+8618117273997weixin
InggrisInggris
中文简体 中文简体 en English ru Русский es Español pt Português tr Türkçe ar العربية de Deutsch pl Polski it Italiano fr Français ko 한국어 th ไทย vi Tiếng Việt ja 日本語

Perangkat Uji Retensi Pin Soket

Pengguna manual

berlaku LISUN models: GNGPL-3607、LS-B12、LS-B12A、LS-B12B、LS-B12C

1. Pemahaman Dasar tentang Peralatan

1.1 Properti Dasar

This device is intended for specific testing of the retention strength of plugs for household and similar uses.

It is primarily used in electrical product quality inspection laboratories, the quality control departments of plug manufacturers, third-party testing organizations, and for the inspection of products entering or leaving the country.

1.2 Standar yang Berlaku

In strict accordance with Clause 24.11 and Figure 30 of GB/T 2099.1-2021, Clause 24.11 and Figure 33 of IEC 60884-1:2022, Clause 22.3 of GB/T 4706-2024, Clause 22.3 of IEC 60335-1:2016, Clause 12.11.1 and Figure 7 of BS 1363-1, Clause 13.4 of EN 50075, Clause 24.10 and Figure 31 of VDE 0620-2-1, and other relevant standards, and conducted testing.

In accordance with the tensile force values and application duration specified in the standard, a vertical tensile force is applied to the plug and prong test specimens under specific environmental conditions. The deformation and damage to the specimens are then observed to determine whether they meet the requirements of the standard.

1.3 Mengunduh Dokumen Elektronik

Kami telah mengirimkan tautan unduhan untuk manual pengguna elektronik, sertifikat kalibrasi, kartu garansi, dan dokumen lainnya melalui email pada saat pengiriman. Silakan unduh sesegera mungkin. Jika Anda belum menerima email tersebut, silakan hubungi kami untuk meminta tautan unduhan lagi.

2. Tindakan Pencegahan untuk Penggunaan

2.1 Persyaratan Keselamatan Lingkungan Operasi

Letakkan perangkat di atas meja kerja yang rata dan stabil dengan jarak minimal 30 cm di semua sisi, jauh dari sumber getaran.

2.2 Pedoman Keselamatan untuk Operator

Do not reach into the test area during testing to prevent crush injuries.

Samples must be securely positioned, and the weight clamps must firmly hold the pins in place to prevent the weights from falling off or popping out during testing, which could pose a safety hazard.

Conduct testing strictly in accordance with the standards. Do not add weights without authorization to prevent equipment damage and safety incidents.

3. Pendahuluan tentang Pengaturan Pengujian

Lihat Gambar 3-1.

Jumlah Note
Steel plate with test holes; the diameter of the test holes is customized according to the applicable standard.
Weights, which provide a constant vertical test force; the weights are customized according to the corresponding standards.
Weight clamp—used during testing to secure weights and apply tension to the pin
Adjustable base to ensure the unit is placed level

Socket-Outlets Pins Retention Test Device-Figure1

Gambar-3 1

4. Latch Strength Test

4.1 Pemilihan Sampel

Select brand-new, unused finished plugs that have not undergone any destructive testing, and exclude any nonconforming specimens with surface dents, cracks, or obvious deformation.

4.2 Confirmation of Test Standards and Parameters

4.2.1 Basis for the Standard

Note: This test procedure uses Clause 24.11 of IEC 60884-1:2022 as an example. Test procedures may vary slightly across different standards; please strictly follow the applicable standards for the product and the target market when conducting tests.

4.2.2 Parameter Uji

See Figure 4-1 (taken from the IEC standard) for a diagram illustrating the interrelationship between plug pins, test holes, and tensile force P.

Socket-Outlets Pins Retention Test Device-Figure2

Gambar-4 1

The tensile force shall comply with the maximum pull-out force specified in Table 19 of this standard; see Figure 4-2 (taken from the IEC standard).

Socket-Outlets Pins Retention Test Device-Figure3

Gambar-4 2

4.3 Practical Experiment

The plug shall be placed in a High Temperature Chamber maintained at a stable temperature of (70±2) °C for 1 hour. Then, in accordance with 4.2.2, the test specimen shall be securely clamped inside the High Temperature Chamber for testing and shall not be removed from the Chamber.

Apply a tensile force P to each pin sequentially along the pin’s longitudinal axis for 1 minute; no impact is permitted during this process. See Figure 4-3 for a schematic diagram of the application of tensile force P to the LS-B12.

Socket-Outlets Pins Retention Test Device-Figure4

Gambar-4 3

4.4 Interpretasi Hasil Tes

After the test is complete, remove the device and the test specimen plug from the test chamber and allow them to cool to ambient temperature. If it is verified that the displacement of any pin in the plug body does not exceed 1 mm, the test specimen passes this test.

5. Prosedur Pasca-Ujian

5.1 Reset Perangkat

When the device is not in use, do not load it with test specimens or weights, and avoid applying force to it.

5.2 Persiapan Sampel

Pisahkan sampel yang lulus dan gagal dalam pengujian, beri label dengan jelas, dan hindari kebingungan.

5.3 Membersihkan dan Merapikan Lokasi Pengujian

Use a lint-free, soft cloth to wipe away dust and debris from the surfaces of the steel plate, brackets, weights, weight clamps, and other components. Do not rinse the equipment directly with water.

Tidy up the workbench and remove any debris left over from the experiment.

6. Perawatan dan Pemeliharaan Rutin

6.1 Pemeliharaan Reguler

Depending on the storage environment of the device, periodically apply rust-preventive oil to all components of the device—especially the test holes in the steel plates, the weights, and the clamps—to prevent rust.

6.2 Persyaratan Interval Kalibrasi

To ensure the device’s measurement accuracy and operational reliability, it is recommended that it be sent periodically to a qualified third-party metrology organization for calibration. The recommended calibration interval is 12 months. Users may determine the specific calibration interval based on frequency of use, environmental conditions, and quality system requirements.

7. Pemecahan Masalah dan Penyelesaian Masalah Umum

Gejala Kesalahan Kemungkinan penyebab Prosedur
The clamp doesn’t hold securely; the weight falls off The device was not placed level, preventing the clamp from engaging with the latch. If the device is tilted, it will not only affect the test results, but the weights may also fall off during the test; please reinstall the device and level it before retrying.