Alat Uji Kompresi Saluran Pipa

Pengguna manual

berlaku LISUN Model: LS-B50

1. Pemahaman Dasar tentang Peralatan

1.1 Parameter Utama

1.1.1 Spesifikasi Teknis Peralatan

Parameter Teknis Note
Tekanan Maksimal 5000 N
Akurasi Tekanan ± 1%
Specifications for Pressurized Steel Blocks 50 × 50 mm
Steel Base Dimensions 300 × 150 mm
Rentang pengangkatan 200 mm

1.1.2 Standar yang Berlaku

The LS-B50 Conduits Compression Test Apparatus is designed and manufactured in strict accordance with Clause 10.2 and Figure 1 (IEC 61386-1:2017, Clause 10.2, Figure 1).

1.2 Verifikasi Konfigurasi Peralatan

1.2.1 Membongkar

Saat membuka kemasan perangkat, tangani dengan hati-hati untuk menghindari goresan pada casing dengan alat tajam. Setelah membuka kemasan, periksa terlebih dahulu bagian luar perangkat untuk memastikan tidak ada deformasi atau penyok yang terlihat, dan tombol-tombolnya tidak rusak atau longgar. Jika Anda melihat kelainan pada perangkat atau aksesorinya, jangan menyalakannya. Segera hubungi kami.

1.2.2 Unit Utama dan Aksesoris

Jumlah Note
Unit Utama LS-B50
Kabel Listrik

1.2.3 Mengunduh Dokumen Elektronik

Kami telah mengirimkan tautan unduhan untuk manual pengguna elektronik, sertifikat kalibrasi, kartu garansi, dan dokumen lainnya melalui email pada saat pengiriman. Silakan unduh sesegera mungkin. Jika Anda belum menerima email tersebut, silakan hubungi kami untuk meminta tautan unduhan lagi.

2. Tindakan Pencegahan Keamanan

2.1 Keamanan Listrik

Before connecting the power cord, verify that the supply voltage is AC 220 V ±10% and the frequency is 50 Hz or 60 Hz. You must use a three-prong outlet with a protective ground, and the protective ground resistance must be ≤4 Ω. Never cut or disconnect the protective ground wire, as doing so poses a risk of electric shock.

2.2 Keselamatan Lingkungan

Do not use in environments with flammable or explosive substances, strong electromagnetic interference, high temperatures and humidity, or excessive dust levels. Place the equipment on a level, stable surface, and ensure there is at least 0.5 m of operating space around it. Operating environment: temperature 15°C to 35°C, relative humidity 45% to 75%.

2.3 Prosedur Operasional

During testing, do not place your hands or any other part of your body within the pressure test area to prevent injuries from being crushed; test specimens must be securely fastened to prevent them from falling and causing safety incidents during testing; it is strictly prohibited to disassemble the equipment’s transmission mechanism or dismantle the machine without authorization.

3. Ikhtisar Peralatan

3.1 Diagram Standar

Lihat Gambar 3-1 (diambil dari standar IEC).

Conduits Compression Test Apparatus-Figure1

Gambar-3 1

3.2 Foto Produk

Lihat Gambar 3-2.

Jumlah Note
Saklar daya
Buzzer; a buzzer sounds when the test is complete.
Touchscreen
Pressure block, used to apply pressure to the conduit specimen
Sensor tekanan
Basis Baja

Conduits Compression Test Apparatus-Figure2

Gambar-3 2

4. Pengantar Program Layar Sentuh

4.1 Layar Awal

Lihat Gambar 4-1.

Jumlah Note
Test Interface, Chinese
Test Interface, English
Pressure Calibration Interface, Chinese
Pressure Calibration Interface, English

Conduits Compression Test Apparatus-Figure3

Gambar-4 1

Note: The pressure readings have been calibrated before the device left the factory. If you need to recalibrate the pressure during use and have any questions about operating the calibration interface, please contact us.

4.2 Antarmuka Pengujian

Lihat Gambar 4-2.

Jumlah Note
To preset the load pressure value, click the ⑧ Pressure Value button. Do not set non-standard pressure values on your own.
Set the test duration; according to the standard, it should be set to 60 seconds.
Current Load Pressure Value
Current Test Duration
Number of loading pulses, preset according to different pressure values
Time from 0 to the preset pressure value
Reset, Release Pressure
Pressure values for different catheters—simply click to preset them
Start/Stop Test
Reset: Clear the data from the last test and clear the beep alarm.
Reset the pressure value; set the current force value to 0
1 Up—click to raise the steel block
Drop; click to lower the steel block
Kembali ke Layar Beranda

Conduits Compression Test Apparatus-Figure4

Gambar-4 2

5. Pressure Testing of Electrical Conduits

5.1 Persiapan

5.1.1 Persiapan Spesimen Uji

Cut a test specimen from the conduit with a length of (200±5) mm. Maintain the test environment temperature at (23±2) °C. Before the test, measure the outer diameter of the conduit specimen.

5.1.2 Determination of Pressure Values

Kode Klasifikasi Catheter Types Pressure N (tolerance ±4%)
1 Sangat ringan 125
2 Cahaya 320
3 Cina 750
4 Berat 1250
5 Kegemukan 4000

5.1.3 Positioning of Test Specimens and Equipment

Place the pipe specimen on the steel base, click “Up/Down” to move the steel block toward the pipe specimen. Stop just before contact is made, then click to reset the pressure reading to zero.

5.2 Pressure Test Procedures

5.2.1 Normal Test Procedure

Set the test duration to 60 seconds. Refer to Section 5.1.2, click the button for the desired pressure value to preset the pressure value and loading pulse, and then click “Start” to begin the test.

After the test begins, the steel block slowly applies pressure to the guide tube. When the pressure value begins to change, the timer starts counting the time from 0 until the preset pressure value is reached. When the current pressure value reaches the preset pressure value, the timer stops counting the time from 0 until the preset pressure value is reached, and the timer begins counting the current test duration.

When the current test duration reaches the preset test duration, the device sounds an alarm, but the current pressure value remains at the preset pressure value. According to the standard, after the test is complete, the outer diameter of the specimen at the flattened section must be measured without relieving the pressure. The difference between the original outer diameter and the outer diameter of the specimen after flattening must not exceed 25% of the original outer diameter measured before the test.

Click “Reset” to release the pressure and return the steel block to its initial position. After (60±2) seconds, measure the outer diameter of the flattened portion of the specimen again. The difference between the original outer diameter and the outer diameter of the flattened specimen must not exceed 10% of the outer diameter measured before the test.

After the test, the test specimen must not exhibit any cracks visible to the naked eye or with corrected vision without any additional magnification.

5.2.2 Debugging Load Pulses

Depending on the catheter’s compressive strength, the preset loading pulses may require adjustment. During normal testing, if the timer stops while the load is being applied from 0 to the preset pressure value and the duration is not within the range of (30±3) seconds, the number of loading pulses must be adjusted.

Taking 4000 N as an example, the corresponding preset number of loading pulses is 2700. If the actual loading duration is 25 seconds, which is less than the standard requirement of (30±3) seconds, you can change the number of loading pulses from 2700 to 2600 and try again; If the actual loading duration is 35 seconds—which exceeds the standard requirement of (30 ± 3) seconds—then try changing the number of loading pulses from 2700 to 2800. Continue adjusting until the loading duration meets the standard requirement.

6. Pemeliharaan dan Perawatan Peralatan Harian

6.1 Perawatan Harian Setelah Pemutusan Daya

At the end of each day’s testing, clean dust and specimen debris from the equipment surfaces, and keep the steel blocks and steel bases clean and free of debris; use a special touchscreen cleaning cloth to clean the touchscreen; check the pressure-applied steel blocks for deformation or displacement.

6.2 Perawatan Rutin Mingguan

Apply a small amount of special-purpose lubricating grease to the equipment’s drive lead screw to ensure smooth, jam-free operation.

6.3 Pemeliharaan Selama Penyimpanan Jangka Panjang

After the equipment has been powered off and routine maintenance has been completed, cover the entire unit with a dust cover and store it in a dry, well-ventilated warehouse; power it on for 30 minutes each month to prevent components from being damaged by moisture.

6.4 Persyaratan Interval Kalibrasi

Untuk memastikan akurasi pengukuran dan keandalan operasional peralatan, disarankan agar peralatan tersebut secara berkala dikirim ke laboratorium metrologi pihak ketiga yang berkualifikasi untuk kalibrasi. Interval kalibrasi yang disarankan adalah 12 bulan. Pengguna dapat menentukan interval kalibrasi spesifik berdasarkan frekuensi penggunaan, kondisi lingkungan, dan persyaratan sistem mutu.

7. Pemecahan Masalah dan Penyelesaian Masalah Umum

Gejala Kesalahan Kemungkinan penyebab Prosedur
Touchscreen lags; taps don’t register Screen surface smudges causing interference, program crashes Turn off the device, clean the screen, then restart it and try again.
No pressure reading is displayed for an extended period after the test starts The specimen guide tube was not positioned beneath the steel block, and the steel block was too far from the specimen guide tube. Reposition the specimen guide tube; use the down arrow key to move the steel block closer to the specimen guide tube, then try again.
The load time is not within the (30±3)s range. Since catheters have different compressive strength, the preset loading pulses may need to be adjusted. Follow the instructions in Section 5.2.2 for debugging.
Pressure Drift After the equipment has been in use for a long time, it is important to calibrate the force readings. We recommend having the equipment calibrated by a third-party metrology organization. You can adjust the pressure reading by accessing the calibration screen.