosen the screw here to adjust the height of the standard lamp.
There are three screws here that allow you to adjust the pitch of the standard lamp.

High Precision Rotation Luminaire Goniophotometer-Figure56

Gambar 64

Lampu standar harus disesuaikan sehingga berkas laser berbentuk salib diarahkan ke tengah cermin kecil, dan berkas laser horizontal dan vertikal yang dipantulkan dari cermin bertepatan persis dengan berkas laser berbentuk salib aslinya.

High Precision Rotation Luminaire Goniophotometer-Figure57

Gambar 65

Based on our experience, follow these steps for calibration: First, adjust the height of the standard lamp to the appropriate position. The vertical reflected laser can be calibrated by rotating the calibration target, while the horizontal reflected laser can be calibrated by adjusting the pitch of the standard lamp. After calibration, the center height of the standard lamp may change, requiring you to readjust its height. Continue the reflection adjustment until the cross-shaped laser beam is projected onto the center of the small mirror, and the reflected horizontal and vertical laser beams perfectly align with the original cross-shaped laser beam.

Once the standard lamp is properly aligned, carefully remove the small reflector, taking care not to alter the alignment of the standard lamp during removal. Then turn off the cross-line laser.

6.2 Kalibrasi Probe Fotometri

All standard lamps are constant-current DC light sources. Set the cabinet’s AC/DC switch to DC.

High Precision Rotation Luminaire Goniophotometer-Figure58

Gambar 66

Buka pengontrol sudut perangkat lunak.

Jumlah Note
Enter -90
Click to rotate the Gamma axis to the -90-degree position (if the crosshair laser is installed in a different position on your darkroom drawing, you will need to rotate it to the 90-degree position).

High Precision Rotation Luminaire Goniophotometer-Figure59

Gambar 66.1

Align the round hole in the standard lamp mount so that it faces the photometric probe.

High Precision Rotation Luminaire Goniophotometer-Figure60

Gambar 67

Turn on the calibration lamp and enter the calibration interface.

Jumlah Note
Select a DC Power Supply
The standard lamp is a constant-current light source. Select CC (constant-current mode) and enter the reference voltage and rated current as specified in the standard lamp certificate; typically, these are 30 V and 6.1 A.
Click to turn on the standard lamp and turn off all other lights in the darkroom. The standard lamp takes 15 minutes to reach a steady state.
Setelah lampu standar stabil, klik untuk masuk ke antarmuka kalibrasi fotometer.

High Precision Rotation Luminaire Goniophotometer-Figure61

Gambar 68

Close the darkroom door and make sure that all light sources in the darkroom are turned off, except for the standard lamp.

Jumlah Note
Memilih Probe Fotometri Medan Dekat
Luminous flux calibration—our software includes this option, but current standard practices require the use of luminous intensity calibration, and the standard lamps included in our standard configuration are also luminous intensity standard lamps. Therefore, there is no need to use luminous flux calibration at this time.
Metode Kalibrasi Intensitas Cahaya
Rujuklah pada sertifikat kalibrasi lampu standar dan masukkan nilai intensitas cahaya yang telah dikalibrasi.
Klik untuk Memulai
Perangkat lunak akan secara otomatis menghitung jarak uji.
To verify the results, click “Check,” and the software will display the current measured light intensity value. The measured light intensity value should match the calibrated light intensity value.
Klik “OK” untuk menyimpan dan keluar.

High Precision Rotation Luminaire Goniophotometer-Figure62

Gambar 69

Note: The test distance is the distance from the center of the Goniophotomoeter to the surface of the photometric probe. The distance calculated by the software should be close to the actual distance; an error of 1% or less is considered normal. If the error is too large, please refer to Section 2.3.1 to verify that the darkroom is functioning properly, and refer to Chapter 5 to review the calibration steps.

After calibration is complete, you can turn off the standard lamp using the software. Since the standard lamp becomes quite hot during use, wait for it to cool down before placing it back in its packaging for safe storage. Remove the calibration rod and store it safely.

6.3 Kalibrasi dan Pemeliharaan

Tidak ada perubahan di ruang gelap. If there have been no changes to the computer, recalibration is not necessary.
No changes were made to the computer; simply unplug and replug the communication cable or reinstall the Goniophotomoeter software. No recalibration is necessary.
Jika Anda beralih ke komputer baru atau menginstal ulang sistem operasi, tidak perlu melakukan kalibrasi ulang; cukup masukkan jarak uji sebelumnya dan simpan.
Perubahan di Ruang Gelap The structure of the darkroom remains unchanged; however, after prolonged use, the interior paint may fade or peel. Depending on how well you maintain the darkroom and the level of accuracy you require for your test results, you may repaint the darkroom periodically. After repainting, the darkroom must be recalibrated.
Terdapat perubahan pada konfigurasi ruang gelap, sehingga diperlukan kalibrasi ulang.

7. Persiapan Sebelum Pengujian

7.1 Pemilihan dan Penuaan Lampu

Pilih luminer representatif untuk pengujian yang sesuai dengan standar nasional atau perusahaan; luminer baru harus menjalani perawatan penuaan (untuk luminer LED, periode penuaan 24 jam disarankan);

Periksa apakah perlengkapan lampu tidak memiliki kerusakan yang terlihat, apakah kabel internal terhubung dengan benar, dan bersihkan noda apa pun dari permukaan perlengkapan lampu.

7.2 Introduction to Test Mode

Mode C-Gamma: Cocok untuk sebagian besar perlengkapan pencahayaan, seperti lampu jalan, lampu downlight, lampu panel, dan lampu pertumbuhan tanaman.

High Precision Rotation Luminaire Goniophotometer-Figure63

Gambar 74

Mode Beta: Cocok untuk lampu sorot, lampu banjir, dan perlengkapan serupa; memerlukan pemasangan lengan tambahan.

High Precision Rotation Luminaire Goniophotometer-Figure64

Gambar 75

Mode Alpha: Cocok untuk lampu otomotif dan lampu sein.

High Precision Rotation Luminaire Goniophotometer-Figure65

Gambar 76

Catatan: LSG-1890BCCD can perform C-Gamma and B-Beta tests. For A-Alpha testing, a Type A goniophotomoeter is required; please contact us if needed.

8. Tes C-Gamma

8.1 Memasang dan Menyalakan Lampu

8.1.1 Pemasangan Perlengkapan Lampu

Kami memiliki dua alat uji C-Gamma; prosedur ini menggunakan alat uji berbentuk salib sebagai contoh.

Jumlah Note
Large cross-shaped clamps are used for various types of lighting fixtures
E27 Fixture for Testing Globe Bulbs

High Precision Rotation Luminaire Goniophotometer-Figure66

Gambar 77

Attach the fixture to the Goniophotomoeter. Turn on the wall-mounted cross-line laser and verify that the cross-line laser is projected accurately onto the markings made on the opposite wall.

High Precision Rotation Luminaire Goniophotometer-Figure67

Gambar 78

On the angle control interface, select the near-field photometric probe. Rotate the Gamma axis 90 degrees so that the crosshair laser points directly at the center of the crosshair fixture. Mount the lamp under test onto the crosshair fixture. Adjust the lamp’s position so that the crosshair laser is centered on the light-emitting surface of the lamp under test, and the light-emitting surface is parallel to the crosshair fixture.

High Precision Rotation Luminaire Goniophotometer-Figure68

Gambar 79

Then rotate the Gamma axis back to 0 degrees. Turn the dial here.

High Precision Rotation Luminaire Goniophotometer-Figure69

Gambar 80

Align the light-emitting surface of the fixture with the vertical laser beam of the crosshair.

High Precision Rotation Luminaire Goniophotometer-Figure70

Gambar 81

Caution is required when installing directional luminaires. When the angle of the C-plane on the Goniophotomoeter is 0 degrees, according to streetlight installation standards, the light pole should point directly upward, as shown in the figure below.

High Precision Rotation Luminaire Goniophotometer-Figure71

Gambar 81.1

When the angle of the C-plane on the goniophotomoeter is 0 degrees, the strip light should be installed vertically in accordance with industry standards, as shown in the figure below.

High Precision Rotation Luminaire Goniophotometer-Figure72

Gambar 81.2

8.1.1 Menyalakan Lampu

Connect the power cord to the luminaire under test. The connection method for the luminaire under test is the same as that for the standard luminaire; refer to Section 6.1.1.

Sesuaikan sakelar kabinet sesuai dengan parameter luminer yang sedang diuji.

Jumlah Note
If the sample luminaire is AC-powered, switch to the AC setting; if it is DC-powered, switch to the DC setting.
Internal/External Sampling Switch for Electrical Parameters. External sampling uses a four-wire configuration and provides more accurate voltage readings. Unless otherwise specified, please set this switch to “External Sampling”; no further adjustments are necessary.

High Precision Rotation Luminaire Goniophotometer-Figure73

Gambar 82

Jika lampu sampel menggunakan daya AC.

Jumlah Note
Select AC Power
Based on the sample parameters, enter the correct voltage and frequency.

High Precision Rotation Luminaire Goniophotometer-Figure74

Gambar 83

Jika lampu sampel menggunakan daya DC.

Jumlah Note
Select a DC Power Supply
Berdasarkan parameter sampel, masukkan tegangan dan arus yang benar.
Select the output mode: CC stands for constant current, and CV stands for constant voltage. If you select constant current, you need to enter the sample’s rated current value; for the voltage, simply enter the maximum output voltage of this DC power supply. If you select constant voltage, enter the sample’s rated voltage and the DC power supply’s maximum output current.
Catatan: Jika model catu daya DC Anda adalah DC3010, its maximum output voltage is 30V and its maximum output current is 10A.

High Precision Rotation Luminaire Goniophotometer-Figure75

Gambar 84

The software offers two methods for turning the sample lamp on and off, and users can choose between them.

Jumlah Note
If you check the “Auto Turn On” and “Auto Turn Off” options, the software will automatically power on the sample according to the selected power source and the entered electrical parameters when the test begins, and will automatically cut off the sample’s power when the test ends.
If the “Auto On” and “Auto Off” options are not checked, you can still click this button; the software will illuminate the sample based on the selected power source and the entered electrical parameters. Clicking this button again will cut off the power to the sample.

High Precision Rotation Luminaire Goniophotometer-Figure76

Gambar 85

8.2 Pengaturan Parameter Pemanasan Awal

Jumlah Note
If you check the “Preheat” box, the software will preheat the luminaires according to the specified preheat parameters when the test begins, and the test will start automatically once preheating is complete.
If you do not check the “Preheat” box, the software will start the test immediately.
Pengaturan Parameter Pemanasan Awal

High Precision Rotation Luminaire Goniophotometer-Figure77

Gambar 86

Click “Preheat Parameter Settings.”

Jumlah Note
Pengaturan Durasi Pemanasan: Tujuan dari pemanasan adalah untuk memungkinkan perlengkapan lampu tetap menyala selama jangka waktu tertentu hingga parameter cahaya dan warnanya stabil, setelah itu pengujian akan dilanjutkan secara otomatis.
Sampling interval: The software automatically samples data at the set interval throughout the entire warm-up period to obtain real-time data. We recommend setting this to 5 seconds or longer.
If checked, the test will automatically skip the remaining warm-up time and proceed once the light fixture has reached a stable state; if unchecked, the test will begin only after the warm-up period has been completed according to the set duration.
Definition of a luminaire’s steady-state condition. If the values are set to 20 and 0.5%, the luminaire is considered to be in a steady state if the parameter has not varied by more than 0.5% over the past 20 minutes.
Anda dapat memilih parameter mana yang harus digunakan perangkat lunak untuk menentukan apakah perlengkapan lampu stabil: parameter optik, parameter listrik.
Perangkat lunak akan menyimpan kurva perubahan untuk parameter yang dipilih selama periode pemanasan dalam laporan pengujian.
Setelah Anda selesai mengkonfigurasi pengaturan, klik OK untuk menyimpan dan keluar.

High Precision Rotation Luminaire Goniophotometer-Figure78

Gambar 87

8.3 Mengonfigurasi Parameter Pengujian dan Memulai Pengujian

Klik “Mulai Tes.”

High Precision Rotation Luminaire Goniophotometer-Figure79

Gambar 88

Buka layar pengaturan parameter pengujian.

Jumlah Note
Jenis perlengkapan pencahayaan: lampu dalam ruangan, lampu luar ruangan, atau lampu pertumbuhan tanaman—ketiga jenis perlengkapan pencahayaan ini tunduk pada pengujian C-Gamma.
Select whether the luminaire is symmetrical or asymmetrical. Unless there are specific requirements, select the first option (asymmetrical) regardless of whether the luminaire being tested is symmetrical or not (to obtain the luminaire’s actual data).
Rentang sudut bidang C: Pilih 0 hingga 180 derajat; tidak perlu modifikasi. Untuk data yang berkisar dari 180 hingga 360 derajat, perangkat lunak akan secara otomatis mengonversinya.
C-plane angle intervals. Symmetrical luminaires, such as indoor lights, can use a larger interval (e.g., 30 degrees), while asymmetrical luminaires, such as streetlights, can use a smaller interval (e.g., 10 degrees). The smaller the angle interval, the higher the test accuracy, but the longer the test will take. If time permits, we recommend selecting a smaller angle interval.
Gamma plane angle range: If the sample’s light irradiation angle exceeds 180 degrees (e.g., incandescent bulbs), select -180 degrees to 180 degrees; if the sample’s light irradiation angle is less than 180 degrees (e.g., panel lights), you may select a Gamma
plane angle interval ranging from -90 degrees to 90 degrees. Unless there are special requirements, a 1-degree interval is sufficient. The specific angle interval selected here has no effect on test speed.
If the system has multiple photometric probes, please select the one you want to use.
For standard luminaires undergoing standard testing, select “Normal Speed”; if the luminaire being tested has an extremely narrow beam angle, select “Slow Speed” to obtain more accurate test results; for expedited testing, select “Fast Speed.”
Fixed Range: If you are consistently measuring the same type of light fixture and the luminous intensity data does not vary significantly, you can check this box to save time. If you have ample time, or if you frequently test lamps with different luminous intensities, we recommend leaving this box unchecked; in this case, the photometric probe will automatically select the appropriate range for each lamp and enter
continuous scanning mode. This is suitable when the angular interval of the Gamma plane is between -180 degrees and 180 degrees. During the test, when the Gamma angle approaches ±180 degrees, simultaneously rotating the C-axis to the next scanning position—that is, the Gamma axis rotates continuously without stopping to save time. If you have ample time, we recommend leaving this option unchecked.
To set the monitoring angle during lamp warm-up, simply set it to (0, 0).
Normalisasi mengacu pada keputusan apakah akan menerapkan nilai seragam pada baris pertama data dalam tabel intensitas cahaya pada laporan pengujian; standar tertentu, seperti standar lampu jalan AS, mungkin mengharuskan Anda untuk melakukan penyesuaian serupa. Ada tiga pilihan: tanpa normalisasi, normalisasi ke nilai maksimum (yang akan sedikit meningkatkan fluks cahaya total luminer), dan normalisasi ke nilai rata-rata (yang pada dasarnya tidak mengubah fluks cahaya).
If you need to remove all stray light outside a specific angle, check this box and enter the desired angle
. For example, if you check this option and enter 60, the data outside ±60 degrees of the luminaire’s gamma plane will automatically be set to 0.
Enter the number of light sources in the luminaire and the rated luminous flux of each light source. If you do not know these values, you may leave them blank; this will not affect the test results.
1 Masukkan dimensi permukaan pemancar cahaya sampel. Untuk perlengkapan pencahayaan dalam ruangan, perangkat lunak memerlukan data ini untuk menghitung kurva batas luminansi, UGR, dan parameter lainnya.
Enter information about the sample light fixture
Enter the temperature and humidity of the test environment (i.e., the darkroom)
3 Klik OK untuk menuju layar pemanasan, atau menuju layar pengujian.

High Precision Rotation Luminaire Goniophotometer-Figure80

Gambar 89

Then wait for the software testing to finish.

9. B-Beta Testing

9.1 Install and Turn On the Lights

9.1.1 Lighting Fixture Installation

Turn on the crosshair laser and make sure it is still shining on the mark you made on the opposite wall.

Find the square B-Beta test fixture.

High Precision Rotation Luminaire Goniophotometer-Figure81

Gambar 92

When performing the B-Beta test, reattach the Goniophotomoeter’s auxiliary arm. Then, install the B-Beta fixture vertically as shown in the figure.

High Precision Rotation Luminaire Goniophotometer-Figure82

Gambar 93

Jumlah Note
Click to switch to the angle display mode
Klik ① untuk menampilkan sudut B-Beta di sini, bukan sudut C-Gamma.
Klik untuk membuka antarmuka kontrol sudut.

High Precision Rotation Luminaire Goniophotometer-Figure83

Gambar 94

Jumlah Note
When angle B is 0 degrees, the fixture should be in a vertical position; this can be verified using a level.
Masukkan 0
Click to set the Beta angle to 0 degrees. Note: The 0-degree position for the Gamma angle is not the same as the 0-degree position for the Beta angle.

High Precision Rotation Luminaire Goniophotometer-Figure84

Gambar 95

After rotating Beta to the 0-degree position, mount the floodlight onto the fixture, ensuring that the light-emitting surface of the floodlight faces the photometric probe. Adjust the mounting position of the fixture so that the cross-shaped vertical laser beam aligns with the light-emitting surface of the floodlight.

High Precision Rotation Luminaire Goniophotometer-Figure85

Gambar 96

To ensure the sample is centered in the Goniophotomoeter, rotate the Beta fixture to the position shown in the figure below so that the sample’s light-emitting surface faces the crosshairs. Use the crosshairs to fine-tune the position of the sample holder.

High Precision Rotation Luminaire Goniophotometer-Figure86

Gambar 97

Once the light fixture is installed, you can turn off the crosshair laser.

9.1.2 Menyalakan Lampu

Please refer to 8.1.2.

9.2 Pengaturan Parameter Pemanasan Awal

Please refer to Section 8.2.

9.3 Mengonfigurasi Parameter Pengujian dan Memulai Pengujian

Klik “Mulai Pengujian” untuk menuju ke layar pengaturan parameter pengujian.

Jumlah Note
Silakan pilih “Floodlight,” dan mode pengujian akan secara otomatis beralih ke B-Beta.
Rentang sudut untuk Bidang B adalah -90 derajat hingga 90 derajat. Tidak perlu melakukan penyesuaian.
the angle intervals for Plane B; you can select smaller angle intervals, such as 5 degrees or 10 degrees. The smaller the interval, the higher the test accuracy, but the longer it takes.
The Beta plane angle range is -90 degrees to 90 degrees. There is no need to adjust the
Beta plane interval angle; unless there are special requirements, select 1 degree. Different interval angles here have no effect on test speed.

High Precision Rotation Luminaire Goniophotometer-Figure87

Gambar 98

For other parameter settings, refer to Section 8.1.3. Once you have configured the settings, click “Start” and wait for the software to complete the test.

10. Pemrosesan Laporan Tes

10.1 Organizing Database Files

Jumlah Note
By default, all test reports are automatically stored in the currently open database file in the order in which the tests were run.
Untuk menghapus laporan, pilih laporan tersebut lalu klik “Hapus.”
To make it easier to find test reports, click the blank button shown in the image below to create a new database file, and store reports for different lighting fixtures in separate database files. After creating a new database file, you can proceed directly with testing without having to recalibrate or reconfigure the system.
Click “Open” to view test reports from other database files.

High Precision Rotation Luminaire Goniophotometer-Figure88

Gambar 100

Note 1: The next time the software is launched, it will automatically load the database file used during the previous session. If the path to that database file has changed or its name has been modified, the software will prompt you to create a new database file.

Catatan 2: Sebisa mungkin hindari menempatkan file basis data di desktop komputer Anda.

10.2 Pengaturan Tampilan Laporan Pengujian

Jumlah Note
You can double-click to upload a picture of the light fixture.
Anda dapat mengklik dua kali konten yang ditampilkan di area ini untuk mengeditnya.

High Precision Rotation Luminaire Goniophotometer-Figure89

Gambar 101

Jumlah Note
Klik "Pengaturan"
For unit selection, “cd” is generally used; however, some older standards—such as those regarding energy-saving light bulbs—may require the use of “cd/klm.”
Streetlight C0 angle setting. The CIE/GB and IESNA standards have inconsistent requirements; please select the standard you are following.
Customize the divergence angle. The software automatically displays the half-peak angle (commonly referred to as the beam angle, calculated as 50% of the maximum luminous intensity) and the effective beam angle (calculated as 10% of the maximum luminous intensity). If you wish to customize the divergence angle, enter the percentage of the maximum light intensity and check the box (e.g., 0.25 for 25%). In this case, the test report will display not only the half-peak angle and effective beam angle but also your custom divergence angle.
Centang kotak ini untuk menampilkan sudut kerucut intensitas cahaya maksimum dalam laporan.

High Precision Rotation Luminaire Goniophotometer-Figure90

Gambar 102

Jumlah Note
Switch to the “Energy Efficiency Rating” settings screen, then select the applicable standard and luminaire type.
Klik untuk menerapkan pada laporan tes saat ini
Klik OK untuk menyimpan dan keluar.

High Precision Rotation Luminaire Goniophotometer-Figure91

Gambar 103

Each display window has several settings that allow you to customize the display. Users can experiment with adjusting these settings to achieve their desired display effects.

High Precision Rotation Luminaire Goniophotometer-Figure92

Gambar 103.1

10.3 Memproses Data Uji

10.3.1 Double-sided luminaires

For lamps that emit light from both sides, use the indoor lamp test mode to test each side individually—that is, set the gamma plane angle range to -90 degrees to 90 degrees during testing—and then generate two test reports.

Jumlah Note
Tekan dan tahan tombol “Ctrl” pada keyboard komputer Anda, lalu klik untuk memilih dua laporan pengujian yang baru saja Anda terima.
Klik “Sintesis Data” di bilah navigasi perangkat lunak.
Di jendela pop-up, Anda dapat mengklik untuk beralih antara pancaran cahaya ke atas dan ke bawah.
Pilih “Komposit”
Klik OK untuk menyimpan dan keluar.

High Precision Rotation Luminaire Goniophotometer-Figure93

Gambar 104

Ini akan menghasilkan laporan uji sintetis baru, seperti yang ditunjukkan pada gambar di bawah ini.

High Precision Rotation Luminaire Goniophotometer-Figure94

Gambar 105

10.3.2 Conversion of Test Report Types

Jumlah Note
Click “Type Conversion”
Laporan pengujian dapat dikonversi di antara empat jenis.
When converting to a C-Gamma test report, you can use the “termination angle” of the C-plane to generate a symmetric test report. If the termination angle is set to 360 degrees, there is no symmetry; if the termination angle is set to 180 degrees, the report can be converted to an axially symmetric test report; if the termination angle is set to 90 degrees, the report can be converted to a quadrant-symmetric test report; if the termination angle is set to 0 degrees, the report can be converted to a fully symmetric test report. This feature is primarily intended for symmetric indoor lighting fixtures; certain lighting design software programs, such as DIALUX, may require you to use this setting when importing fully symmetric IES files.

High Precision Rotation Luminaire Goniophotometer-Figure95

Gambar 106

10.4 Mengekspor Data Uji dan Mencetak Laporan Uji

10.4.1 Mengekspor Data Uji

Jumlah Note
Klik logo merah di pojok kiri atas perangkat lunak.
Click “Export Data” in the drop-down menu

High Precision Rotation Luminaire Goniophotometer-Figure96

Gambar 107

File dapat diekspor dalam berbagai format standar.

High Precision Rotation Luminaire Goniophotometer-Figure97

Gambar 108

10.4.2 Mencetak Laporan Uji

Pengaturan pencetakan.

Jumlah Note
Klik untuk membuka layar “Pengaturan Cetak”
Laporan ini menampilkan pilihan bahasa; Anda dapat memilih antara bahasa Mandarin dan bahasa Inggris.
Format Nomor Halaman
Informasi Header Laporan
Judul Laporan dalam Bahasa Mandarin dan Inggris
General Report Content—Applicable to Reports on All Types of Lighting Fixtures
Karena lampu dalam ruangan, lampu jalan, lampu sorot, dan lampu pertumbuhan tanaman memiliki parameter kunci yang berbeda, isi laporan akan bervariasi sesuai dengan itu. Anda dapat memilih item spesifik yang ingin Anda sertakan dalam laporan.
Opsi TM-30
CIE color space diagram or SDCM color tolerance diagram—you can choose only one of them.

High Precision Rotation Luminaire Goniophotometer-Figure98

Gambar 109

Jumlah Note
Klik untuk mencetak laporan pengujian yang saat ini dipilih
Klik untuk mencetak semua laporan pengujian dalam file basis data saat ini.

High Precision Rotation Luminaire Goniophotometer-Figure99

Gambar 110

11. Pemeliharaan dan Perawatan Peralatan Harian

11.1 Perawatan Rutin

Start-up procedure: First, turn on the cabinet’s main power switch; then, turn on the power switches for the equipment inside the cabinet and in the darkroom one by one; finally, launch the software. Shutdown procedure: First, close the software; then, turn off the power switches for the equipment inside the cabinet and in the darkroom one by one; finally, turn off the cabinet’s main power switch.

Keep the darkroom control room clean and dust-free to prevent dust from entering the equipment. If there is surface dust on the equipment, wipe it off with a dry, soft cloth.

Hindari membiarkan ruang gelap terpapar kelembapan tinggi dalam waktu lama; sediakan alat pengering udara jika diperlukan.

11.2 Perawatan Rutin

11.2.1 Membersihkan Probe Fotometri

Jika ruang gelap tidak dapat dijaga agar bebas debu, bersihkan permukaan peka cahaya pada probe fotometer secara teratur (misalnya, sebulan sekali) berdasarkan kondisi aktual untuk mencegah penumpukan debu dan memengaruhi akurasi pengujian.

When cleaning, please use a professional lint-free cloth to remove dust without scratching the device.

11.2.2 Standard Lamp

Lampu standar berlaku selama satu tahun. Jika Anda masih perlu mengkalibrasi ruang gelap setelah satu tahun, kami sarankan untuk membeli lampu standar yang baru.

Perangkat Lunak 11.2.3

Kami akan memperbarui perangkat lunak dari waktu ke waktu untuk menambahkan fitur dan memperbaiki bug. Jika Anda mengalami masalah saat menggunakan perangkat lunak, jangan ragu untuk menghubungi kami kapan saja untuk mendapatkan versi terbaru.

11.3 Penonaktifan dan Pemeliharaan Jangka Panjang

11.3.1 Pembersihan dan Perlindungan Peralatan

Bersihkan semua permukaan peralatan secara menyeluruh, dan tutupi unit utama, probe, dan kabinet dengan kain anti debu; putuskan semua sumber daya listrik.

11.3.2 Perlindungan Lingkungan

Ensure that the darkroom is dry and well-ventilated to prevent equipment from getting damp. If the ambient humidity is high, keep a dehumidifier on hand.

11.3.3 Monthly Power-On

Lakukan uji penyalaan daya sebulan sekali untuk memastikan semua perangkat keras sistem berfungsi dengan baik.

12. Pemecahan Masalah dan Penyelesaian Masalah Umum

12.1 Kesalahan Komunikasi

Cause of the problem: Check whether the communication cable is securely connected and whether it is plugged into the wrong port; verify that the driver is installed correctly; try a different USB port or communication cable to rule out hardware issues.

Solusi: Instal ulang driver, sambungkan kabel komunikasi sesuai dengan pengenal perangkat, mulai ulang perangkat lunak, dan coba lagi.

12.2 Kesalahan Pengujian

12.2.1 Anomali Fluks Cahaya

Cause of the malfunction: Check whether the light source is stable and whether the warm-up time is sufficient; verify that the equipment calibration is valid, that the test distance is correct, and that the standard light is functioning properly; check whether there is any external light interference in the darkroom and whether the light path is obstructed.

Solution: Extend the warm-up time, recalibrate the instrument, turn off other light sources in the darkroom, and remove any obstructions from the optical path.

12.2.2 Anomali UGR

Cause of error: Dimensions of the luminaire’s light-emitting surface were not entered; the luminaire’s beam angle is less than 45°; the calculation standard was selected incorrectly.

Solusi: Masukkan dimensi permukaan pemancar cahaya luminer (panjang/lebar/tinggi) dengan benar; verifikasi bahwa luminer tersebut adalah luminer dalam ruangan (sudut pancaran harus memadai); pilih standar CIE 190:2010 dan hitung ulang.

12.3 Goniophotometer tidak berputar

Cause of the malfunction: Check whether the host power is on and whether the wiring is correct; check whether the angle controller inside the cabinet is operating normally; confirm whether any foreign objects are jammed in the rotating shaft.

Solusi: Periksa catu daya dan kabel, dan singkirkan benda asing apa pun; mulai ulang sistem dan perangkat lunak, lalu coba lagi.

12.4 Software Malfunctions

12.4.1 Kerusakan Perangkat Lunak

Penyebab masalah: Periksa apakah Anda masuk ke sistem operasi komputer dengan akun administrator; verifikasi bahwa perangkat lunak telah terinstal sepenuhnya; periksa apakah file basis data rusak.

Solution: Log in to the computer using an administrator account; reinstall the software and drivers; create a new database file.

12.4.2 Perangkat lunak menampilkan pesan “Berkas bukan basis data”

Penyebab masalah: File basis data telah dihapus atau dipindahkan; program default yang digunakan untuk membuka file LGF telah diubah.

Solution: Create a new database file and save it; restore the default association for LGF files.

12.5 Kegagalan Daya

12.5.1 DC Power Output Abnormalities

Causes of Failure: The cabinet’s AC/DC switch was not set to DC; the sample voltage or current exceeded the DC power supply limits; a short circuit in the sample;

Troubleshooting: Please verify that the sample’s voltage and current are within the DC power supply’s output range. For example, the DC3005 Model ini memiliki tegangan keluaran maksimum 30 V dan arus keluaran maksimum 5 A. Periksa apakah sampel mengalami korsleting; jika ya, perbaiki sampel sebelum melakukan pengujian ulang.

12.5.2 Tidak Ada Output AC

Cause of the fault: The cabinet’s AC/DC switch was not set to the AC position; the overload protection was triggered;

Troubleshooting Steps: Check the AC power output wiring and sample connections to ensure there are no loose connections or short circuits; if the overload protection is triggered, turn off the power, allow the unit to cool for 5 minutes, and then restart it. Ensure that the sample power remains within the AC power output range; for example, the LSP-500VARC memiliki daya keluaran maksimum 500W.

Lampiran & Unduhan

Video

LSG-1800A Goniofotometer Luminer Rotasi Presisi Tinggi - Video Pengoperasian

LSG-1890B Goniofotometer Luminer Rotasi Presisi Tinggi - Video Pengoperasian

LSG-1800B Rotasi Presisi Tinggi Luminaire Goniophotometer - Video Operasi