+8618117273997weixin
InggrisInggris
中文简体 中文简体 en English ru Русский es Español pt Português tr Türkçe ar العربية de Deutsch pl Polski it Italiano fr Français ko 한국어 th ไทย vi Tiếng Việt ja 日本語

Catu Daya DC CC dan CV Digital

Pengguna manual

berlaku LISUN models: DC100020B、DC12010B、DC12030B、DC3050B、DC50010B

1. Membuka Kemasan dan Mengenal Peralatan Secara Dasar

1.1 Spesifikasi Teknis Utama

Spesifikasi/Nomor Model DC3050B DC12030B DC50010B
Maximum Output Voltage and Current 30 V / 50 A 120 V / 30 A 500 V / 10 A
Resolusi Tegangan 0.01 V (<100 V); 0.1 V (>100 V)
Akurasi Tegangan ± 0.1 V.
Resolusi Saat Ini 0.01 A (<100 A); 0.1 A (>100 A)
Akurasi saat ini ± 0.1 A

1.2 Pemeriksaan dan Verifikasi Aksesori Saat Membuka Kemasan

1.2.1 Membuka Kemasan dan Pemeriksaan Visual

Saat membuka kemasan perangkat, tangani dengan hati-hati untuk menghindari goresan pada casing luar dengan alat tajam. Setelah membuka kemasan, periksa terlebih dahulu bagian luar perangkat untuk memastikan tidak ada deformasi atau penyok yang terlihat, dan tombol-tombolnya tidak rusak atau longgar. Jika Anda melihat kelainan pada perangkat atau aksesorinya, jangan menyalakannya. Segera hubungi kami.

1.2.2 Unit Utama dan Aksesoris

Host: 1 DC-B Series power supply, with the input power supply cord pre-installed.

Digital CC And CV DC Power Supply-Figure1

Gambar 1

Other accessories are as follows:

Jumlah Note
Equipment Handle Mounting Accessories
Kabel USB
Kabel RS-232
Sampling Line Accessories

Digital CC And CV DC Power Supply-Figure2

Gambar 2

Note: The standard DC-B unit does not include PC software and does not require a USB or RS-232 cable. If you need to use PC software to control the DC power output, please contact us for a custom solution.

Links to download the electronic user manual, warranty card, and other documents have been sent via email. Please download them as soon as possible. If you missed the email, please feel free to contact us to receive the download links again.

2. Tindakan Pencegahan Keamanan

2.1 Standar Keselamatan Listrik

The equipment requires a single-phase AC 220 V ±10 V input power supply that complies with safety standards and is equipped with a proper protective grounding system to prevent electric shock.

2.2 Standar Keselamatan Lingkungan

For indoor use only; avoid direct sunlight. Place on a sturdy, stable surface to avoid severe vibrations. Keep away from dust, corrosive gases, and flammable or explosive materials. Keep away from strong magnetic and electric fields. Ensure good ventilation; leave at least 10 cm of space around the device for heat dissipation, and do not block the cooling fan vents. Operating Environment: Temperature 5°C to 35°C, relative humidity <65% RH.

2.3 Pedoman Keselamatan Operasional

Do not operate the equipment or plug in/unplug the power cord with wet hands; the output must be turned off before connecting or disconnecting wires; allow the unit to warm up for at least 10 minutes after powering on before enabling the output; Do not touch the output terminals, internal components, the load, or the load wiring while the equipment is operating; allow at least 3 minutes between powering on and off to prevent damage from electrical surges; do not shut down the equipment directly while it is in output mode; first shut off the output before powering down; if you notice an unusual odor, smoke, unusual noise, or overheating, immediately disconnect the power and do not continue operation.

3. Gambaran Umum Peralatan dan Pengkabelan

3.1 Pengantar Panel Depan

Jumlah Note
Saklar daya
Sakelar Output
Voltage Display Window V
Current Display Window A
Power display window W; displays “OFF” when no power is being output
Equivalent impedance display window R; displays “—-” when no output is present.
Perlindungan Arus Lebih
Perlindungan Overvoltage
Perlindungan Over-Power
Perlindungan Over-Temperature
Lampu Indikator Keluaran Tegangan Konstan (CV)
1 Kenop Kontrol Tekanan
Lampu Indikator Keluaran Arus Konstan (CC)
Flow Control Knob

Digital CC And CV DC Power Supply-Figure3

Gambar 3

3.2 About the Back Panel

3.2.1 Regarding the Input Power Supply Cord

Depending on the power rating of the equipment, the pre-installed power cord will vary.

Jumlah Note
Low-power devices come pre-installed with standard mains power plugs
High-wattage appliances come pre-installed with thicker power cords; do not plug them directly into low-wattage outlets.

Digital CC And CV DC Power Supply-Figure4

Gambar 4

3.2.2 Pengenalan Panel Belakang dan Pengkabelan

Jumlah Note
① Kabel daya Connect the device to an appropriate power source
② Power Circuit Breaker
③ Output Terminals Connect to your DC load
④ Sampling Terminal It also connects to your DC load, allowing you to obtain a more accurate load-side voltage. If you require high voltage sampling accuracy, please connect it; leaving it unconnected does not affect the DC power supply’s output functionality.
⑤ Ground Terminal If your load requires a ground connection, you can connect its ground wire to this terminal.
⑥ Communication Port If you need the device to communicate with a computer, connect a USB/RS232/RS485 cable to the computer.

Digital CC And CV DC Power Supply-Figure5

Gambar 5

Note: If the sampling terminals are connected, this is a 4-wire sampling configuration, which eliminates cable voltage drop; if the sampling terminals are not connected, cable voltage drop is included.

4. Pengoperasian Peralatan

4.1 Keluaran Tegangan Konstan

4.1.1 Current Adjustment

If constant-voltage output is required, the current serves as a current-limiting reference. You can set it directly to the maximum value, or at least to a value sufficient to maintain the set constant voltage. If the current is set too low, the current limit will be triggered, and constant-voltage output cannot be maintained.

petunjuk Note
Press the current adjustment knob, and the last digit in the current display window will begin to flash.
Turn the flow control knob to change the digit in that position.
Continue pressing the flow adjustment knob to cycle through the other digits, and use the same method to change each digit.
After completing the setup, wait 2 seconds; the device will automatically exit setup mode.

4.1.2 Adjusting the Voltage

If a constant-voltage output is required, set the voltage to the load’s rated voltage.

petunjuk Note
Press the voltage adjustment knob, and the last digit of the voltage display will begin to flash.
Turn the volume control knob to change this digit.
Continue pressing the voltage adjustment knob to cycle through the other digits, and use the same method to change that digit.
After completing the setup, wait 2 seconds; the device will automatically exit setup mode.

4.1.3 Turn Output On/Off

Example: I need a constant-voltage output of 22V. Set it up as follows.

Digital CC And CV DC Power Supply-Figure6

Gambar 6

Jumlah Note
Press the output switch to activate the output.
The constant-voltage indicator light is on, and the device is in constant-voltage output mode.
The voltage display window shows the output voltage, which matches the set value.
The current display window shows the output current. It does not need to match the set value, but the set value must be higher than the actual value; otherwise, constant-voltage output will not be possible.
The power display window shows the output power.
The Equivalent Impedance window displays the equivalent impedance of the current output line.

Digital CC And CV DC Power Supply-Figure7

Gambar 7

Press the output switch again to turn off the output.

4.2 Keluaran Arus Konstan

4.2.1 Adjusting the Voltage

If constant-current output is required, the voltage serves as the over-voltage limit reference. You can set it directly to the maximum value, or at least to a value that satisfies the set constant-current value. If the voltage is set too low, it will trigger the over-voltage limit, and the constant-current output cannot be maintained.

For instructions on adjusting the voltage, see Section 4.1.2.

4.2.2 Current Adjustment

If a constant current output is required, set the current value to the load’s rated current.

For instructions on adjusting the current, see Section 4.1.1.

4.2.3 Turning the Output On/Off

Example: I need a constant current output of 4.00 A. Set it up as follows.

Digital CC And CV DC Power Supply-Figure8

Gambar 8

Jumlah Note
Press the output switch to activate the output.
The constant-current indicator is lit, and the device is in constant-current output mode.
The voltage display window shows the output voltage, which does not need to match the set value; however, the set value must be higher than the actual value, otherwise constant-current output will not be possible.
The current display window shows the output current, which matches the set value.
The power display window shows the output power.
The Equivalent Impedance window displays the equivalent impedance of the current output line.

Digital CC And CV DC Power Supply-Figure9

Gambar 9

Press the output switch again to turn off the output.

5. Pemeliharaan dan Perawatan Peralatan Harian

5.1 Perawatan Rutin

5.1.1 Pembersihan Harian

Pastikan untuk mematikan daya dan mencabut kabel daya sebelum membersihkan. Kami menyarankan untuk menyeka perangkat dengan kain katun lembut anti-statis dan deterjen ringan; jangan menyemprotkan cairan langsung ke perangkat. Jangan gunakan pelarut kimia seperti benzena, toluena, atau aseton.

5.1.2 Pemeriksaan Rutin

Before each use, check that the power cord and terminal connections are secure and free of burn marks; ensure that the fan vents are not blocked; and after turning on the unit, verify that the indicator lights and display are functioning properly.

5.2 Spesifikasi Penyimpanan

5.2.1 Persyaratan Lingkungan Penyimpanan

Temperature: 5°C to 35°C; relative humidity: <65% RH; dry, dust-free environment free of corrosive gases.

5.2.2 Operasi Penyimpanan Jangka Panjang

Matikan daya dan cabut semua kabel; bersihkan peralatan dan bungkus dengan plastik anti debu; hidupkan daya selama 10 menit setiap bulan untuk mencegah kerusakan akibat kelembapan pada komponen.

Memecahkan Masalah dan Menyelesaikan Masalah Umum

6.1 Equipment Alarm Indicator Lights

Gejala Kesalahan Kemungkinan penyebab Prosedur
The OV indicator light is on Perlindungan Overvoltage Turn off the output, check that the voltage setting does not exceed the device’s maximum output voltage, and then turn the output back on.
The OC indicator light is on Overcurrent Protection, Load Short Circuit Turn off the output; check that the current setting does not exceed the device’s maximum output current; check the load, rule out a short circuit, and then resume output.
The OP indicator light is on Perlindungan Over-Power Turn off the output, reduce the voltage/current, and lower the load power, then resume output
The OT indicator light is on Perlindungan Over-Temperature Turn off the output, let the device cool down by disconnecting the power, clean the fan vents to improve heat dissipation, and then restart the device once it has cooled down and try again.

6.2 Kerusakan Lainnya

Gejala Kesalahan Kemungkinan penyebab Prosedur
Tidak dapat menyala Kabel catu daya input longgar, sekering putus Turn off the power and check the input power supply cord connection; turn off the power, troubleshoot for short circuits or overloads, replace the fuse with one of the same rating, and do not use a fuse with a higher current rating.
Output parameters are incorrect Incorrect voltage and current settings Please refer to Chapter 4 to set the parameters correctly.
Kipasnya tidak berputar Penumpukan debu, kerusakan motor Turn off the power and clean the dust from the fan. If the problem persists, please contact us for repair or to have the fan replaced.