Perangkat Uji Tekanan Suhu Tinggi
Pengguna manual
berlaku LISUN models: GNGPL-3610、GNGPL-3610-2PA、GNGPL-3610-3PA
1. Pemahaman Dasar tentang Peralatan
Spesifikasi Teknis 1.1
| LISUN Pilih Model | GNGPL-3610 | GNGPL-3610-2PA |
| Jumlah stasiun kerja | 1 | 2 |
| Mata pisau | 1 round blade (Ф6) | One rectangular blade and one Ф6 circular blade |
| Ketebalan pisau | 0.7 mm | |
| berat | 2.5N weight (including the weight of the blade): 1 | 2.5N weights (including the weight of the blade): 2 |
| Standar yang berlaku | BS 1363-1 Gambar 10 | GB/T 2099.1, Clause 30.1 and Figure 41 (IEC 60884-1, Clause 30.1 and Figure 44) |
Note: This device must be used in conjunction with the LISUN GW-225 High Temperature Chamber and is specifically designed for high-temperature pressure testing of insulation sheaths on plugs and prongs for household and similar purposes.
1.2 Skenario Aplikasi
Plug and Socket Manufacturers: Inspection of incoming plastic materials, 100% final inspection of finished products, and comparative testing of new materials; Third-Party Safety Testing Laboratories: CCC certification, CE (IEC) certification, and mandatory testing for UK BS certification; Quality Inspection / Market Regulatory Authorities: Random inspections of plug products in e-commerce and brick-and-mortar markets; Small Appliance Manufacturers: In-house reliability verification of plugs for adapters, rice cookers, electric kettles, and desk lamps.
1.3 Mengunduh Dokumen Elektronik
Kami telah mengirimkan tautan unduhan untuk manual pengguna elektronik, sertifikat kalibrasi, kartu garansi, dan dokumen lainnya melalui email pada saat pengiriman. Silakan unduh sesegera mungkin. Jika Anda belum menerima email tersebut, silakan hubungi kami untuk meminta tautan unduhan lagi.
2. Tindakan Pencegahan Keamanan
During high-temperature testing, temperatures may reach up to 200°C. When opening the door to remove or place test specimens, wear heat-resistant gloves; do not touch the inner chamber, blades, weights, or fixtures with bare hands to prevent burns.
Weights are heavy metal objects; handle them gently. Do not throw or stack them. The blades are sharp; be careful of the cutting edges when installing or replacing test specimens to avoid cutting your hands. When moving the entire press fixture, lift it steadily to prevent the blades from causing scratches.
3. Ikhtisar Peralatan
3.1 Diagram Standar
See Figure 3-1 (taken from IEC 60884-1, Figure 44).

Gambar-3 1
3.2 Photographs of the Actual Items
3.2.1 GNGPL-3610
Lihat Gambar 3-2.
| Jumlah | Note |
| ① | Pisau Bulat |
| ② | berat |
| ③ | Basis disesuaikan |

Gambar-3 2
3.2.2 GNGPL-3610-2PA
Lihat Gambar 3-3.
| Jumlah | Note |
| ① | Rectangular Blade |
| ② | Pisau Bulat |
| ③④ | berat |
| ⑤ | Basis disesuaikan |

Gambar-3 3
4. High-Temperature Pressure Test of the Insulation Sheath
4.1 Preparation of Specimens and Apparatus
Rectangular blades are used for round-pin tests; circular blades are used for other pin tests. Refer to Section 3.1 to position the specimen correctly.
4.2 Prosedur Uji
Set the high-temperature chamber to 200°C and allow it to stabilize at the target temperature. Place the test setup, along with the test specimen mounted in its normal position, in the high-temperature chamber for 2 hours.
Next, remove the test apparatus and the specimen, detach the specimen from the apparatus, and immerse it in cold water within 10 seconds to cool it.
4.3 Evaluasi Hasil Tes
Measure the thickness of the insulating material at the indentation point; the reduction in thickness compared to the original measurement at the start of the test should not exceed 50 percent.
4.4 Post-Test Procedures
4.4.1 Reset Perangkat
After the test is complete, remove the weights and blades from the test rig bracket. Do not allow the weights and blades to remain suspended on the test rig for extended periods when the test is not in progress.
4.4.2 Persiapan Sampel
Pisahkan sampel yang lulus dan gagal dalam pengujian, beri label dengan jelas, dan hindari kebingungan.
5. Pemeliharaan dan Perawatan Peralatan Harian
5.1 Daily Maintenance After Shutdown
At the end of each day’s testing, clean dust and sample debris from the equipment surfaces, and keep the chucks and slide rails clean and free of debris; check that the blades are not deformed.
5.2 Pemeliharaan Selama Penyimpanan Jangka Panjang
After routine maintenance is complete, apply a neutral rust-preventive oil to the test apparatus, blades, weights, and other components to prevent rust; cover the entire unit with a dust cover and store it in a dry, well-ventilated warehouse.
5.3 Persyaratan Interval Kalibrasi
Untuk memastikan akurasi pengukuran dan keandalan operasional peralatan, disarankan agar peralatan tersebut secara berkala dikirim ke laboratorium metrologi pihak ketiga yang berkualifikasi untuk kalibrasi. Interval kalibrasi yang disarankan adalah 12 bulan. Pengguna dapat menentukan interval kalibrasi spesifik berdasarkan frekuensi penggunaan, kondisi lingkungan, dan persyaratan sistem mutu.
6. Pemecahan Masalah dan Penyelesaian Masalah Umum
| Gejala Kesalahan | Kemungkinan penyebab | Prosedur |
| The blade is not perpendicular to the sample | The test setup is not level. | Please adjust the base of the test apparatus so that it is level. |