Rotasi Presisi Tinggi Luminer Goniophotometer
Pengguna manual
berlaku LISUN Model: LSG-1890B
1. Membuka Kemasan dan Mengenal Peralatan Secara Dasar
1.1 Membuka Kemasan dan Pemeriksaan
When unpacking the device, handle it gently to avoid scratching the casing with sharp objects. After unboxing, first inspect the exterior of the device to ensure there is no visible deformation or dents, and that the buttons are not damaged or loose. If you notice any abnormalities with the device or its accessories, do not turn it on. Please contact us immediately.
1.2 Konfigurasi dan Gambaran Umum Sistem
1.2.1 Komponen Perangkat Keras Inti
Catatan: Tabel di bawah ini mencantumkan komponen perangkat keras inti dari keseluruhan sistem. Peralatan yang Anda beli mungkin hanya mencakup beberapa komponen ini; harap perhatikan hal ini saat mengikuti petunjuk dalam manual ini.
| Nama komponen | Pilih Model | Fitur Inti |
| Unit Utama Goniophotometer | LSG-1890B | Used in conjunction with a photometric probe to measure photometric parameters and their spatial distribution, such as luminous flux and light distribution curves. |
| Probe fotometri medan dekat | PM400C (Utara) | Suitable for standard darkrooms, such as those around 10 meters in size |
| Far-Field Photometric Probe | PM400C (F) | Optional. Suitable for extra-long darkrooms, such as those longer than 20 meters. |
| Lampu Standar | SLS-150W | Used to calibrate the darkroom (luminance sensor) |
| E27 Fixture | / | Untuk pengujian C-Gamma; cocok untuk lampu pijar. |
| Penjepit Silang | Untuk pengujian C-Gamma; cocok untuk perlengkapan pencahayaan dalam dan luar ruangan lainnya. | |
| Penjepit Persegi | Untuk pengujian B-Beta; cocok untuk lampu sorot | |
| Laser Garis Silang | Digunakan selama instalasi dan pengoperasian awal; digunakan selama kalibrasi dan pengujian untuk menentukan posisi dan orientasi lampu standar dan luminer sampel. | |
| Kabinet | CASE-19 | Used to house angle controllers, power meters, and DC and AC power supplies. Includes a power strip, an RS485 communication module, and more. |
| Pengontrol Sudut | RT-200C | Sudut di mana Goniophotometer diposisikan untuk tampilan dan kontrol. Umumnya, tidak diperlukan penyesuaian manual; kontrol ditangani oleh perangkat lunak. |
| Pengukur daya | Seri LS | Measure electrical parameters such as voltage, current, power, and power factor of the sample |
| DC Power Supply | Seri DC | Menyediakan catu daya DC yang stabil untuk sampel DC, lampu standar, dan lampu bantu. |
| AC Power | LSP Series | Provides a stable, pure sine-wave AC power supply for AC samples |
1.2.2 Perangkat lunak, manual, sertifikat, dll.
At the time of shipment, a download link containing all the documents was sent via email. Please download them promptly. If you missed the email, you can contact us to request a new download link.
Includes software, user manuals, warranty cards, standard lamp calibration certificates, and other important documents, as well as other relevant documents.
1.4 Parameter Pengukuran
Illuminance data, illuminance distribution, area luminous flux, luminaire efficiency, luminance distribution (optional), utilization factor, luminance limit curve, glare rating, isophote curves, maximum allowable height-to-distance ratio, luminaire curve vs. illumination area, isointensity curves, effective beam angle, EEI, UGR, PAR, PPF, and PPFD spatial distribution tests, etc.
2. Catatan Penting
2.1 Keamanan Listrik
The system must be connected to a stable AC power source rated at 220 V ±5% with a frequency of 50 Hz or 60 Hz. Ensure proper grounding with a grounding resistance of ≤4 Ω. Do not use the equipment if the power connections are loose or the cables are damaged.
2.2 Keselamatan Lingkungan
When using the optical darkroom, the ambient temperature must be maintained at 25±1°C, and the relative humidity must be maintained between 10% and 65% RH;
Both the control room and the darkroom must be kept dry, well-ventilated, dust-free, and free of corrosive gases;
Avoid placing the device near strong magnetic or electric fields; it must be placed on a level, sturdy surface.
2.3 Prosedur Operasi Ruang Gelap
2.3.1 Deskripsi Ruang Gelap
Please confirm that the darkroom has been constructed exactly according to the darkroom drawings we provided. If there are any discrepancies in the actual dimensions, please inform us before installing and commissioning the equipment so that we can adjust the plan, ensuring that the installation and commissioning proceed smoothly and that the test results are accurate;
The interior of the darkroom must be completely painted matte black—that is, a non-reflective pure black—including all walls, ceilings, floors, partitions, doors, and so on. A small number of components, such as lighting fixtures, air conditioning units, and power switches, may not be painted black, provided they do not appear in the optical path between the Goniophotomoeter’s main unit and the photometric probe;
Once the equipment has been installed and commissioned, do not move any of the hardware components inside the darkroom, including the main unit base, the photometer probe base, the photometer probe orientation, and the cross-laser base. Doing so may result in inaccurate test results.
2.3.2 Precautions for Using the Darkroom
Lepaskan sepatu Anda atau kenakan penutup sepatu saat memasuki ruang gelap untuk mencegah debu masuk;
Luminer sampel harus dipasang dengan aman untuk mencegahnya jatuh dan merusak Goniophotometer selama pengujian;
Do not place any objects on the main unit of the Goniophotomoeter, and do not place any objects within the Goniophotomoeter’s rotation range;
During testing, the darkroom must remain sealed to prevent external light from interfering with the test results.
2.3.3 Precautions When the Darkroom Is Not in Use
Keep the darkroom closed; unauthorized persons are not allowed to enter; prevent dust from entering the darkroom.
2.4 Prosedur Pengoperasian Lampu Standar
The standard lamp has a lifespan of one year and is used for calibrating the photometer probe in a darkroom during installation and commissioning;
Jika laboratorium Anda diharuskan untuk mengkalibrasi ruang gelapnya setiap tahun, Anda dapat membeli lampu standar baru;
Lampu standar akan terasa panas segera setelah digunakan; jangan disentuh. Biarkan dingin selama 10 menit sebelum mengembalikannya ke wadah lampu;
The standard lamp used in the darkroom is a directional standard lamp, and its direction is controlled by its bracket. Do not remove the bulb from its bracket; doing so will render the standard lamp inoperable.
2.5 Lainnya
Untuk hal-hal yang tidak tercakup dalam panduan ini, harap berhati-hati atau hubungi kami.
3. Perakitan Perangkat Keras
3.1 Cabinet Equipment Assembly
3.1.1 Introduction to the Cabinet
Place the cabinet on a level, sturdy workbench, leaving sufficient space for operation.
| Jumlah | Note |
| ① | Wiring Diagram: Please assemble the equipment according to the wiring diagram. |
| ② | Panel distribusi daya, digunakan untuk menghubungkan kabel daya semua perangkat di dalam kabinet. |
| ③ | Cooling fan power cord: Please connect it to the power strip inside the cabinet. |
| ④ | Kotak Komunikasi RS-485, digunakan untuk menghubungkan semua perangkat melalui kabel komunikasi RS-485. |
| ⑤ | Harap sambungkan adaptor daya untuk kotak komunikasi RS485 ke stopkontak di dalam kabinet. |
| ⑥ | The USB cable for the RS485 communication box must be connected to the computer. |
| ⑦ | Pengkabelan Pengontrol Sudut RT-200C |
| ⑧ | LS Series Power Meter Wiring |
| ⑨ | Koneksi Daya DC |
| ⑩ | Pengkabelan Daya AC |
| ⑪ | The main power cable for the cabinet, which supplies power to all equipment inside the cabinet, should be routed out through cable opening ② and connected to the power source only after all equipment has been fully assembled. |
| 1 | Cable opening—the main power cord can be routed through here. Note: The main power cord plug is a Chinese standard plug; you may need to provide your own power adapter. |

Gambar 1
3.1.2 Instalasi RT-200C
Untuk saat ini, yang perlu Anda lakukan hanyalah mencolokkannya.

Gambar 2
3.1.3 Installation of the LS Power Meter
| Jumlah | Note |
| ① | Connected via an RS-485 communication cable; the other end is connected to the RS-485 communication box inside the cabinet. |
| ② | Kabel daya sudah terhubung; ujung lainnya sudah terhubung ke panel distribusi daya di dalam kabinet. |
| ③ | V1 test lead, connected to the red terminal of the power meter’s voltage V |
| ④ | Kabel uji V2, dihubungkan ke terminal hitam tegangan V pada meteran daya. |
| ⑤ | A1 test lead, connected to the red terminal of the power meter’s current A |
| ⑥ | A2 test lead, connected to the black terminal for current A on the power meter |

Gambar 3
3.1.4 Instalasi Catu Daya DC
| Jumlah | Note |
| ① | Terhubung melalui kabel komunikasi RS-485; ujung lainnya sudah terhubung ke kotak komunikasi RS-485 di dalam kabinet. |
| ② | Hubungkan kabel keluaran daya DC, pastikan untuk mencocokkan terminal positif dan negatif. |
| ③ | The power cord is connected; the other end is already connected to the power strip inside the cabinet. |

Gambar 4
3.1.5 Instalasi Daya AC
Please locate the following AC power output cable in the accessories:

Gambar 5
| Jumlah | Note |
| ① | Connected via an RS485 communication cable; the other end is already connected to the RS485 communication box inside the cabinet. |
| ② | The power cord is connected; the other end is already connected to the power strip inside the cabinet. |
| ③ | Lewatkan kabel keluaran daya AC melalui lubang kabel ④ dan sambungkan ke posisi ⑤. |
| ④ | Lubang kawat |
| ⑤ | AC power output cable: Pass it through cable hole ④ and connect it to position ③. |

Gambar 6

Gambar 7
3.2 Pemasangan Unit Utama Goniophotometer dan Probe Goniophotometer
3.2.1 Host Assembly
The same number is engraved on the main unit’s base tapered sleeve ① and the corresponding position on the main unit ②. Rotate the tapered sleeve until the two numbers are aligned, then secure the tapered sleeve using the hex socket screws provided as standard.

Gambar 8

Gambar 9
Remove the screws securing the dust cover on the turntable, and pull the dust cover off.

Gambar 10
Use a forklift to lift the turntable; since the turntable is quite heavy, three to four people can also carefully lift it. Align the dust cover on the turntable with the terminal block on the base, then slowly lower the turntable onto the base. Hold the terminal block in place with your hand so that it slides into the turntable along with the tapered sleeve, preventing the terminal block from being crushed by the turntable.

Gambar 11
To fine-tune the turntable’s direction, follow these steps in the order listed below.
| Jumlah | Note |
| ① | Align the six screw holes on the turntable with the six screw holes on the base, and use the standard hex screws and washers provided to secure the turntable to the base . Note: Be sure to use all the standard washers; otherwise, the screws may not fit properly, which could damage the equipment during turntable rotation. |
| ② | Plug the base terminal block into the corresponding terminal block socket on the turntable. |
| ③ | Then put the dust cover back in place |

Gambar 12

Gambar 13
The four feet on the bottom of the unit are height-adjustable.
| Jumlah | Note |
| ① | Turn the screw at the bottom to raise the leg until it is off the ground. |
| ② | At the same time, make sure the top screw isn’t pressing against anything; you can unscrew it slightly. |

Gambar 14
Position the goniophotomoeter in the correct location according to the darkroom blueprints. Align the four sides of the base as closely as possible with the darkroom walls. Then remove the auxiliary arm.

Gambar 15
Next, please turn all four legs of the Goniophotomoeter until they are in contact with the floor.
3.2.2 Assembly of the Photometric Probe and Probe Holder
Take, for example, a darkroom with only one photometer.
Rakitlah probe fotometri dan dudukan probe, lalu posisikan keduanya di lokasi yang sesuai menurut gambar ruang gelap; untuk saat ini tidak perlu mengencangkannya dengan sekrup.

Gambar 16

Gambar 17
3.2.3 Pengkabelan Sistem
| Jumlah | Note |
| ① | The four terminal blocks ①on the cabinet are connected to the Goniophotomoeter⑩; simply match them one-to-one based on the numbers. |
| ② | Kedua terminal tersebut terhubung ke blok terminal di dalam kabinet dan digunakan untuk memasok daya ke semua peralatan di ruang gelap, termasuk Goniophotometer dan probe fotometrik. Hal ini memungkinkan kontrol yang mudah terhadap semua peralatan ruang gelap melalui sakelar daya kabinet. Catatan: Peralatan kamar gelap juga dapat dioperasikan langsung dari listrik utama. |
| ③ | Cabinet grounding terminal: Please connect it to ground using a wire. |
| ④ | Kabel daya kabinet, terhubung ke sumber daya 220V, 50/60 Hz. |
| ⑤ | Power Cord for the Main Unit of the Goniophotomoeter |
| ⑥ | Hubungkan terminal ground Goniophotometer ke ground menggunakan kawat. |
| ⑦ | Ini adalah kabel komunikasi satu-ke-dua; satu ujung kabel RS-485 terhubung ke kotak komunikasi RS-485 di dalam kabinet, dan ujung lainnya terhubung ke posisi ⑭ di bagian belakang RT-200C. |
| ⑧ | Photometer probe trigger wire; connect to probe ⑬. When only one probe is installed, the default setting is near-field (N). |
| ⑨ | If you have only one photometer in your darkroom, there is no need to connect the far-field (F) sensor. |
| ⑩ | Keempat blok terminal ⑩ pada Goniophotometer terhubung ke kabinet ①; cukup pasangkan satu per satu sesuai dengan nomornya. |
| ⑪ | Adaptor Daya untuk Probe Fotometri |
| 1 | Kabel komunikasi RS-485 untuk probe fotometri; ujung lainnya terhubung ke kotak komunikasi RS-485 di dalam kabinet. |
| ⑬ | Kabel pemicu probe fotometri, terhubung ke unit utama ⑧ dari Goniophotometer |
| ⑭ | Kabel komunikasi RT-200C, dengan ujung lainnya terhubung ke unit utama ⑦ dari Goniophotometer |

Gambar 18

Gambar 19

Gambar 20

Gambar 21
4. Instalasi dan Konfigurasi Perangkat Lunak
4.1 Instalasi Perangkat Lunak dan Driver
4.1.1 Lingkungan Eksekusi Perangkat Lunak
Persyaratan Sistem: Windows 7/8/10/11 (32-bit/64-bit); komputer harus memiliki setidaknya satu port USB.
Instalasi 4.1.2
| Jumlah | Note |
| ① | LSG Series Goniophotomoeter Software—Double-click to install |
| ② | File APK ini dapat disalin ke ponsel Android untuk diinstal. Aplikasi ini hanya ditujukan untuk memutar Goniophotomoeter dan tidak dapat digunakan untuk pengujian. Karena perangkat lunak ini mencakup semua fitur aplikasi, manual ini menggunakan perangkat lunak sebagai contoh pengoperasian. |
| ③ | RS485 Communication Box Driver—Please double-click to install |

Gambar 27
4.1.3 Penanganan Pengecualian
If your antivirus software or the Windows Firewall flags a file as suspicious during the software installation process, please temporarily disable the antivirus software or the Windows Firewall and try installing the software again.
If the software fails to open when you double-click its icon after a successful installation, please check whether you are logged in to your computer with an administrator account. If the software still does not open after logging in with an administrator account, right-click the software icon to open its Properties window. On the Compatibility tab, check the box next to “Run this program as an administrator,” click “Apply,” and then double-click the icon again to open the software.

Gambar 28
4.2 Pengaturan Perangkat Lunak
First, turn on the cabinet’s power switch, then power on all system devices, including all equipment inside the cabinet, the Goniophotomoeter main unit, and the photometric probes. Double-click to launch the software. The first time you run the software, it will automatically open the system configuration interface. Select the correct model for each device you have purchased.
| Jumlah | Note |
| ① | To select a goniophotomoeter model, click the drop-down list on the right. The model you purchased is the LSG-1890B. |
| ② | Probe fotometer medan dekat; pilih model probe fotometer PM400C |
| ③ | Far-field photometric probe—if you have not purchased one, please select “Not Installed.” |
| ④ | The LSG-1890B does not include a near-field Spectroradiometer; please select “Not Installed.” |
| ⑤ | AC power supply models are generally from the LSP Series |
| ⑥ | Model catu daya DC umumnya disebut sebagai Seri DC. |
| ⑦ | Model meteran daya: Nomor model tercetak di panel depan meteran daya. Silakan periksa dan pilih model yang sesuai. |
| ⑧ | Once you have selected all the correct models, click “Auto-detect Port.” |
| ⑨ | Setelah pencarian semua port selesai, klik “OK” untuk menyimpan dan keluar. |

Gambar 29
Note: If a communication error occurs during the automatic port search, please read the error message carefully, check whether the corresponding device is powered on, and verify that the RS485 communication cable is properly connected. Once you have confirmed these points, click “Automatic Port Search” again until communication is established with all devices.
After clicking OK, the software automatically opens the database file save dialog. From then on, all test reports will be automatically saved in this database file.
| Jumlah | Note |
| ① | Please select the path where you want to save the database file. Please do not save it to the desktop or inside the Desktop folder. |
| ② | Masukkan nama file |
| ③ | Just click “Save” |

Gambar 30
5. Calibration of darkroom equipment (Each step must be performed carefully and thoroughly; otherwise, calibration data may be inaccurate and test results may be abnormal.)
5.1 Horizontal Commissioning of the Main Unit
5.1.1 Laser Ready
There is a laser switch on the back of the goniophotomoeter’s main arm. When you turn on the line laser switch, you will see a red line laser beam emitted directly in front of the main arm.

Gambar 31
Find the crosshair laser device and tripod.

Gambar 32
Remove the cross-line laser and the batteries. Insert the batteries into the cross-line laser. The power switch is located on the side of the device. The device turns on when the switch is set to the “unlocked” position.
| Jumlah | Note |
| ① | Power switch: Turn it to the “Unlock” position to turn on the power. |
| ② | VH refers to vertical and horizontal laser switches, respectively. |

Gambar 33
Mount the crosshair laser onto the tripod.
Note: The cross-line laser is self-leveling. If the cross-line laser is not self-leveled, an audible alarm will sound. This may be due to an uneven surface or a tripod that is not properly set up. Please ensure that the cross-line laser remains self-leveled at all times during use.
5.1.2 Steps for Horizontal Commissioning of the Host Machine
According to the darkroom drawings, the center height of the LSG-1890B goniophotomoeter is 1510 mm. Using a tripod, adjust the height of the horizontal laser beam from the crosshair laser to 1510 mm. At this height, the horizontal laser beam should pass through the center of all light-transmitting holes in the darkroom partition.

Gambar 34
Please note: Do not place any objects on the Goniophotomoeter or within its rotation range to prevent damage to the equipment during rotation.
Buka perangkat lunak.
| Jumlah | Note |
| ① | Click to switch angle modes |
| ② | Click ① to display the C-Gamma angle here instead of the B-Beta angle. |
| ③ | Kemudian klik untuk membuka antarmuka kontrol sudut. |

Gambar 35
Click the continuous rotation button for any direction. The Gamma axis of the Goniophotomoeter will then begin to rotate continuously.

Gambar 36
At this point, observe the horizontal laser beams emitted by the single-line and cross-line lasers from the main axis of the Goniophotomoeter.

Gambar 37
Then, using an appropriate wrench, raise the goniophotomoeter and level it by adjusting the height of its four legs. The final result of the calibration: Regardless of the Gamma angle of the goniophotomoeter, its straight laser beam perfectly aligns with the horizontal laser beam, as shown in the figure below.

Gambar 38
You’ll need plenty of patience to complete this step. Once adjusted, make sure all four legs of the main unit are firmly on the ground. Then click the “Stop” button on the software’s angle control interface to stop the Goniophotomoeter from rotating. Next, lock the four legs of the Goniophotomoeter in place using the screws at the top. Once the position and level of the Goniophotomoeter main unit have been properly adjusted, do not move it again, as this will cause deviations in the test results.
5.2 Kalibrasi Probe Fotometri
5.2.1 Calibration of the Photometric Probe at the 0-Degree Position
| Jumlah | Note |
| ① | Masukkan 0 |
| ② | Click any rotation direction to rotate the Gamma axis back to the 0-degree position. |

Gambar 39
Verify that the single-line laser beam is passing through the exact horizontal center of the last partition at this point.

Gambar 40
If not, follow the steps below.
| Jumlah | Note |
| ① | Please enter a small angle value here based on the actual situation, such as 1, 0.1, 0.05, etc. |
| ② | Select a rotation direction and rotate the Gamma axis until the single-line laser passes through the exact center of the last partition horizontally. |
| ③ | Click “Set to 0 Degrees” to set the current position of the Gamma axis on the near-field luminance probe to 0 degrees. The password is LISUNGROUP. |

Gambar 40.1
Note 1: After setting the Gamma axis of the near-field photometer to 0 degrees, do not arbitrarily set it to 0 degrees again, as this will cause deviations in the test results.
Note 2: Definition of the Gamma angle for a goniophotomoeter. The direction directly facing the photometer probe is defined as 0 degrees/360 degrees. When rotating clockwise, the angle value increases from 0 degrees to 360 degrees; when rotating counterclockwise, the angle value decreases from 360 degrees to 0 degrees. Since angle notation may vary, -90 degrees and 270 degrees correspond to the same position.
5.2.2 Mounting the Photometric Probe
Rotate the light sensor’s lens cover counterclockwise and remove it.

Gambar 41
Kotak kemasan lampu standar berisi cermin kecil.

Gambar 42

Gambar 43
Remove the small reflector and attach it to the probe.

Gambar 44
At this point, the Gamma axis of the near-field photometer is still at the 0-degree position. Slightly move the probe base and adjust the screws at points ① and ② to adjust the probe’s height and orientation. Position the probe so that the straight laser beam coming from the main axis hits the exact center of the mirror, and ensure that the reflected beam travels back along the same path.

Gambar 45
Kemudian, kencangkan alas braket probe ke lantai dengan sekrup.

Gambar 46
Once secured, you may need to make slight adjustments to the probe’s height and orientation so that the straight laser beam continues to hit the exact center of the reflector and the reflected beam returns along the same path. After calibration is complete, ensure that all screws are tightened. Then remove the reflector and reinstall the light probe’s light shield.
Note: Once the probe’s position and orientation have been set, do not adjust them, as doing so may cause the test data to be inaccurate.
You can secure the photometer probe’s power adapter to the probe mount as shown in the figure below to protect the probe’s power jack.

Gambar 47
5.3 Cross-Line Laser Installation
Finally, the crosshair laser must be mounted on the side wall of the darkroom. Depending on the specific layout of each darkroom, the crosshair laser can be mounted on either the left or right wall. Take the darkroom layout shown in the figure below as an example.

Gambar 52
Go to the angle control screen.
| Jumlah | Note |
| ① | Enter 90 (if the installation position of the crosshair laser in your darkroom blueprint is different, enter -90) |
| ② | Click to rotate the gamma axis to a 90-degree position, so that the main-axis laser is directly aligned with the position where the crosshair laser is installed. |

Gambar 53
Secure the crosshair laser to the triangular mount we provide.

Gambar 54
Turn on the crosshair laser and use the direction of the single-line laser to roughly determine the position of the crosshair laser.

Gambar 55
Fine-Tune the Position of the Crosshair Laser: The centerline of the straight laser should be perfectly aligned with the centerline of the crosshair laser. You can verify and adjust the position using the following method. Place a sheet of white paper at the exit of the cross-shaped laser and at the exit of the single-line laser. If the centerline of the single-line laser aligns with the centerline of the cross-shaped laser at both locations, you can confirm that the position and orientation of the cross-shaped laser are correct.

Gambar 56

Gambar 57
Next, secure the cross-laser tripod to the wall using screws. Once secured, you can still loosen the screws if necessary to make slight adjustments to the cross-laser in both the horizontal and vertical directions, ensuring it remains properly aligned. After calibration is complete, all screws should be tightened securely.

Gambar 58
Then rotate the Gamma axis back to the 0-degree position. The cross-shaped laser beam is projected onto the opposite wall. You can use this to create a permanent mark. That is, when the cross-shaped laser beam is projected in this direction, it indicates the horizontal and vertical center positions of the Goniophotomoeter. This mark is needed for both calibration and testing.

Gambar 59
Installation and commissioning are complete. Turn off the line laser on the main shaft of the Goniophotomoeter.
5.4 Membersihkan
You may shut down all equipment. When doing so, first turn off the power switch on each individual device, and then turn off the main power switch for the cabinet. Next, thoroughly clean the darkroom and control room. To prevent dust from entering the equipment, cover the devices with dust-proof cloths before cleaning. After cleaning, turn all equipment back on and continue with the calibration and testing procedures.
Catatan: Jaga agar ruang gelap sebisa mungkin bebas debu selama penggunaan.
6. Kalibrasi
6.1 Pemasangan dan Pengoperasian Lampu Standar
6.1.1 Pemasangan Lampu Standar
First, turn on the crosshair laser on the wall and rotate it to ensure that it is aimed at the mark on the opposite wall.
Open the Standard Lamp box.

Gambar 60
Then assemble the standard lamp bracket as shown in the figure below.

Gambar 61
The main shaft of the Goniophotomoeter should still be aligned with the photometric probe, i.e., the probe’s 0-degree position. Locate the calibration rod and install it at the center of the Goniophotomoeter. Then remove the top section.

Gambar 62
Please follow the instructions below to connect the wires.
| Jumlah | Note |
| ① | The main shaft of the goniophotomoeter has four terminals for connecting the power cables and sampling cables of the standard lamp and sample luminaire. Use short wires to bridge the red and yellow terminals together, and the black and blue terminals together. |
| ② | Connect the power cord of the standard lamp/sample fixture to the red and blue terminals on the main shaft. |

Gambar 63
6.1.2 Calibration of the Standard Light Direction
| Jumlah | Note |
| ① | Loosen the screw here to adjust the height of the standard lamp. |
| ② | There are three screws here that allow you to adjust the pitch of the standard lamp. |

Gambar 64
Lampu standar harus disesuaikan sehingga berkas laser berbentuk salib diarahkan ke tengah cermin kecil, dan berkas laser horizontal dan vertikal yang dipantulkan dari cermin bertepatan persis dengan berkas laser berbentuk salib aslinya.

Gambar 65
Based on our experience, follow these steps for calibration: First, adjust the height of the standard lamp to the appropriate position. The vertical reflected laser can be calibrated by rotating the calibration target, while the horizontal reflected laser can be calibrated by adjusting the pitch of the standard lamp. After calibration, the center height of the standard lamp may change, requiring you to readjust its height. Continue the reflection adjustment until the cross-shaped laser beam is projected onto the center of the small mirror, and the reflected horizontal and vertical laser beams perfectly align with the original cross-shaped laser beam.
Once the standard lamp is properly aligned, carefully remove the small reflector, taking care not to alter the alignment of the standard lamp during removal. Then turn off the cross-line laser.
6.2 Kalibrasi Probe Fotometri
All standard lamps are constant-current DC light sources. Set the cabinet’s AC/DC switch to DC.

Gambar 66
Buka pengontrol sudut perangkat lunak.
| Jumlah | Note |
| ① | Enter -90 |
| ② | Click to rotate the Gamma axis to the -90-degree position (if the crosshair laser is installed in a different position on your darkroom drawing, you will need to rotate it to the 90-degree position). |

Gambar 66.1
Align the round hole in the standard lamp mount so that it faces the photometric probe.

Gambar 67
Turn on the calibration lamp and enter the calibration interface.
| Jumlah | Note |
| ① | Select a DC Power Supply |
| ② | The standard lamp is a constant-current light source. Select CC (constant-current mode) and enter the reference voltage and rated current as specified in the standard lamp certificate; typically, these are 30 V and 6.1 A. |
| ③ | Click to turn on the standard lamp and turn off all other lights in the darkroom. The standard lamp takes 15 minutes to reach a steady state. |
| ④ | Setelah lampu standar stabil, klik untuk masuk ke antarmuka kalibrasi fotometer. |

Gambar 68
Close the darkroom door and make sure that all light sources in the darkroom are turned off, except for the standard lamp.
| Jumlah | Note |
| ① | Memilih Probe Fotometri Medan Dekat |
| ② | Luminous flux calibration—our software includes this option, but current standard practices require the use of luminous intensity calibration, and the standard lamps included in our standard configuration are also luminous intensity standard lamps. Therefore, there is no need to use luminous flux calibration at this time. |
| ③ | Metode Kalibrasi Intensitas Cahaya |
| ④ | Rujuklah pada sertifikat kalibrasi lampu standar dan masukkan nilai intensitas cahaya yang telah dikalibrasi. |
| ⑤ | Klik untuk Memulai |
| ⑥ | Perangkat lunak akan secara otomatis menghitung jarak uji. |
| ⑦ | To verify the results, click “Check,” and the software will display the current measured light intensity value. The measured light intensity value should match the calibrated light intensity value. |
| ⑧ | Klik “OK” untuk menyimpan dan keluar. |

Gambar 69
Note: The test distance is the distance from the center of the Goniophotomoeter to the surface of the photometric probe. The distance calculated by the software should be close to the actual distance; an error of 1% or less is considered normal. If the error is too large, please refer to Section 2.3.1 to verify that the darkroom is functioning properly, and refer to Chapter 5 to review the calibration steps.
After calibration is complete, you can turn off the standard lamp using the software. Since the standard lamp becomes quite hot during use, wait for it to cool down before placing it back in its packaging for safe storage. Remove the calibration rod and store it safely.
6.3 Kalibrasi dan Pemeliharaan
| Tidak ada perubahan di ruang gelap. | If there have been no changes to the computer, recalibration is not necessary. |
| No changes were made to the computer; simply unplug and replug the communication cable or reinstall the Goniophotomoeter software. No recalibration is necessary. | |
| Jika Anda beralih ke komputer baru atau menginstal ulang sistem operasi, tidak perlu melakukan kalibrasi ulang; cukup masukkan jarak uji sebelumnya dan simpan. | |
| Perubahan di Ruang Gelap | The structure of the darkroom remains unchanged; however, after prolonged use, the interior paint may fade or peel. Depending on how well you maintain the darkroom and the level of accuracy you require for your test results, you may repaint the darkroom periodically. After repainting, the darkroom must be recalibrated. |
| Terdapat perubahan pada konfigurasi ruang gelap, sehingga diperlukan kalibrasi ulang. |
7. Persiapan Sebelum Pengujian
7.1 Pemilihan dan Penuaan Lampu
Pilih luminer representatif untuk pengujian yang sesuai dengan standar nasional atau perusahaan; luminer baru harus menjalani perawatan penuaan (untuk luminer LED, periode penuaan 24 jam disarankan);
Periksa apakah perlengkapan lampu tidak memiliki kerusakan yang terlihat, apakah kabel internal terhubung dengan benar, dan bersihkan noda apa pun dari permukaan perlengkapan lampu.
7.2 Introduction to Test Mode
Mode C-Gamma: Cocok untuk sebagian besar perlengkapan pencahayaan, seperti lampu jalan, lampu downlight, lampu panel, dan lampu pertumbuhan tanaman.

Gambar 74
Mode Beta: Cocok untuk lampu sorot, lampu banjir, dan perlengkapan serupa; memerlukan pemasangan lengan tambahan.

Gambar 75
Mode Alpha: Cocok untuk lampu otomotif dan lampu sein.

Gambar 76
Catatan: LSG-1890BCCD can perform C-Gamma and B-Beta tests. For A-Alpha testing, a Type A goniophotomoeter is required; please contact us if needed.
8. Tes C-Gamma
8.1 Memasang dan Menyalakan Lampu
8.1.1 Pemasangan Perlengkapan Lampu
Kami memiliki dua alat uji C-Gamma; prosedur ini menggunakan alat uji berbentuk salib sebagai contoh.
| Jumlah | Note |
| ① | Large cross-shaped clamps are used for various types of lighting fixtures |
| ② | E27 Fixture for Testing Globe Bulbs |

Gambar 77
Attach the fixture to the Goniophotomoeter. Turn on the wall-mounted cross-line laser and verify that the cross-line laser is projected accurately onto the markings made on the opposite wall.

Gambar 78
On the angle control interface, select the near-field photometric probe. Rotate the Gamma axis 90 degrees so that the crosshair laser points directly at the center of the crosshair fixture. Mount the lamp under test onto the crosshair fixture. Adjust the lamp’s position so that the crosshair laser is centered on the light-emitting surface of the lamp under test, and the light-emitting surface is parallel to the crosshair fixture.

Gambar 79
Then rotate the Gamma axis back to 0 degrees. Turn the dial here.

Gambar 80
Align the light-emitting surface of the fixture with the vertical laser beam of the crosshair.

Gambar 81
Caution is required when installing directional luminaires. When the angle of the C-plane on the Goniophotomoeter is 0 degrees, according to streetlight installation standards, the light pole should point directly upward, as shown in the figure below.

Gambar 81.1
When the angle of the C-plane on the goniophotomoeter is 0 degrees, the strip light should be installed vertically in accordance with industry standards, as shown in the figure below.

Gambar 81.2
8.1.1 Menyalakan Lampu
Connect the power cord to the luminaire under test. The connection method for the luminaire under test is the same as that for the standard luminaire; refer to Section 6.1.1.
Sesuaikan sakelar kabinet sesuai dengan parameter luminer yang sedang diuji.
| Jumlah | Note |
| ① | If the sample luminaire is AC-powered, switch to the AC setting; if it is DC-powered, switch to the DC setting. |
| ② | Internal/External Sampling Switch for Electrical Parameters. External sampling uses a four-wire configuration and provides more accurate voltage readings. Unless otherwise specified, please set this switch to “External Sampling”; no further adjustments are necessary. |

Gambar 82
Jika lampu sampel menggunakan daya AC.
| Jumlah | Note |
| ① | Select AC Power |
| ② | Based on the sample parameters, enter the correct voltage and frequency. |

Gambar 83
Jika lampu sampel menggunakan daya DC.
| Jumlah | Note |
| ① | Select a DC Power Supply |
| ② | Berdasarkan parameter sampel, masukkan tegangan dan arus yang benar. |
| ③ | Select the output mode: CC stands for constant current, and CV stands for constant voltage. If you select constant current, you need to enter the sample’s rated current value; for the voltage, simply enter the maximum output voltage of this DC power supply. If you select constant voltage, enter the sample’s rated voltage and the DC power supply’s maximum output current. Catatan: Jika model catu daya DC Anda adalah DC3010, its maximum output voltage is 30V and its maximum output current is 10A. |

Gambar 84
The software offers two methods for turning the sample lamp on and off, and users can choose between them.
| Jumlah | Note |
| ① | If you check the “Auto Turn On” and “Auto Turn Off” options, the software will automatically power on the sample according to the selected power source and the entered electrical parameters when the test begins, and will automatically cut off the sample’s power when the test ends. |
| ② | If the “Auto On” and “Auto Off” options are not checked, you can still click this button; the software will illuminate the sample based on the selected power source and the entered electrical parameters. Clicking this button again will cut off the power to the sample. |

Gambar 85
8.2 Pengaturan Parameter Pemanasan Awal
| Jumlah | Note |
| ① | If you check the “Preheat” box, the software will preheat the luminaires according to the specified preheat parameters when the test begins, and the test will start automatically once preheating is complete. If you do not check the “Preheat” box, the software will start the test immediately. |
| ② | Pengaturan Parameter Pemanasan Awal |

Gambar 86
Click “Preheat Parameter Settings.”
| Jumlah | Note |
| ① | Pengaturan Durasi Pemanasan: Tujuan dari pemanasan adalah untuk memungkinkan perlengkapan lampu tetap menyala selama jangka waktu tertentu hingga parameter cahaya dan warnanya stabil, setelah itu pengujian akan dilanjutkan secara otomatis. |
| ② | Sampling interval: The software automatically samples data at the set interval throughout the entire warm-up period to obtain real-time data. We recommend setting this to 5 seconds or longer. |
| ③ | If checked, the test will automatically skip the remaining warm-up time and proceed once the light fixture has reached a stable state; if unchecked, the test will begin only after the warm-up period has been completed according to the set duration. |
| ④ | Definition of a luminaire’s steady-state condition. If the values are set to 20 and 0.5%, the luminaire is considered to be in a steady state if the parameter has not varied by more than 0.5% over the past 20 minutes. |
| ⑤ | Anda dapat memilih parameter mana yang harus digunakan perangkat lunak untuk menentukan apakah perlengkapan lampu stabil: parameter optik, parameter listrik. |
| ⑥ | Perangkat lunak akan menyimpan kurva perubahan untuk parameter yang dipilih selama periode pemanasan dalam laporan pengujian. |
| ⑦ | Setelah Anda selesai mengkonfigurasi pengaturan, klik OK untuk menyimpan dan keluar. |

Gambar 87
8.3 Mengonfigurasi Parameter Pengujian dan Memulai Pengujian
Klik “Mulai Tes.”

Gambar 88
Buka layar pengaturan parameter pengujian.
| Jumlah | Note |
| ① | Jenis perlengkapan pencahayaan: lampu dalam ruangan, lampu luar ruangan, atau lampu pertumbuhan tanaman—ketiga jenis perlengkapan pencahayaan ini tunduk pada pengujian C-Gamma. |
| ② | Select whether the luminaire is symmetrical or asymmetrical. Unless there are specific requirements, select the first option (asymmetrical) regardless of whether the luminaire being tested is symmetrical or not (to obtain the luminaire’s actual data). |
| ③ | Rentang sudut bidang C: Pilih 0 hingga 180 derajat; tidak perlu modifikasi. Untuk data yang berkisar dari 180 hingga 360 derajat, perangkat lunak akan secara otomatis mengonversinya. C-plane angle intervals. Symmetrical luminaires, such as indoor lights, can use a larger interval (e.g., 30 degrees), while asymmetrical luminaires, such as streetlights, can use a smaller interval (e.g., 10 degrees). The smaller the angle interval, the higher the test accuracy, but the longer the test will take. If time permits, we recommend selecting a smaller angle interval. |
| ④ | Gamma plane angle range: If the sample’s light irradiation angle exceeds 180 degrees (e.g., incandescent bulbs), select -180 degrees to 180 degrees; if the sample’s light irradiation angle is less than 180 degrees (e.g., panel lights), you may select a Gamma plane angle interval ranging from -90 degrees to 90 degrees. Unless there are special requirements, a 1-degree interval is sufficient. The specific angle interval selected here has no effect on test speed. |
| ⑤ | If the system has multiple photometric probes, please select the one you want to use. |
| ⑥ | For standard luminaires undergoing standard testing, select “Normal Speed”; if the luminaire being tested has an extremely narrow beam angle, select “Slow Speed” to obtain more accurate test results; for expedited testing, select “Fast Speed.” |
| ⑦ | Fixed Range: If you are consistently measuring the same type of light fixture and the luminous intensity data does not vary significantly, you can check this box to save time. If you have ample time, or if you frequently test lamps with different luminous intensities, we recommend leaving this box unchecked; in this case, the photometric probe will automatically select the appropriate range for each lamp and enter continuous scanning mode. This is suitable when the angular interval of the Gamma plane is between -180 degrees and 180 degrees. During the test, when the Gamma angle approaches ±180 degrees, simultaneously rotating the C-axis to the next scanning position—that is, the Gamma axis rotates continuously without stopping to save time. If you have ample time, we recommend leaving this option unchecked. |
| ⑧ | To set the monitoring angle during lamp warm-up, simply set it to (0, 0). |
| ⑨ | Normalisasi mengacu pada keputusan apakah akan menerapkan nilai seragam pada baris pertama data dalam tabel intensitas cahaya pada laporan pengujian; standar tertentu, seperti standar lampu jalan AS, mungkin mengharuskan Anda untuk melakukan penyesuaian serupa. Ada tiga pilihan: tanpa normalisasi, normalisasi ke nilai maksimum (yang akan sedikit meningkatkan fluks cahaya total luminer), dan normalisasi ke nilai rata-rata (yang pada dasarnya tidak mengubah fluks cahaya). |
| ⑩ | If you need to remove all stray light outside a specific angle, check this box and enter the desired angle . For example, if you check this option and enter 60, the data outside ±60 degrees of the luminaire’s gamma plane will automatically be set to 0. |
| ⑪ | Enter the number of light sources in the luminaire and the rated luminous flux of each light source. If you do not know these values, you may leave them blank; this will not affect the test results. |
| 1 | Masukkan dimensi permukaan pemancar cahaya sampel. Untuk perlengkapan pencahayaan dalam ruangan, perangkat lunak memerlukan data ini untuk menghitung kurva batas luminansi, UGR, dan parameter lainnya. |
| ⑬ | Enter information about the sample light fixture |
| ⑭ | Enter the temperature and humidity of the test environment (i.e., the darkroom) |
| 3 | Klik OK untuk menuju layar pemanasan, atau menuju layar pengujian. |

Gambar 89
Then wait for the software testing to finish.
9. B-Beta Testing
9.1 Install and Turn On the Lights
9.1.1 Lighting Fixture Installation
Turn on the crosshair laser and make sure it is still shining on the mark you made on the opposite wall.
Find the square B-Beta test fixture.

Gambar 92
When performing the B-Beta test, reattach the Goniophotomoeter’s auxiliary arm. Then, install the B-Beta fixture vertically as shown in the figure.

Gambar 93
| Jumlah | Note |
| ① | Click to switch to the angle display mode |
| ② | Klik ① untuk menampilkan sudut B-Beta di sini, bukan sudut C-Gamma. |
| ③ | Klik untuk membuka antarmuka kontrol sudut. |

Gambar 94
| Jumlah | Note |
| ① | When angle B is 0 degrees, the fixture should be in a vertical position; this can be verified using a level. |
| ② | Masukkan 0 |
| ③ | Click to set the Beta angle to 0 degrees. Note: The 0-degree position for the Gamma angle is not the same as the 0-degree position for the Beta angle. |

Gambar 95
After rotating Beta to the 0-degree position, mount the floodlight onto the fixture, ensuring that the light-emitting surface of the floodlight faces the photometric probe. Adjust the mounting position of the fixture so that the cross-shaped vertical laser beam aligns with the light-emitting surface of the floodlight.

Gambar 96
To ensure the sample is centered in the Goniophotomoeter, rotate the Beta fixture to the position shown in the figure below so that the sample’s light-emitting surface faces the crosshairs. Use the crosshairs to fine-tune the position of the sample holder.

Gambar 97
Once the light fixture is installed, you can turn off the crosshair laser.
9.1.2 Menyalakan Lampu
Please refer to 8.1.2.
9.2 Pengaturan Parameter Pemanasan Awal
Please refer to Section 8.2.
9.3 Mengonfigurasi Parameter Pengujian dan Memulai Pengujian
Klik “Mulai Pengujian” untuk menuju ke layar pengaturan parameter pengujian.
| Jumlah | Note |
| ① | Silakan pilih “Floodlight,” dan mode pengujian akan secara otomatis beralih ke B-Beta. |
| ② | Rentang sudut untuk Bidang B adalah -90 derajat hingga 90 derajat. Tidak perlu melakukan penyesuaian. the angle intervals for Plane B; you can select smaller angle intervals, such as 5 degrees or 10 degrees. The smaller the interval, the higher the test accuracy, but the longer it takes. |
| ③ | The Beta plane angle range is -90 degrees to 90 degrees. There is no need to adjust the Beta plane interval angle; unless there are special requirements, select 1 degree. Different interval angles here have no effect on test speed. |

Gambar 98
For other parameter settings, refer to Section 8.1.3. Once you have configured the settings, click “Start” and wait for the software to complete the test.
10. Pemrosesan Laporan Tes
10.1 Organizing Database Files
| Jumlah | Note |
| ① | By default, all test reports are automatically stored in the currently open database file in the order in which the tests were run. |
| ② | Untuk menghapus laporan, pilih laporan tersebut lalu klik “Hapus.” |
| ③ | To make it easier to find test reports, click the blank button shown in the image below to create a new database file, and store reports for different lighting fixtures in separate database files. After creating a new database file, you can proceed directly with testing without having to recalibrate or reconfigure the system. |
| ④ | Click “Open” to view test reports from other database files. |

Gambar 100
Note 1: The next time the software is launched, it will automatically load the database file used during the previous session. If the path to that database file has changed or its name has been modified, the software will prompt you to create a new database file.
Catatan 2: Sebisa mungkin hindari menempatkan file basis data di desktop komputer Anda.
10.2 Pengaturan Tampilan Laporan Pengujian
| Jumlah | Note |
| ① | You can double-click to upload a picture of the light fixture. |
| ② | Anda dapat mengklik dua kali konten yang ditampilkan di area ini untuk mengeditnya. |

Gambar 101
| Jumlah | Note |
| ① | Klik "Pengaturan" |
| ② | For unit selection, “cd” is generally used; however, some older standards—such as those regarding energy-saving light bulbs—may require the use of “cd/klm.” |
| ③ | Streetlight C0 angle setting. The CIE/GB and IESNA standards have inconsistent requirements; please select the standard you are following. |
| ④ | Customize the divergence angle. The software automatically displays the half-peak angle (commonly referred to as the beam angle, calculated as 50% of the maximum luminous intensity) and the effective beam angle (calculated as 10% of the maximum luminous intensity). If you wish to customize the divergence angle, enter the percentage of the maximum light intensity and check the box (e.g., 0.25 for 25%). In this case, the test report will display not only the half-peak angle and effective beam angle but also your custom divergence angle. |
| ⑤ | Centang kotak ini untuk menampilkan sudut kerucut intensitas cahaya maksimum dalam laporan. |

Gambar 102
| Jumlah | Note |
| ① | Switch to the “Energy Efficiency Rating” settings screen, then select the applicable standard and luminaire type. |
| ② | Klik untuk menerapkan pada laporan tes saat ini |
| ③ | Klik OK untuk menyimpan dan keluar. |

Gambar 103
Each display window has several settings that allow you to customize the display. Users can experiment with adjusting these settings to achieve their desired display effects.

Gambar 103.1
10.3 Memproses Data Uji
10.3.1 Double-sided luminaires
For lamps that emit light from both sides, use the indoor lamp test mode to test each side individually—that is, set the gamma plane angle range to -90 degrees to 90 degrees during testing—and then generate two test reports.
| Jumlah | Note |
| ① | Tekan dan tahan tombol “Ctrl” pada keyboard komputer Anda, lalu klik untuk memilih dua laporan pengujian yang baru saja Anda terima. |
| ② | Klik “Sintesis Data” di bilah navigasi perangkat lunak. |
| ③ | Di jendela pop-up, Anda dapat mengklik untuk beralih antara pancaran cahaya ke atas dan ke bawah. |
| ④ | Pilih “Komposit” |
| ⑤ | Klik OK untuk menyimpan dan keluar. |

Gambar 104
Ini akan menghasilkan laporan uji sintetis baru, seperti yang ditunjukkan pada gambar di bawah ini.

Gambar 105
10.3.2 Conversion of Test Report Types
| Jumlah | Note |
| ① | Click “Type Conversion” |
| ② | Laporan pengujian dapat dikonversi di antara empat jenis. |
| ③ | When converting to a C-Gamma test report, you can use the “termination angle” of the C-plane to generate a symmetric test report. If the termination angle is set to 360 degrees, there is no symmetry; if the termination angle is set to 180 degrees, the report can be converted to an axially symmetric test report; if the termination angle is set to 90 degrees, the report can be converted to a quadrant-symmetric test report; if the termination angle is set to 0 degrees, the report can be converted to a fully symmetric test report. This feature is primarily intended for symmetric indoor lighting fixtures; certain lighting design software programs, such as DIALUX, may require you to use this setting when importing fully symmetric IES files. |

Gambar 106
10.4 Mengekspor Data Uji dan Mencetak Laporan Uji
10.4.1 Mengekspor Data Uji
| Jumlah | Note |
| ① | Klik logo merah di pojok kiri atas perangkat lunak. |
| ② | Click “Export Data” in the drop-down menu |

Gambar 107
File dapat diekspor dalam berbagai format standar.

Gambar 108
10.4.2 Mencetak Laporan Uji
Pengaturan pencetakan.
| Jumlah | Note |
| ① | Klik untuk membuka layar “Pengaturan Cetak” |
| ② | Laporan ini menampilkan pilihan bahasa; Anda dapat memilih antara bahasa Mandarin dan bahasa Inggris. |
| ③ | Format Nomor Halaman |
| ④ | Informasi Header Laporan |
| ⑤ | Judul Laporan dalam Bahasa Mandarin dan Inggris |
| ⑥ | General Report Content—Applicable to Reports on All Types of Lighting Fixtures |
| ⑦ | Karena lampu dalam ruangan, lampu jalan, lampu sorot, dan lampu pertumbuhan tanaman memiliki parameter kunci yang berbeda, isi laporan akan bervariasi sesuai dengan itu. Anda dapat memilih item spesifik yang ingin Anda sertakan dalam laporan. |
| ⑧ | Opsi TM-30 |
| ⑨ | CIE color space diagram or SDCM color tolerance diagram—you can choose only one of them. |

Gambar 109
| Jumlah | Note |
| ① | Klik untuk mencetak laporan pengujian yang saat ini dipilih |
| ② | Klik untuk mencetak semua laporan pengujian dalam file basis data saat ini. |

Gambar 110
11. Pemeliharaan dan Perawatan Peralatan Harian
11.1 Perawatan Rutin
Start-up procedure: First, turn on the cabinet’s main power switch; then, turn on the power switches for the equipment inside the cabinet and in the darkroom one by one; finally, launch the software. Shutdown procedure: First, close the software; then, turn off the power switches for the equipment inside the cabinet and in the darkroom one by one; finally, turn off the cabinet’s main power switch.
Keep the darkroom control room clean and dust-free to prevent dust from entering the equipment. If there is surface dust on the equipment, wipe it off with a dry, soft cloth.
Hindari membiarkan ruang gelap terpapar kelembapan tinggi dalam waktu lama; sediakan alat pengering udara jika diperlukan.
11.2 Perawatan Rutin
11.2.1 Membersihkan Probe Fotometri
Jika ruang gelap tidak dapat dijaga agar bebas debu, bersihkan permukaan peka cahaya pada probe fotometer secara teratur (misalnya, sebulan sekali) berdasarkan kondisi aktual untuk mencegah penumpukan debu dan memengaruhi akurasi pengujian.
When cleaning, please use a professional lint-free cloth to remove dust without scratching the device.
11.2.2 Standard Lamp
Lampu standar berlaku selama satu tahun. Jika Anda masih perlu mengkalibrasi ruang gelap setelah satu tahun, kami sarankan untuk membeli lampu standar yang baru.
Perangkat Lunak 11.2.3
Kami akan memperbarui perangkat lunak dari waktu ke waktu untuk menambahkan fitur dan memperbaiki bug. Jika Anda mengalami masalah saat menggunakan perangkat lunak, jangan ragu untuk menghubungi kami kapan saja untuk mendapatkan versi terbaru.
11.3 Penonaktifan dan Pemeliharaan Jangka Panjang
11.3.1 Pembersihan dan Perlindungan Peralatan
Bersihkan semua permukaan peralatan secara menyeluruh, dan tutupi unit utama, probe, dan kabinet dengan kain anti debu; putuskan semua sumber daya listrik.
11.3.2 Perlindungan Lingkungan
Ensure that the darkroom is dry and well-ventilated to prevent equipment from getting damp. If the ambient humidity is high, keep a dehumidifier on hand.
11.3.3 Monthly Power-On
Lakukan uji penyalaan daya sebulan sekali untuk memastikan semua perangkat keras sistem berfungsi dengan baik.
12. Pemecahan Masalah dan Penyelesaian Masalah Umum
12.1 Kesalahan Komunikasi
Cause of the problem: Check whether the communication cable is securely connected and whether it is plugged into the wrong port; verify that the driver is installed correctly; try a different USB port or communication cable to rule out hardware issues.
Solusi: Instal ulang driver, sambungkan kabel komunikasi sesuai dengan pengenal perangkat, mulai ulang perangkat lunak, dan coba lagi.
12.2 Kesalahan Pengujian
12.2.1 Anomali Fluks Cahaya
Cause of the malfunction: Check whether the light source is stable and whether the warm-up time is sufficient; verify that the equipment calibration is valid, that the test distance is correct, and that the standard light is functioning properly; check whether there is any external light interference in the darkroom and whether the light path is obstructed.
Solution: Extend the warm-up time, recalibrate the instrument, turn off other light sources in the darkroom, and remove any obstructions from the optical path.
12.2.2 Anomali UGR
Cause of error: Dimensions of the luminaire’s light-emitting surface were not entered; the luminaire’s beam angle is less than 45°; the calculation standard was selected incorrectly.
Solusi: Masukkan dimensi permukaan pemancar cahaya luminer (panjang/lebar/tinggi) dengan benar; verifikasi bahwa luminer tersebut adalah luminer dalam ruangan (sudut pancaran harus memadai); pilih standar CIE 190:2010 dan hitung ulang.
12.3 Goniophotometer tidak berputar
Cause of the malfunction: Check whether the host power is on and whether the wiring is correct; check whether the angle controller inside the cabinet is operating normally; confirm whether any foreign objects are jammed in the rotating shaft.
Solusi: Periksa catu daya dan kabel, dan singkirkan benda asing apa pun; mulai ulang sistem dan perangkat lunak, lalu coba lagi.
12.4 Software Malfunctions
12.4.1 Kerusakan Perangkat Lunak
Penyebab masalah: Periksa apakah Anda masuk ke sistem operasi komputer dengan akun administrator; verifikasi bahwa perangkat lunak telah terinstal sepenuhnya; periksa apakah file basis data rusak.
Solution: Log in to the computer using an administrator account; reinstall the software and drivers; create a new database file.
12.4.2 Perangkat lunak menampilkan pesan “Berkas bukan basis data”
Penyebab masalah: File basis data telah dihapus atau dipindahkan; program default yang digunakan untuk membuka file LGF telah diubah.
Solution: Create a new database file and save it; restore the default association for LGF files.
12.5 Kegagalan Daya
12.5.1 DC Power Output Abnormalities
Causes of Failure: The cabinet’s AC/DC switch was not set to DC; the sample voltage or current exceeded the DC power supply limits; a short circuit in the sample;
Troubleshooting: Please verify that the sample’s voltage and current are within the DC power supply’s output range. For example, the DC3005 Model ini memiliki tegangan keluaran maksimum 30 V dan arus keluaran maksimum 5 A. Periksa apakah sampel mengalami korsleting; jika ya, perbaiki sampel sebelum melakukan pengujian ulang.
12.5.2 Tidak Ada Output AC
Cause of the fault: The cabinet’s AC/DC switch was not set to the AC position; the overload protection was triggered;
Troubleshooting Steps: Check the AC power output wiring and sample connections to ensure there are no loose connections or short circuits; if the overload protection is triggered, turn off the power, allow the unit to cool for 5 minutes, and then restart it. Ensure that the sample power remains within the AC power output range; for example, the LSP-500VARC memiliki daya keluaran maksimum 500W.

中文简体