Alat Uji untuk Pengujian pada Pin Steker

Pengguna manual

berlaku LISUN models: GNGPL-3632

1. Pemahaman Dasar tentang Peralatan

1.1 Ruang Lingkup Penerapan

The GNGPL-3632 Test Apparatus for Tests on Plug Pins of BS 1363-1 Figure 32 is a specialized testing device designed to evaluate the mechanical strength of British 13A plug pins. By applying a specified static force to the plug pins, it verifies whether they possess sufficient strength to withstand stresses encountered during normal use, thereby ensuring that the plug products meet safety requirements.

1.2 Standar yang Berlaku

This equipment is designed and manufactured in strict accordance with the requirements of BS 1363-1:2023, “13A Plugs, Sockets, Adapters, and Connectors—Part 1: Specification for Fused Plugs, Reconnectable and Non-Reconnectable,” specifically clause 13.9.4 and the requirements of Figure 32a and Figure 32b, and fully meets the standard’s strength test requirements for both solid and non-solid pins.

1.3 Prinsip Operasi

1.3.1 Principles of Mechanical Testing

The equipment employs a anvil-pressurized testing principle, in which a fixed anvil supports one side of the test pin, while a movable anvil applies pressure perpendicular to the pin’s main axis plane from the other side. By controlling the magnitude of the force, the loading rate, and the number of cycles, the test evaluates the plug’s deformation and structural integrity under specified stress conditions.

1.3.2 Principles of Standards Compliance

The dimensions, shape, clearances, and alignment methods of the equipment’s anvil blocks are strictly processed in accordance with the specifications in BS 1363-1:2023, Figure 32b. Key dimensions—such as the fit clearance between the fixed and movable anvil bases, the fillet radius, and the working surface width—are all controlled within the standard tolerance limits to ensure that test results are consistent with and traceable to the requirements of the standard.

1.4 Mengunduh Dokumen Elektronik

We have already emailed the download links for the electronic user manual, calibration certificate, warranty card, and other documents at the time of shipment. Please download them promptly. If you have not received the email, please contact us to request the download links again.

2. Tindakan Pencegahan Keamanan

2.1 Keamanan Listrik

Before connecting the power cord, verify that the supply voltage is AC 220 V ±10% and the frequency is 50 Hz or 60 Hz. You must use a three-prong outlet with a protective ground, and the protective ground resistance must be ≤4 Ω. Never cut or disconnect the protective ground wire, as doing so poses a risk of electric shock.

2.2 Keselamatan Lingkungan

Ambient temperature: 15°C to 35°C; relative humidity: ≤80% RH, no condensation; Installation surface: Level and stable, with minimal vibration; free from strong electromagnetic interference; Do not use in environments with direct sunlight or excessive dust; Do not operate in environments containing flammable, explosive, or corrosive gases; Allow at least 200 mm of clearance around the equipment for operation and heat dissipation.

2.3 Keselamatan Operasional

Never reach into the anvil’s working area with your hands or any other objects during loading to prevent pinching injuries; when installing plug specimens, ensure they are securely clamped to prevent them from flying out during the test; never strike, grind, or modify the dimensions of the anvil on your own, as this may affect test accuracy; Keep the equipment level when moving or transporting it to avoid severe vibrations and impacts; ensure that no foreign objects are jammed in the drive leadscrew or guide rails, and periodically check for smooth operation.

3. Ikhtisar Peralatan

3.1 Diagram Standar

3.1.1 Schematic Diagram of the Plug Test

See Figure 3-1 (taken from BS 1363-1:2023, Figure 32a).

Test Apparatus for Tests on Plug Pins-Figure1

Gambar-3 1

3.1.2 Schematic Diagram of the Fixed Anvil and Moving Anvil

See Figure 3-2 (taken from BS 1363-1:2023, Figure 32b).

Test Apparatus for Tests on Plug Pins-Figure2

Gambar-3 2

3.2 Foto Objek Sebenarnya

3.2.1 Overall Appearance of the Equipment

Lihat Gambar 3-3.

Jumlah Note
Saklar daya
Current Test Force Display
A spirit level to ensure the equipment is level
berat
The position of the weight suspension point is adjustable to utilize the principle of leverage to generate different test forces.
Lifting screw: When not conducting tests, you can use the screw attached to the screw to lift the lever.
Turn the screw by hand to raise or lower the lead screw.
Control the up-and-down movement of the lever to ensure that the force is applied vertically downward at the point of application.
Point of Application
Sensor Angkatan
The plug body is supported to ensure the plug is securely installed
1 Moving Anvil Retaining Screw
Moving anvil, corresponding to item 3 in Figure 32b
Fixed anvil, corresponding to item 2 in Figure 32b

Test Apparatus for Tests on Plug Pins-Figure3

Gambar-3 3

3.2.2 Detailed Views of the Fixed Anvil and Moving Anvil

Lihat Gambar 3–4.

Jumlah Note
Moving anvil, corresponding to item 3 in Figure 32b
Fixed anvil, corresponding to item 2 in Figure 32b
The plug body is supported to ensure the plug is securely installed
Moving Anvil Retaining Screw

Test Apparatus for Tests on Plug Pins-Figure4

Gambar-3 4

4. Plug Test Procedure

4.1 Zero Calibration of Force Values

Raise the lead screw to lift the lever, and check whether the current test force reading shows 0. If a force value is displayed, click “CLEAR” on the screen to reset the current force value to zero.

4.2 Pemasangan Spesimen

Install the specimen as shown in Figure 32a.

Place the plug under test horizontally in the slot of the fixed anvil, with the contact surface of the plug lying in the horizontal plane; adjust the plug support so that the end face of the plug is in close contact with the end face of the fixed anvil, and the longitudinal axis of the plug is perpendicular to the slot; confirm that the plug is securely positioned and will not slide or tip over on its own.

4.3 Force Calibration

Loosen the moving anvil retaining screw so that the moving anvil contacts the specimen pin, then tighten the moving anvil retaining screw to prevent force transmission to the specimen pin when the test is not in progress, and then lower the lever. Using different weights and different suspension positions, calibrate the force value displayed on the screen to the desired force value, ensuring that the lever remains level.

4.4 Prosedur Uji

Loosen the moving anvil retaining screw so that the force from the moving anvil is transmitted to the specimen pin. Once the test duration has reached the required time specified in the standard, manually remove the weights, raise the moving anvil, and remove the specimen.

4.5 Evaluasi Hasil

Visually inspect the plug pins for cracks or obvious deformation; use the gauge specified in Figure 5 to check whether the plug passes the test in accordance with the method described in 13.2.1; if all three pins pass, the test is considered passed.

5. Pemeliharaan dan Perawatan Peralatan Harian

5.1 Daily Maintenance After Shutdown

5.1.1 Reset Perangkat

Use the lead screw to raise the lever, and avoid applying force when the test is not in progress. Secure the moving anvil in a position away from the fixed anvil using the moving anvil set screw to prevent deformation caused by collision or compression.

5.1.2 Equipment Cleaning and Inspection

At the end of each day’s testing, clean dust and sample debris from the equipment surfaces, and keep the fixtures and slide rails clean and free of debris; clean the touchscreen with a dedicated screen-cleaning cloth; and check that the hooks are not loose or deformed.

5.2 Weekly Routine Maintenance

Apply a small amount of neutral lubricating grease to the equipment’s lever and lever pivot, the moving anvil slide rail, and the lead screw to ensure smooth, jamming-free operation.

5.3 Pemeliharaan Selama Penyimpanan Jangka Panjang

After the equipment has been powered off and maintenance has been completed, apply a small amount of rust-preventive oil to the working surface of the anvil base, cover the entire unit with a dust cover, and store it in a dry, well-ventilated warehouse; apply rust-preventive oil to the weights and store them separately; power on the equipment for 30 minutes each month to prevent moisture damage to the components.

5.4 Persyaratan Interval Kalibrasi

Untuk memastikan akurasi pengukuran dan keandalan operasional peralatan, disarankan agar peralatan tersebut secara berkala dikirim ke organisasi metrologi pihak ketiga yang berkualifikasi untuk kalibrasi. Interval kalibrasi yang disarankan adalah 12 bulan. Pengguna dapat menentukan interval kalibrasi spesifik berdasarkan frekuensi penggunaan, kondisi lingkungan, dan persyaratan sistem mutu.

6. Pemecahan Masalah dan Penyelesaian Masalah Umum

Gejala Kesalahan Kemungkinan penyebab Prosedur
Force value drifts from 0 Force value has not been reset Refer to Section 4.1; reset the force value to zero and try again.
Abnormal Force Value Moving Anvil Stuttering Make sure to completely loosen the screws securing the movable anvil and ensure that the movable anvil slide rail is sufficiently lubricated.