Ruang Uji Tahan Air
Pengguna manual
berlaku LISUN Model: JL-123456CP-R400,JL-123456CP-R600,JL-123456CP-R800A、JL-1234CP-R200、JL-1234CP-R400、JL-1234CP-R600、JL-1234CP-R800、JL-12CP-1000M、JL-12CP-1200M,JL-12CP-600M、JL-12CP-800M、JL-3456CP-R400,JL-3456CP-R600,JL-3456CP-R800A、JL-34CP-R200,JL-34CP-R400,JL-34CP-R600、JL-34CP-R800、JL-56CP-1000L、JL-56CP-1400L、JL-56CP-200L、JL-56CP-2200LA、JL-56CP-500L、JL-XC
1. Membuka Kemasan dan Mengenal Peralatan Secara Dasar
1.1 Spesifikasi Teknis Utama
The equipment complies with the test requirements for IPX1, IPX2, IPX3, IPX4, IPX5, and IPX6 specified in GB/T 4208-2017, Sections 14.2.1 through 14.2.6, and Figures 3, 4, and 6.
Note 1: A complete Waterproof Test Chamber covers all six waterproof protection levels, from IPX1 to IPX6. The Waterproof Test Chamber you purchased may only include some of these levels. This manual is a general guide; please take this into account when operating the equipment.
Note 2: The appearance of the Waterproof Test Chamber may vary depending on the waterproof rating and dimensions, but the operating procedures remain the same.
1.2 Membuka Kemasan dan Pemeriksaan
1.2.1 Membuka Kemasan dan Pemeriksaan Visual
When unpacking the device, handle it gently to avoid scratching the casing with sharp objects. After unpacking, first inspect the exterior of the device to ensure there are no visible deformations or dents, and that the buttons are not damaged or loose. If you notice any abnormalities with the device or its accessories, do not turn it on. Please contact us immediately.
1.2.2 Unit Utama dan Aksesoris
Test Chamber Main Unit: 1 unit (IPX1, IPX2, IPX3, and IPX4 components are pre-installed in the test chamber).
The IPX5 and IPX6 components are shown in the figure below.

Gambar 1
Other accessories are shown in the figure below. They include a peristaltic tube, spare IPX3/IPX4 stopper needles, a waterproof connector (for samples), spare filter cartridges, a cleaning needle (for unclogging needles), straps (for securing samples), installation tools, and other accessories.

Gambar 2
Download links for the electronic user manual, software, warranty card, and other materials have been sent via email. Please download them as soon as possible. If you missed the email, you may contact us to request the download links again.
2. Tindakan Pencegahan Keamanan
2.1 Keamanan Listrik
2.1.1 Spesifikasi Koneksi Daya
Based on the nameplate, connect the test chamber to the correct input power supply: 220V/380V 50Hz (test chambers without IPX5 or IPX6 ratings require a single-phase 220V input, while those with IPX5 or IPX6 ratings require a three-phase 380V input;60 Hz is available upon request), and ensure the voltage and frequency are correct. Please use a dedicated circuit to avoid sharing the circuit with other high-power equipment. Confirm that all connections are secure and free of looseness.
2.1.2 Persyaratan Pembumian
The test chamber must be reliably grounded, with a grounding resistance of ≤4 Ω. Do not use gas pipes, non-metallic water pipes, or the neutral wire as a substitute for the grounding wire. After grounding is complete, use the Grounding Resistance Tester to measure the ground resistance. Power may only be applied after confirming that the resistance meets the requirements, to prevent electric shock, test chamber malfunctions, or measurement errors caused by poor grounding.
2.1.3 Tindakan Pencegahan Pengoperasian Listrik
Before wiring, inspecting, or maintaining equipment, the main power supply must be disconnected, and operations may only proceed after confirming that there is no residual voltage;
Do not operate the test chamber’s power switch, touchscreen, or instrument buttons with wet hands to avoid the risk of electric shock;
Jangan meletakkan benda asing apa pun (seperti perkakas atau kertas) di dalam kabinet kontrol listrik untuk mencegah debu atau benda asing menyebabkan korsleting atau merusak komponen;
Jangan mencolokkan atau mencabut kabel daya atau kabel sinyal saat peralatan sedang beroperasi untuk mencegah korsleting yang dapat merusak pengontrol atau komponen listrik.
2.2 Keselamatan Lingkungan
Sufficient clearance must be maintained between the equipment and walls, as well as surrounding fixtures (such as other test equipment and shelving), to ensure proper operation of the equipment and to facilitate operation and maintenance;
The test chamber should be placed on a solid, level surface; its adjustable feet can be used to ensure that the test chamber remains level.
2.3 Lainnya
Untuk hal-hal yang tidak tercakup dalam panduan ini, harap berhati-hati atau hubungi kami.
3. Test Chamber Assembly and Wiring
3.1 Assembly of IPX5 and IPX6 Components
Please refer to the illustration below for assembly instructions.

Gambar 3
IPX5 and IPX6 water outlets.

Gambar 4
The nozzle direction is adjustable.

Gambar 5
Note: After assembly, all four swivel casters on the stand should be in contact with the floor, ensuring the stand is stable and does not wobble.
3.2 Introduction to the Test Chamber Front Panel
| Nomor | Note |
| ① | Test the chamber door; it must remain closed and locked during the test. |
| ② | Touchscreen |
| ③ | Saklar daya |
| ④ | Tombol berhenti darurat |
| ⑤ | IPX1 and IPX2 Flow Meters |
| ⑥ | IPX3 and IPX4 Flow Meters |
| ⑦ | IPX5 Flow Meter |
| ⑧ | IPX6 Flow Meter |

Gambar 6
3.3 Introduction to and Connection of the Test Chamber Back Panel
| Nomor | Note |
| ① Sample Power Supply | If your sample requires power during testing, please connect the appropriate power source here based on your sample’s specific requirements. Short circuits may occur during testing, so please connect an external ground-fault circuit interrupter. |
| ② Filtration System | The filter cartridge must be replaced when it has discolored significantly. (Note: If only IPX5 and IPX6 testing is included, there is no filtration device.) |
| ③ Air Inlet | Used to connect to an air compressor (Note: This is used solely to quickly drain water from the IPX1/IPX2 drip tray; not connecting it to an air compressor does not affect the normal operation of the Waterproof Test Chamber at any waterproof rating.) |
| ④ Inlet | Connect the corrugated hose to a clean water source, and keep the water supply valve open at all times. |
| ⑤ Drain opening | The valve should remain closed at all times; it should only be opened when the tank needs to be cleaned or drained. |
| ⑥ Power cord | Refer to the product nameplate and connect the device to the correct power source. |

Gambar 7
Note 1: Before using the unit for the first time, inspect the test chamber, water tank, and all water lines, remove any debris, and clean them with clean water.
Note 2: We recommend using deionized water or purified water (a water purifier may be used). If tap water is used, the minerals and impurities it contains can easily cause scale buildup, which affects the test water flow rate. This may require frequent replacement of filter cartridges, unclogging of needles, cleaning of solenoid valves, and other maintenance, and may even damage the water pump.
3.4 Sample Connection and Installation
| Nomor | Note |
| ① | If the sample requires testing under power, use the standard waterproof plug and socket for connection. |
| ② | The test specimen must be secured to the center of the turntable using straps (Note: For IPX1 and IPX2 testing, the axis of the test specimen must be 100 mm from the axis of the turntable). If the test specimen has any protruding parts, ensure that it will not collide with the equipment while the turntable is rotating to prevent damage to both the equipment and the test specimen. |
| ③ | The turntable is height-adjustable; for IPX2 testing, the turntable must be tilted 15° and then secured in place. |

Gambar 8
3.5 Menghidupkan Peralatan
Verify that all components are properly connected, close and lock the test chamber door, and turn on the equipment.

Gambar 9

Gambar 10
4. Pengantar Program Layar Sentuh
4.1 Komponen Antarmuka Utama Program
4.1.1 Layar Awal
| Nomor | Note |
| ① | Go to the IPX1 and IPX2 Test Interface |
| ② | Go to the IPX3 and IPX4 Test Interface |
| ③ | Go to the IPX5/IPX6 Test Interface |
| ④ | Debug interface—no need to enter |

Gambar 11
4.1.2 IPX1 and IPX2 User Interfaces
4.1.2.1 Introduction to the Test Interface
| Nomor | Note |
| ① Peringkat ketahanan air: IPX1 | Once selected, you can configure the IPX1 test parameters and perform the IPX1 dripping water test. |
| ② Peringkat ketahanan air: IPX2 | Once selected, you can configure the IPX2 test parameters and perform the IPX2 dripping water test. |
| ③ Add water | Click to fill the water tank with water (Note: The water tank is equipped with upper and lower water level limit switches; if the water level is too low, a low-water alert will appear, and the system will automatically stop filling once the tank is full). |
| ④ Add water to the drip tray | Click to turn the drip tray water pump on or off |
| ⑤ Drainage from the drip tray | Click to turn the air compressor on or off, to drain the water from the drip tray |
| ⑥ Contoh Catu Daya | The sample power supply can be turned on or off. When the sample power supply is turned on for testing, the program will automatically supply power to the sample according to the sample on/off times set in the parameter settings interface. |
| ⑦ Manual Window | Before the formal test, you can check in the manual window to confirm that all device functions are working properly. |
| ⑧Test Parameters | Click to go to the test parameter settings screen |
| ⑨Lighting | Click to turn the lighting inside the test chamber on or off to facilitate observation of the test status. |
| ⑩ Reset to Default | Before the test, click here to clear the data from the previous test. |
| ⑪ Experiment Start | Click to start the water drip test |
| ⑫ Test Stopped | Click to stop the drip test |
| ⑬ Swivel Pipe Warning | If your test chamber includes IPX3 and IPX4, when you enter the IPX1 and IPX2 interface, the IPX3 and IPX4 swing arms will automatically swing to the side to facilitate IPX1 and IPX2 testing. Please wait until this red warning disappears before proceeding. |
| ⑭ Test Duration Display | Tampilkan durasi pengujian saat ini |
| ⑮ Standard Flow Display | Displays the standard water flow rate required for the current drip tray size |
| ⑯ Sample exposure time | Display the current power-on duration for this sample |
| ⑰ Sample power-off time | Tampilkan durasi pemadaman listrik untuk sampel saat ini. |
| ⑱Back | Kembali ke Layar Beranda |

Gambar 12
4.1.2.2 Introduction to the Test Parameter Settings Interface
| Nomor | Note |
| ① Durasi Tes | The total test duration can be set; the GB/T 4208-2017 standard specifies that the test duration for both IPX1 and IPX2 is 10 minutes. |
| ② Sample energization time | Atur durasi pasokan daya kontinu ke sampel selama pengujian, dalam detik. |
| ③ Time the sample is de-energized | Set the duration of the power interruption for the sample during the test, in seconds |
| ④ Turntable Speed | Only IPX1 can be configured; the standard setting is 1 r/min, with a customizable range of 0–5 r/min. IPX2 cannot be configured; rotation during testing is performed automatically according to the standard. Note: During IPX1 testing, for samples that require power to be tested, the turntable will automatically rotate in both directions to prevent the sample’s power cord from becoming tangled. |
| ⑤ Drainage duration | When the drip tray is activated to drain, the duration for which the air compressor automatically turns on can be set to 0.5 minutes; that is, the air compressor will automatically shut off after half a minute. |
| ⑥ Duration of water addition | Before officially starting the test, there is no water in the drip tray. You must first tap the drip tray to add water; this can be set to 2 minutes. |
| ⑦ Turntable Interval Time | Applies to IPX2. According to the standard, set the duration to 150 seconds, which means 2.5 minutes per angle for a total of 10 minutes across all four angles. |
| ⑧Back | Click to return to the IPX1 and IPX2 test interface |

Gambar 13
Note: Continuous power supply: Power-on time = test duration; power-off time = 0 min; No power supply: Power-on time = 0 min; power-off time = test duration.
4.1.2.3 Introduction to the Manual Window
| Nomor | Note |
| ① Water Pump | The water pump can be turned on or off; once turned on, water begins to flow into the drip tray at the flow rate set by the flow meter. |
| ② Sample Power Supply | Jika catu daya sampel dan sampel terhubung, Anda dapat menghidupkan atau mematikan catu daya sampel. |
| ③ Turntable Reversal | Click to rotate the turntable clockwise or counterclockwise; release to stop. Note: If a sample with a power cord attached is placed on the turntable, do not rotate it in one direction for an extended period to prevent the cord from becoming tangled. |
| ④ Meja Putar Berputar Searah Jarum Jam | |
| ⑤ Back | Click to return to the IPX1 and IPX2 test interface |

Gambar 14
Note: Before the official test, you can start the water pump in the manual window and adjust the IPX1 and IPX2 flow meters to the standard water flow rate. This ensures that the flow rate will be correct immediately after the test begins, eliminating the need for further adjustments.
4.1.3 IPX3 and IPX4 User Interfaces
4.1.3.1 Introduction to the Test Interface
| Nomor | Note |
| ① Peringkat ketahanan air: IPX3 | Once selected, you can configure the IPX3 test parameters and perform the IPX3 swing-nozzle spray test. |
| ② Peringkat ketahanan air: IPX4 | Once selected, you can configure the IPX4 test parameters and conduct the IPX4 oscillating spray test. |
| ③ Lighting | Click to turn the lighting inside the test chamber on or off to facilitate observation of the test status. |
| ④ Add water | Click to fill the water tank (Note: The water tank is equipped with upper and lower water level limit switches; if the water level is too low, a low-water alert will appear, and the system will automatically stop filling once the tank is full). |
| ⑤ Sample Power Supply | Daya sampel dapat dihidupkan atau dimatikan. Saat daya sampel dihidupkan untuk pengujian, program akan secara otomatis memasok daya ke sampel sesuai dengan waktu hidup/mati sampel yang telah diatur di antarmuka pengaturan parameter. |
| ⑥ Jendela Manual | Sebelum pengujian resmi, Anda dapat memeriksa apakah semua fungsi perangkat bekerja dengan benar di jendela manual. |
| ⑦ Test Parameters | Klik untuk mengakses layar pengaturan parameter pengujian. |
| ⑧ Return the swing arm to zero | If the pendulum arm is not in the 0-degree position (directly overhead), click it to reset it to zero before the test. |
| ⑨ Clear and Reset | Sebelum memulai tes, klik di sini untuk menghapus data dari tes sebelumnya. |
| ⑩ Experiment Start | Click to start the swing-arm rain test |
| ⑪ Test Stopped | Click to stop the swing-arm rain test |
| ⑫ Swivel Pipe Warning | If your test chamber supports IPX1 and IPX2, the swing arm will automatically return to the 0 position when you navigate from the IPX1/IPX2 screen to the IPX3/IPX4 screen, to facilitate IPX3/IPX4 testing. Please wait until this red warning disappears before proceeding. |
| ⑬ Selection of Swivel Pipe Dimensions | If your test chamber contains multiple swing arms, you must select the size of the swing arm currently in use. |
| ⑭ Test Duration Display | Tampilkan durasi pengujian saat ini |
| ⑮ Standard Flow Display | Displays the standard water flow rate required for the current swing pipe size and waterproofing rating |
| ⑯ Sample exposure time | Tampilkan durasi penyalaan daya saat ini untuk sampel tersebut. |
| ⑰ Sample power-off time | Tampilkan durasi pemadaman listrik untuk sampel saat ini. |
| ⑱Back | Kembali ke Layar Beranda |

Gambar 15
4.1.3.2 Introduction to the Test Parameter Settings Interface
| Nomor | Note |
| ① Durasi Tes | The total test duration can be set; the GB/T 4208-2017 standard specifies that the test duration for both IPX3 and IPX4 is 10 minutes. |
| ② Swing Angle | The IPX3 standard is set at 120°, the IPX4 standard at 350°, with an upper limit of 350°. |
| ③ Turntable Speed | Unit: r/min; adjustable from 1 to 5 r/min; factory default is 1 r/min; no adjustment is required for standard testing. Note: During IPX3 testing, in accordance with the standard, the turntable will automatically rotate to two angles, and the test is conducted with the sample at rest; During IPX4 testing, the turntable rotates continuously at a speed of 1 r/min. For samples that require power during testing, the turntable automatically rotates in both directions to prevent the sample’s power cord from becoming tangled. |
| ④ Time the sample is energized | Atur durasi pasokan daya kontinu ke sampel selama pengujian, dalam detik. |
| ⑤ Time the sample is de-energized | Set the duration of the continuous power interruption for the sample during the test, in seconds |
| ⑥ Back | Click to return to the IPX3 and IPX4 test interface |

Gambar 16
Note: Continuous power supply: Power-on time = test duration; power-off time = 0 min; No power supply: Power-on time = 0 min; power-off time = test duration.
4.1.3.3 Introduction to Manual Windows
| Nomor | Note |
| ① Saklar Pompa | The water pump can be turned on or off; once turned on, water begins to flow into the swivel pipe at the flow rate set by the flow meter. |
| ② Sample Power Supply | Jika catu daya sampel dan sampel terhubung, Anda dapat menghidupkan atau mematikan catu daya sampel. |
| ③ Forward rotation of the swivel pipe | Click to rotate the swivel pipe clockwise or counterclockwise; release to stop. Be careful not to let the swivel pipe collide with the turntable. |
| ④ Swing Arm Reversal | |
| ⑤ Turntable Rotates Clockwise | Click to make the turntable rotate clockwise or counterclockwise; release to stop. Note: If a sample with a power cord attached is on the turntable, do not rotate it in one direction for an extended period to prevent the cord from becoming tangled. |
| ⑥ Turntable Reversal | |
| ⑦ Back | Click to return to the IPX3 and IPX4 test interface |

Gambar 17
Note: Before the official test, you can start the water pump in the manual window and adjust the IPX3 and IPX4 flow meters to the standard water flow rate. This ensures that the flow rate will be correct immediately upon starting the official test, eliminating the need for further calibration.
4.1.4 IPX5 and IPX6 User Interfaces
4.1.4.1 Introduction to the Test Interface
| Nomor | Note |
| ① Peringkat ketahanan air: IPX5 | Setelah dipilih, Anda dapat mengkonfigurasi parameter uji IPX5 dan melakukan uji semprotan air IPX5. |
| ② Peringkat ketahanan air: IPX6 | Setelah dipilih, Anda dapat mengkonfigurasi parameter uji IPX6 dan melakukan uji semprotan air IPX6. |
| ③ Lighting | Click to turn the lighting inside the test chamber on or off, making it easier to observe the test status. |
| ④ Add water | Click to fill the water tank (Note: The water tank is equipped with upper and lower water level limit switches; if the water level is too low, a low-water alert will appear, and the system will automatically stop filling once the tank is full). |
| ⑤ Sample Power Supply | Daya sampel dapat dihidupkan atau dimatikan. Saat daya sampel dihidupkan untuk pengujian, program akan secara otomatis memasok daya ke sampel sesuai dengan waktu hidup/mati sampel yang telah diatur di antarmuka pengaturan parameter. |
| ⑥ Jendela Manual | Before the official test, you can check in the manual window to confirm that all device functions are working properly. |
| ⑦ Test Parameters | Click to go to the test parameter settings screen |
| ⑧ Clear and Reset | Before the test, click to clear the data from the previous test. |
| ⑨ Experiment Start | Klik untuk memulai uji semprotan air |
| ⑩ Test Stopped | Klik untuk menghentikan uji semprotan air |
| ⑪ Swivel Pipe Warning | If your test chamber includes IPX1 and IPX2, when you navigate from the IPX1/IPX2 screen to the IPX5/IPX6 screen, the swing arm will automatically return to the 0 position to facilitate IPX5/IPX6 testing. Please wait until this red warning disappears before proceeding. |
| ⑫ Test Duration Display | Tampilkan durasi pengujian saat ini |
| ⑬ Standard Flow Display | Displays the standard water flow rate required for testing the current waterproof rating |
| ⑭ Sample exposure time | Tampilkan durasi penyalaan daya saat ini untuk sampel tersebut. |
| ⑮Sample power-off time | Tampilkan durasi pemadaman listrik untuk sampel saat ini. |
| ⑯Back | Click to return to the IPX5 and IPX6 testing interface |

Gambar 18
4.1.4.2 Introduction to the Test Parameter Settings Interface
| Nomor | Note |
| ① Durasi Tes | The total test duration can be set; the GB/T 4208-2017 standard requires that the test duration for both IPX5 and IPX6 be ≥3 minutes. |
| ② Kecepatan Meja Putar | 1–5 r/min; the factory default turntable speed is 1 r/min. If the sample is large in volume or has a large surface area, the speed may be adjusted appropriately to 2–5 r/min Catatan: Untuk sampel yang membutuhkan daya selama pengujian, meja putar akan secara otomatis berputar ke kedua arah untuk mencegah kabel daya sampel kusut. |
| ③ Waktu sampel diberi energi | Set the duration of continuous power supply to the sample during the test, in seconds (s) |
| ④ Time the sample is disconnected from power | Set the duration of the power interruption for the sample during the test, in seconds |
| ⑤ Back | Klik untuk kembali ke antarmuka kerja. |

Gambar 19
Note: Continuous power supply: Power-on time = test duration; power-off time = 0 min; No power supply: Power-on time = 0 min; power-off time = test duration.
4.1.4.3 Introduction to Manual Windows
| Nomor | Note |
| ① IPX5 Water Pump Switch | The IPX5 water pump can be turned on or off; once turned on, water begins to flow into the IPX5 nozzle at the flow rate set by the flow meter. |
| ② IPX6 Water Pump Switch | The IPX6 water pump can be turned on or off; once turned on, water begins to flow into the IPX6 nozzle at the flow rate set by the flow meter. |
| ③ Turntable Rotating Clockwise | Click to rotate the turntable clockwise or counterclockwise; release to stop. Note: If a sample with a power cord attached is placed on the turntable, do not rotate it in one direction for an extended period to prevent the cord from becoming tangled. |
| ④ Turntable Reversal | |
| ⑤ Sample Power Supply | Jika catu daya sampel dan sampel terhubung, Anda dapat menghidupkan atau mematikan catu daya sampel. |
| ⑥ Back | Klik untuk kembali ke antarmuka kerja. |

Gambar 20
Note: Before the official test, you can start the water pump in the manual window, adjust the IPX5 and IPX6 flow meters to the standard water flow rate, and adjust and secure the nozzles so that the water jets are directed at the sample. This way, once the official test begins, no further adjustments will be needed.
4.2 Description of the Program Prompt Interface
IPX1 and IPX2 tests completed.
| Nomor | Note |
| ① | Click to use the air compressor to drain the water from the drip tray. |
| ② | Click to return to the IPX1 and IPX2 test interface |

Gambar 21
Check the filter assembly; if the filter cartridge has changed color significantly, replace it.

Gambar 22
The water tank is low on water.
| Nomor | Note |
| ① | Click to fill the tank with water |
| ② | After adding water, you can click “Back.” |

Gambar 23
Waktu pengujian belum direset. Silakan klik “Kembali,” lalu klik “Reset” untuk menghapus data dari pengujian sebelumnya dan coba lagi.

Gambar 24
No water resistance rating has been selected. Please click “Back,” select the water resistance rating you wish to test, and try again.

Gambar 25
The swing arm has not been reset to zero. Please click “Back,” then click “Reset Swing Arm” to rotate the swing arm to the 0-degree position, and try again.

Gambar 26
The test chamber door is not closed. Please close and lock the test chamber door, then click “Back” and try again. Note: Only the JL-123456CP and JL-3456CP models have this feature.

Gambar 27
The inverter has triggered an alarm. Please shut down the system and disconnect the main power supply to the equipment. Wait half a minute, then restart the system.

Gambar 28
Water Tank Cleaning Reminder: Please inspect and clean the water tank regularly.

Gambar 29
Once the test is complete, click “Back.”

Gambar 30
The emergency stop has not been reset. Please turn the emergency stop button clockwise to reset it, then try again.

Gambar 31
The water tank is full. Just tap “Back.”

Gambar 32
5. Water Resistance Test Procedure
5.1 Test Procedure Steps
5.1.1 IPX1 and IPX2 Operating Procedures
| petunjuk | Note |
| 1 | Enter the IPX1/IPX2 test interface. If a swing arm is present, it will automatically rotate to the side to facilitate testing. |
| 2 | Select the required waterproofing test level: IPX1 or IPX2 |
| 3 | Adjust the flow meter to the standard water flow rate as described in Section 4.1.2.3. |
| 4 | Refer to Section 4.1.2.2 to configure the test parameters. |
| 5 | Following the instructions in Section 3.4, set up the sample and adjust the height of the turntable so that the drip tray is 200 mm away from the sample. |
| 6 | Close and lock the test chamber door. |
| 7 | If there is not enough water in the test chamber, click to add water. |
| 8 | If the sample requires a power-on test, please click “Sample Power Supply.” |
| 9 | Click the drip tray to add water; the water will slowly flow into the drip tray. Wait until the water flow rate on the flow meter matches the standard flow rate displayed on the touchscreen interface. |
| 10 | Click “Reset” to clear the time from the previous test, then click “Start Test.” Throughout the entire test, the water dispenser will automatically begin filling with water. |
| 11 | Once the test duration reaches the set time, the test ends. The drip tray automatically stops dispensing water, the turntable automatically stops rotating, and the power to the sample is automatically cut off (if a powered test was being conducted); however, there will still be water remaining in the drip tray, so the sample can be removed immediately. |
| 12 | If an air compressor is connected, remove a few needles from the drip tray, then press the drain button on the drip tray to quickly use compressed air to drain the water from the drip tray. |
5.1.2 IPX3 and IPX4 Operating Procedures
| petunjuk | Note |
| 1 | When you enter the IPX3 and IPX4 test interface, if IPX1 and IPX2 tests were previously run, the swing arm will automatically rotate from the side to the 0-degree position to facilitate testing. |
| 2 | Select the required waterproof testing level: IPX3 or IPX4 |
| 3 | Selecting the Correct Swivel Pipe Size |
| 4 | For an IPX3 test, water is sprayed within 60° arc segments on either side of the swing tube’s midpoint; the nozzles in the swing tube’s non-60° arc segments must be sealed with plugs, leaving only the nozzles on either side of the midpoint (a total arc of 120°) . For the IPX4 test, water is sprayed within 180° arc segments on either side of the swing arm’s midpoint (approximately 360°); all plugs on the swing arm must be removed, and all nozzles must be installed. |
| 5 | Refer to Section 4.1.3.3 to calibrate the flowmeter to the standard water flow rate. |
| 6 | Refer to Section 4.1.3.2 to set the test parameters. |
| 7 | Following the instructions in Section 3.4, set up the sample and adjust the height of the turntable so that the swing arm is 200 mm from the sample. |
| 8 | Close and lock the test chamber door. |
| 9 | If there is not enough water in the test chamber, click to add water. |
| 10 | If the sample requires a power-on test, please click “Sample Power Supply.” |
| 11 | Click “Reset” to clear the time from the previous test; click “Pendulum Reset” to confirm that the pendulum is still at the 0-degree position; then click “Start Test.” |
| 12 | Once the test duration reaches the set time, the test ends. The water pump automatically stops pumping water, the swivel pipe automatically returns to the 0-degree position, the turntable automatically stops rotating, and the sample’s power is automatically cut off (if an electrically powered test was conducted). The sample can then be removed. |
5.1.3 IPX5 and IPX6 Operating Procedures
| petunjuk | Note |
| 1 | When you enter the IPX5/IPX6 test interface, if IPX1/IPX2 tests were previously run, the swing arm will automatically rotate from the side to the 0-degree position to facilitate the test. |
| 2 | Select the required waterproofing test level: IPX5 or IPX6 |
| 3 | Following the instructions in Section 4.1.4.3, adjust the flow meter to the standard water flow rate, then adjust and secure the nozzle orientation. |
| 5 | Refer to Section 4.1.4.2 to configure the test parameters. |
| 6 | Connect the sample as described in Section 3.4. |
| 7 | Close and lock the test chamber door. |
| 8 | If there is not enough water in the test chamber, click to add water. |
| 9 | If the sample requires a power-on test, please click “Sample Power Supply.” |
| 10 | Click “Reset” to clear the time from the previous test, then click “Start Test.” |
| 11 | Once the test duration reaches the set time, the test ends. The water pump automatically stops pumping water, the turntable automatically stops rotating, and the power to the sample is automatically cut off (if an electrical test was conducted); the sample can then be removed. |
5.2 Memantau Proses Pengujian
5.2.1 Turn on the lights
For easier viewing, you can click the “Lighting” button to turn on the light inside the equipment cabinet.
5.2.2 Monitoring of Exceptional Conditions
| Monitoring Projects | IPX1 IPX2 | IPX3 IPX4 | IPX5 IPX6 |
| Jarum | All needles dispense liquid evenly throughout the test. If a needle becomes clogged, use a needle cleaner to clear it before retrying. | None | |
| Swing Pipe | None | The swing angle of the swing arm is normal; there is no jamming or unusual noise. | |
| curat | None | Verify that water is coming out of the corresponding nozzle and that no water is coming out of the other nozzles. | |
| Tingkat Aliran Air | During testing, the flow rate remained stable; the flow meter can be fine-tuned to keep the water flow rate consistently at the standard level. | ||
| Turntable | The turntable rotates normally according to the settings. If it does not, check the parameter settings and the turntable, then try again after making adjustments. | ||
| Pengiriman | The power to the sample is automatically turned on and off according to the set schedule. If an abnormality occurs during testing—such as a short circuit—the test can be stopped immediately; after inspecting the sample, the test can be repeated. | ||
6. Prosedur Pasca-Ujian
6.1 Tindakan Segera Setelah Tes Selesai
Periksa apakah tampilan waktu pengujian pada antarmuka kerja berfungsi dengan benar, lalu klik “Hapus dan Atur Ulang” untuk bersiap menghadapi pengujian berikutnya.
TIDAK PERNAH Shutdown
After all tests for the day are completed, close the water inlet valve and turn off the equipment’s power switch.
6.3 Evaluasi Hasil Tes
Gunakan kain kering untuk menyeka noda air dari permukaan luar sampel, lalu uji sampel tersebut. Tentukan apakah sampel tersebut lulus uji berdasarkan kriteria berikut.
| Kriteria evaluasi | Persyaratan Khusus |
| Safety/keselamatan | Kekurangan pasokan air tidak memengaruhi pengoperasian normal peralatan atau membahayakan keselamatan. |
| Perlindungan Isolasi | Air tidak akan menumpuk pada komponen berinsulasi di mana hal itu dapat menyebabkan perambatan arus sepanjang jarak rambatan. |
| Perlindungan Komponen Listrik | Air tidak boleh masuk ke bagian yang dialiri listrik atau kumparan yang tidak dirancang untuk beroperasi dalam kondisi lembap. |
| Perlindungan Kabel | Air tidak menumpuk di dekat ujung kabel atau masuk ke dalam kabel. |
| Efektivitas Lubang Drainase | If there are drain holes, it must be verified that incoming water will not accumulate and can drain away without damaging the equipment. |
| Peralatan tanpa lubang pembuangan | Jika air menumpuk dan membahayakan bagian-bagian yang hidup, ikuti kriteria penerimaan yang ditentukan dalam standar produk yang relevan. |
7. Pemeliharaan dan Perawatan Peralatan Harian
7.1 Pembersihan Harian
If the sample contaminates the test chamber, clean it with clean water, drain the water from the tank, and refill it.
Thoroughly clean the water tank at least once a week. Drain any standing water, then wipe the tank dry with a clean cloth. After cleaning, check the interior walls of the tank for rust or scale buildup. If there is minor rust, gently wipe it with a rust remover, then rinse thoroughly with clean water.
7.2 Regular Inspections and Replacement
Needle: Check that the needle on the peristaltic tube is not clogged, loose, or deformed, and verify that the water is dripping evenly. If the needle is clogged, use a needle cleaner to clear the blockage.
Connections: Check all connections on the swing arm, water lines, and electrical circuits to ensure there are no leaks or loose connections.
Filter Cartridge: Check the filter cartridge inside the filter. If the cartridge has turned brown (indicating it is saturated with adsorbed impurities), replace it immediately with a new one.
7.3 Pemeliharaan Peralatan yang Tidak Digunakan
7.3.1 Short-term idling (within 1 month)
Drain all standing water from the water tank and water lines to prevent scale buildup on the needles and tubing;
Cover the entire unit with a dust-proof cloth to prevent dust from entering;
Simpan peralatan di tempat yang kering dan berventilasi baik.
7.3.2 Ketidakaktifan jangka panjang (lebih dari 1 bulan)
Complete all maintenance tasks required for short-term storage; apply a layer of rust-preventative oil to the interior walls of the test chamber and water tank, as well as the surfaces of metal components, to prevent rust; store the equipment in a cool, dry place away from direct sunlight;
Apply a small amount of industrial lubricant to moving parts, such as turntable bearings, to prevent rust and jamming;
The equipment should be powered on once a month. Switch to the manual screen to test each function, and shut it down after 5 to 10 minutes of operation to prevent circuit degradation.
Memecahkan Masalah dan Menyelesaikan Masalah Umum
General 8.1
| Gejala Kesalahan | Kemungkinan penyebab | Prosedur |
| Alarm Layar Sentuh | Lihat 4.2 | |
| The touchscreen is lit, but the buttons do not respond when tapped. | Touchscreen freeze, stains or water marks on the surface, internal touch module malfunction | Turn off the device, wait 5 minutes, and then restart it; wipe the touchscreen with a dry cloth to remove any dirt or water stains; if it still does not respond, contact us for repair. |
8.2 IPX1 IPX2
| Gejala Kesalahan | Kemungkinan penyebab | Prosedur |
| The water drips unevenly from the needle | Test chamber tilt, clogged needle, insufficient water flow | Use a level to ensure the test chamber is vertical and not tilted; use a cleaning needle to unclog any nozzles that are not dispensing water; adjust the flow meter to the standard water flow rate. |
8.3 IPX3 IPX4
| Gejala Kesalahan | Kemungkinan penyebab | Prosedur |
| The water flow from the needle is uneven. | Test chamber tilt, clogged needle, insufficient water flow | Use a spirit level to ensure the test chamber is vertical and not tilted; use a cleaning needle to unclog any nozzles that are not dispensing water; adjust the flow meter to the standard water flow rate. |
| The water flow is insufficient and cannot be increased. | Clogged needle, clogged filter, clogged water pump | Use a cleaning needle to unclog the nozzle that isn’t dispensing water; replace the filter cartridge; disassemble the water pump, clean out the internal debris, and reassemble it. |
8.4 IPX5 IPX6
| Gejala Kesalahan | Kemungkinan penyebab | Prosedur |
| Air mengalir secara terputus-putus dari nosel. | Air di dalam tangki tidak mencukupi; ada udara di dalam saluran air. | Please click to add water to ensure there is enough water in the tank; continue running the machine to purge any air from the water lines. |
| Aliran air tidak mencukupi | Benda asing di dalam meteran aliran, benda asing di dalam katup solenoid, benda asing di dalam pompa air. | Mohon periksa apakah flow meter berfungsi dengan baik, apakah ada kotoran di dalamnya, dan apakah gasketnya terlepas. Lepaskan flow meter dari unit, bersihkan kotoran, atau pasang kembali gasket. Silakan hubungi kami untuk pemeriksaan katup solenoid dan pompa air. |
| IPX5 and IPX6 Water Jets Simultaneously | There is debris in the solenoid valve | Silakan hubungi kami untuk memeriksa katup solenoid. |

中文简体