+8618117273997weixin
InggrisInggris
中文简体 中文简体 en English ru Русский es Español pt Português tr Türkçe ar العربية de Deutsch pl Polski it Italiano fr Français ko 한국어 th ไทย vi Tiếng Việt ja 日本語

Perangkat Benturan Suhu Rendah

Pengguna manual

berlaku LISUN Model: DWC-1

1. Pemahaman Dasar tentang Peralatan

1.1 Lingkup Aplikasi Peralatan

The DWC Series Low-temperature Impact Device is specialized testing equipment designed and manufactured in accordance with multiple national and international standards. It is primarily used to test the impact resistance of PVC-sheathed cables, wires, flexible cords, and unsheathed parallel flexible cords in low-temperature environments, and to verify whether the materials’ low-temperature brittleness and impact strength meet the requirements of these standards.

1.2 Standar yang berlaku

LISUN Pilih Model DWC-1 DWC-2 DWC-3
Menjatuhkan Berat 100 gram, 200 gram, 300 gram, 400 gram, 500 gram, 600 gram, 750 gram, 1000 gram, 1250 gram, 1500 gram 100 g, 1000 g 0.5 kg, 1.0 kg, 2.0 kg, 6.8 kg
Standar yang berlaku GB/T 2951.14, Section 8.5 and Figure 2 (IEC 60811-1-4, Section 8.5 and Figure 2), GB/T 5013.1 (IEC 60245-1), GB/T 5023 (IEC 60227-1) GB/T 2099.1, Section 24.5 and Figure 27, and Section 30.5 and Figure 42 (IEC 60884-1, Section 24.5 and Figure 30, and Section 30.5 and Figure 42), GB/T 17466.1, Figure 21 (IEC 60670-1, Figure 21) GB/T 20041.1, Section 10.3 and Figure 2 (IEC 61386-1, Section 10.3 and Figure 2)

Note: This manual uses the DWC-2 as an example.

1.3 Prinsip Tes

1.3.1 Description of Supporting Equipment

This device is a shock actuator ke harus digunakan setara dengan itu LISUNG DJW series low-temperature test chambers untuk meet the low-temperature environmental testing requirements specified in the standards. The low-temperature test chamber provides a stable low-temperature environment; both the shock device and the test specimens must be pre-cooled in this environment.

1.3.2 Prinsip Pengujian

After holding the test specimen in a specified low-temperature environment for a specified period of time, a specified-weight impact weight is dropped freely from a set height to strike a designated area of the specimen. Following the test, the specimen is inspected for defects such as cracks or breakage, and the material’s low-temperature impact resistance is evaluated based on these findings.

1.3.3 Workflow Logic

Condition the specimen at low temperature → Place the specimen in the impact station → Lock the impact weight at the set height → Release the weight for a free-fall impact → Remove the specimen for inspection → Evaluate the test results

2. Tindakan Pencegahan Keamanan

2.1 Installation Environment for the Equipment

The equipment must be placed on a level and stable test bench with a tilt of no more than 2°;

Leave at least 50 cm of clearance around the area and keep it away from flammable and explosive materials.

2.3 Pedoman Keselamatan Operasional

Operators must undergo training on this manual and become familiar with the equipment’s structure and operating procedures before being allowed to operate it;

Operators with long hair must tie it back to prevent it from getting caught in moving parts of the equipment;

When operating the equipment, wear cryogenic protective gloves to prevent frostbite from contact with cold test specimens or the inner walls of the cryogenic chamber;

Wear safety goggles to prevent injury to the eyes from flying fragments of the specimen;

Damaged test specimens resulting from the impact must be collected properly to avoid cuts from sharp edges;

It is strictly prohibited to test specimens that exceed the device’s rated load capacity;

The use of non-standard weights or impact distances for testing is strictly prohibited.

3. Pengenalan Perangkat

3.1 Diagram Standar

3.1.1 Schematic Diagram of The Low-Temperature Impact Device For Plugs

See Figure 3-1 (taken from IEC 60884-1:2022, Figure 30).

Low-Temperature Impact Device-Figure1
Gambar-3 1

3.1.2 Schematic Diagram of The Low-Temperature Impact Device For Plug Insulation Sleeves

See Figure 3-2 (taken from IEC 60884-1:2022, Figure 45).

Low-Temperature Impact Device-Figure2
Gambar-3 2

3.2 Photographs of the Actual Device

3.2.1 Test Setup Components

Lihat Gambar 3-3.

Jumlah Note
Steel Base and Bracket
Adjustable Weight Holder
Weight Locking Device
Release mechanism: Press to release the weight for the impact test
Adjustable guide rod used to control the vertical descent of weights and secure steel spacers
Steel Spacers for 100g Drop Weights
100-gram drop weight
100-mm steel ruler, to control the impact height
Low-Temperature Impact Device-Figure3
Gambar-3 3

Note: The 1000-gram drop weight comes with a matching steel spacer; see Figures 3-4.

Jumlah Note
1000-gram drop weight
1000g Drop Weight with Steel Spacer Block
100-gram drop weight
Steel Spacers for 100g Drop Weight
Low-Temperature Impact Device-Figure4
Gambar-3 4

3.2.2 Impact Test Setup Diagram

Taking a 100-gram drop weight as an example—specifically, the low-temperature impact test on the insulated sleeve of a plug as described in Section 3.1.2—see Figure 3-4.

Jumlah Note
100g Drop Weight Installation
Installation of Steel Spacers
Note: During the test, place the test specimen pin between the steel spacer and the steel base. See 3.1.2.
Use a 100-mm steel ruler to determine the impact height, which is the distance between the falling hammer and the steel spacer.
Low-Temperature Impact Device-Figure5
Gambar-3 5

For a 1000-gram drop weight, simply replace ② with the corresponding steel spacer.

4. Low-Temperature Impact Test

4.1 Low-Temperature Impact Test for Plugs

See Clause 24.5 of IEC 60884-1: 2022.

Place the test setup, which is positioned on a 40-mm-thick sponge rubber block, along with three intact test specimens, into a test chamber maintained at a temperature of -15°C for at least 16 hours. At the end of this period, place each test specimen in the test position one by one and drop the hammer from a height of 100 mm. The mass of the hammer is 1000 g. After the test, the test specimens shall show no damage within the scope of this document.

4.2 Low-Temperature Impact Test for Plug Insulation Sleeves

See Clause 30.5 of IEC 60884-1: 2022.

A device consisting of a 100-gram drop weight placed on a 40-mm-thick sponge rubber pad, together with three intact test specimens, is placed in a test chamber maintained at a temperature of -15°C for at least 24 hours. At the end of this period, place each specimen in the test position one at a time and allow the drop weight to fall from a height of 100 mm. The same test specimen shall undergo four consecutive impacts; between each impact, the specimen shall be rotated 90°. After the test, the specimen shall be allowed to return to approximately room temperature before inspection. Without any additional magnification, the insulating sheath shall be free of cracks visible to the naked eye or with corrected vision.

5. Pemeliharaan dan Perawatan Peralatan Harian

Pembersihan dan Perawatan Harian

5.1.1 Maintenance After Each Test

Wipe the surface of the device with a dry cloth to remove condensation and dust; check the surfaces of the weights, steel shims, and steel base for stains, and wipe them clean immediately.

5.1.2 Perawatan Mingguan

Clean the surface of the guide rail with anhydrous ethanol to ensure it is smooth and rust-free; check all fastening screws for looseness and tighten them as needed; check whether the spring in the release mechanism is functioning properly.

Regular Lubrication and Rust Prevention

Apply a thin layer of low-temperature grease to the surface of the guide rail to reduce friction; apply a small amount of lubricating oil to the release mechanism and the areas around the mounting screws; check the stainless steel surface for any signs of rust, remove any rust promptly, and apply a rust-preventive treatment.

5.3 Penyimpanan dan Pemeliharaan Jangka Panjang

Pre-treatment for Storage

Thoroughly clean and dry the equipment, and apply rust-preventative oil to all metal surfaces; pack and store weights and other accessories separately to prevent damage; cover the entire unit with a dust cover for protection.

Persyaratan Lingkungan Penyimpanan

Store indoors in a dry, well-ventilated area free of corrosive gases; ambient temperature should be between 5°C and 35°C, with relative humidity not exceeding 80%; keep the equipment in a horizontal position to prevent deformation caused by heavy pressure.

5.4 Persyaratan Interval Kalibrasi

Untuk memastikan keakuratan dan keandalan perangkat, disarankan agar perangkat tersebut secara berkala dikirim ke laboratorium metrologi pihak ketiga yang berkualifikasi untuk kalibrasi. Interval kalibrasi yang disarankan adalah 12 bulan. Pengguna dapat menentukan interval kalibrasi spesifik berdasarkan frekuensi penggunaan, kondisi lingkungan, dan persyaratan sistem mutu.

6. Pemecahan Masalah dan Penyelesaian Masalah Umum

Gejala Kesalahan Prosedur
The drop hammer gets stuck during descent Use a level to ensure the device is placed horizontally; make sure the guide rail is vertical, remove any dirt or foreign objects from the rail, and check the clearance between the drop weight and the guide rail.
Addressing Institutional Failures Inspect the spring; if it is defective, replace it. If it is sticking, apply a neutral lubricant. If the mounting screws are loose, adjust the release mechanism cable until it is at the proper tension, then retighten the screws.